This Statement of Work defines requirements for procuring a complete spectroradiometer measurement system for the FDA consisting of three integrated components: a high-precision spectroradiometer, specialized lens for near-eye displays, and control/analysis software.
The spectroradiometer must feature a wavelength range of 380-780 nm minimum with a back-thinned and cooled CCD detector containing at least 512 elements. Key performance specifications include wavelength accuracy less than 0.5 nm, luminance accuracy within ±3%, spectral resolution under 5 nm, stray light below 0.06%, and sensitivity of minimum 5×10⁻⁵ cd/m² with a 2-degree internal aperture. The unit must include variable internal apertures (2 degrees, 1 degree, and 0.1-0.25 degrees), an integrated neutral density filter, and USB connectivity, with dimensions under 20 cm × 36 cm × 36 cm and weight under 10 kg. The specialized lens for AR/VR displays must have a focal length of 30-50 mm, fixed pupil diameter between 3-5 mm, adjustable focus from 300 mm to infinity, and luminance accuracy within ±3%. The software must control spectroradiometer operations, acquire and display measurements, analyze data, export to Excel/ASCII formats, and provide SDKs or command-line support compatible with Python, MATLAB, and Windows 10 or later. All equipment must be new and non-refurbished with factory calibration prior to delivery, calibration certificates provided, minimum 1-year warranty, and ongoing recalibration support including on-site services at the FDA facility. Quotes must include itemized pricing for each component with all shipping and handling costs included.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combined Synopsis Solicitation- CDRH-2026-133605- Final.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
STATEMENT OF WORK
Procurement of Spectroradiometer System with Specialized Lens and Software
1. PROJECT OVERVIEW
This Statement of Work defines the requirements for the procurement and delivery of a complete spectroradiometer measurement system consisting of a high-precision spectroradiometer, specialized lens for near-eye display measurements, and associated control and analysis software.
2. SCOPE OF WORK
The contractor shall provide the following items as a complete, integrated system:
Item 1: Spectroradiometer
Optical Specifications:
- Wavelength range: minimum 380-780 nm
- Sensor size: minimum 512 elements
- Detector type: back-thinned and cooled CCD
- Spectral resolution/bandwidth: < 5 nm (variable resolution option preferred)
- Datapoint interval/spectral reporting: 1 nm or less
- Digital resolution: 16 bits
Performance Specifications:
- Wavelength accuracy: < 0.5 nm
- Luminance accuracy: < ±3%
- Integration time range: minimum 10 ms to 60 seconds
- Stray light: < 0.06%
- Polarization error: < 2%
- Sensitivity: minimum 5×10⁻⁵ cd/m² with 2-degree internal aperture measuring
Illuminant A
- Color accuracy (x,y): < ±0.002
- Color repeatability (x,y): < ±0.0005
Hardware Features:
- Variable internal aperture diameters: 2 degrees, 1 degree, and 0.1-0.25 degrees
- Integrated neutral density filter: ND1 or ND2
- Communication interface: USB
Physical Specifications:
- Dimensions: < 20 cm × 36 cm × 36 cm
- Weight: < 10 kg
Item 2: Specialized Lens for Near-Eye Displays
The contractor shall provide a specialized lens optimized for measuring augmented reality (AR) and virtual reality (VR) displays with the following specifications:
• Focal length: 30-50 mm
• Front entrance pupil design with fixed pupil diameter between 3-5 mm
• Adjustable focus range: minimum 300 mm to infinity
• Luminance accuracy: < ±3%
Item 3: Software
The contractor shall provide software with the following capabilities:
Core Functions:
- Control spectroradiometer operations
- Acquire measurements
- Display spectrum and color measurements
- Analyze measurement data
Data Export:
- Export raw data in Excel, ASCII, or equivalent formats
Integration:
- Software Development Kits (SDKs) or command line support for third-party software control
- Compatible with Python and MATLAB
- Compatible with Windows 10 Operating System or later versions
3. GENERAL REQUIREMENTS
Equipment Condition
• All equipment must be new and non-refurbished
Support Services
• Software support must be available during normal business hours
Calibration
• Spectroradiometer must be calibrated prior to delivery
• Calibration certificate or test reports must be provided with delivery
• Contractor must provide ongoing recalibration support as needed, including:
• Calibration services at contractor's facility
• On-site calibration services at FDA facility
Warranty
• Minimum 1-year warranty on all equipment
Pricing
• All quotes must include shipping and handling costs
• Pricing must be itemized for each component (spectroradiometer, lens, software)
4. DELIVERABLES
1. Spectroradiometer system (Item 1) - fully calibrated and tested
2. Specialized lens for near-eye displays (Item 2)
3. Control and analysis software (Item 3) with installation media and documentation
4. Calibration certificate(s) or test report(s)
5. User manuals and technical documentation for all equipment
6. Software documentation including SDK/API documentation
7. Warranty documentation
5. ACCEPTANCE CRITERIA
The delivered system will be accepted upon verification that:
8. All technical specifications listed in Section 2 are met
9. Equipment is new and non-refurbished
10. Calibration certificates are provided and current
11. Software installs and operates correctly on Windows 10 or later
12. All components function as an integrated system
13. Complete documentation is provided
14. All shipping and handling is completed without damage
6. CONTRACTOR RESPONSIBILITIES
The contractor shall:
15. Provide all items listed in Section 2 meeting or exceeding stated specifications
16. Perform factory calibration and provide documentation
17. Package equipment appropriately to prevent shipping damage
18. Provide all necessary cables, accessories, and components for system operation
19. Deliver complete documentation and software
20. Provide technical support contact information
21. Honor warranty terms for minimum one year from acceptance
7. GOVERNMENT RESPONSIBILITIES
The FDA shall:
22. Provide clear delivery location and contact information
23. Inspect delivered equipment within reasonable timeframe
24. Notify contractor of acceptance or rejection with specific reasons
25. Provide appropriate operating environment for equipment