C6400000468_Salient Characteristics_Redacted.pdf
PDF 74 KB Posted
- Attached to
- BENCHTOP PARTICLE COUNTER
- Solicitation number
- SPMYM226Q5916
About this file
This document is a salient characteristics specification sheet for a PAMAS SBSS Particle Counter system being procured through Ayalytical Instruments, Inc. via GSA Advantage.
The PAMAS SBSS Particle Counter is designed for fluid contamination analysis and supports multiple applicable test methods including ISO 4406, SAE AS 4059 E, NAS 1638, GJB 420, GOST 17216, and NAVAIR 01-0A-17. The system operates at pressure ranges up to 5 bar with standard accessory pump or up to 10 bar with high pressure pump or external pressure supply. Three volumetric sensor options are specified: the HCB-LD-50/50 (size ranges 1-200 µm or 1-400 µm optional, maximum 24,000 particles/ml at 25 ml/min flow rate); the HCB-LD-25/25 (size ranges 1-200 µm, maximum 120,000 particles/ml at 25 ml/min); and the SLS-25/25 (size ranges 0.5-20 µm, maximum 13,000 particles/ml at 10 ml/min). The counter features 8, 16, or 32 free adjustable size channels with backlit graphical LC-display (320 x 240 pixel resolution), integrated 32-column thermoprinter, RS-232C data transfer capability, and power supply options at 100V, 115V, 220V, or 230V at 50-60Hz. Equipment setup and configuration requires a PAMAS representative to visit TRF Bangor Scientific Services Division for assembly and laboratory specification customization, with an additional six-month delivery extension requested for this service. The supplier is located at 2787 W Fulton Street, Chicago, Illinois 60612.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Solicitation - SPMYM226Q5916.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
PAMAS SBSS Particle Counter Salient Characteristics
1. Applicable Test Methods: ISO 4406, SAE AS 4059 E, NAS 1638, GJB 420, GOST 17216, NAVAIR 01-0A-17
2. Pressure range
a. Standard accessory pump: up to 5 bar
b. High pressure pump or external pressure supply: up to 10 bar
3. PAMAS HCB-LD-50/50 Volumetric Sensor:
a. Size ranges:
• 1-200 µm (ISO 21501-3)
• 1-400 µm: option on request
• 1-100 µm (ISO 4402)
• 4-70 µm(c) according to ISO 11171
b. Max. particle concentration:
• 24,000 P/ml at a flow rate of 25 ml/min
4. PAMAS HCB-LD-25/25 Volumetric Sensor:
a. Size ranges:
• 1-200 µm (ISO 21501-3)
• 1-100 µm (ISO 4402)
• 4-70 µm(c) according to ISO 11171
b. Max. particle concentration:
• 120,000 P/ml at a flow rate of 25 ml/min
5. PAMAS SLS-25/25 Volumetric Sensor:
a. Size ranges:
• 0.5-20 µm (ISO 21501-2)
• 1-20 µm (ISO 4402)
• 1.5-25 µm(c) (ISO 11171)
b. Max. particle concentration:
• 13.000 P/ml at a flow rate of 10 ml/min
6. Dimensions: 400 mm x 300 mm x 600 mm (W x D x H)
7. Counter
a. 8, 16 or 32 free adjustable size channels; the thresholds are free selectable within the measuring range of the particle sensor
b. Backlit graphical LC-display with 320 x 240 pixel resolution.Data printout: 32 column thermo printer
c. Integrated 32 column thermoprinter
d. Data transfer: 8 Bit ASCII code, RS-232C port
e. Power supply via alternating current: 100 V, 115 V, 220 V, 230 V, 50–60Hz
8. Equipment Set-up and Configuration: request PAMAS representative to visit TRF
Bangor Scientific Services Division to assemble and configure the equipment to laboratory’s specifications. Request to add an additional 6 months to the delivery date to allow the representative to perform this task.
10. Company Information:
Ayalytical Instruments, Inc. via GSA Advantage
2787 W Fulton St
Chicago, IL 60612