C04a Specification Document 1232SA26Q0691 .pdf

PDF 84 KB Posted

Attached to
Time-Domain Nuclear Magnetic Resonance (TD-NMR) Body Composition Analyzer
Solicitation number
1232SA26Q0691
Issued by

About this file

This document contains technical requirements for a benchtop time-domain nuclear magnetic resonance (TD-NMR) body composition analyzer. The required equipment must measure composition parameters including fat mass, lean mass, free fluid content, and T2 water properties in intact meat samples up to 650 grams, as well as small poultry carcasses or bone-in poultry carcass parts. The system must also accommodate tissue samples ranging from 0.1 grams to 10 grams. The analyzer shall operate as a benchtop unit suitable for standard laboratory environments, requiring no cryogens and functioning on standard 110-240V AC laboratory power without specialized shielding or infrastructure modifications.

Performance specifications require measurement accuracy of approximately 1% of total sample mass with reproducibility suitable for multi-operator laboratory environments. The system must complete standard measurements in 3 minutes or less per sample and support minimum daily throughput of 30 samples. The software shall provide automated acquisition and analysis with results exportable in open formats (CSV, XML, or equivalent), support user-defined workflows, include longitudinal study tracking tools, and operate on standard Windows-based computers with basic data-integrity safeguards. Deliverables include the TD-NMR analyzer, probes/sample holders for multiple sample sizes, operational software, operating manuals, training materials, maintenance documentation, technical support, and on-site setup with initial operational training for laboratory personnel.

View the file

Other files

File Type Posted
C04 Sol_1232SA26Q0691.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Technical Requirements

Equipment description:

Benchtop time-domain nuclear magnetic resonance (TD-NMR) body composition analyzer capable of measuring composition (moisture, protein, fat, connective tissue, etc.) and T2 properties (transverse relaxation time, proportion of free and tightly bound water populations, etc.) in intact meat samples up to 650 g. Additionally, the system should also be able to measure composition in either small poultry carcasses or bone-in poultry carcass parts.

General Requirements The instrument shall:

1. Be a benchtop TD-NMR system suitable for laboratory environments.

2. Provide non-invasive, in-vivo body composition measurements.

3. Provide T2 properties for assessing meat/muscle water properties (proportion of free water, immobilized water, tightly bound water).

4. Operate without cryogens and on standard 110–240 V AC laboratory power.

5. Support high-throughput workflows necessary for repeated daily measurements in longitudinal studies.

6. Be fully operational indoors without specialized shielding or infrastructure modifications.

Performance Requirements Measurement Capabilities The system shall:

1. Quantify, at minimum:

o Fat mass o Lean mass o Free fluid (water content) o Device must be able to produce T1 data results o Device must be able to produce or be configured to produce T2 water properties of meat samples

2. Produce quantitative results using T2-based relaxation measurements or equivalent

TD-NMR methodology.

3. Provide body composition values as absolute masses or volumes.

Accuracy and Precision The system shall:

1. Provide measurement accuracy of approximately 1% of total sample mass.

2. Maintain reproducibility suitable for multi-operator laboratory environments.

3. Include factory-calibrated or validated protocols requiring no chemical standards or consumables.

Measurement Duration

1. Complete a standard in-vivo measurement in 3 minutes or less per sample.

2. Support minimum daily throughput of 30 samples per day under routine operation.

Sample Size Capacity The system shall accommodate:

1. Intact meat samples up to 650 g

2. Tissue samples 5-10 g

3. Small tissue biopsies down to 0.1 g

Operational Requirements

1. Be designed for continuous operation under standard laboratory conditions.

2. Require no external cooling, cryogens, or RF shielding.

3. Allow transport within the facility using standard lab carts or internal casters.

4. Include all necessary cables, accessories, and sample holders required for immediate operation upon installation.

Software Requirements The system software shall:

1. Provide automated acquisition and analysis for fat, lean, free-fluid measurements and T2 properties.

2. Export results in open formats (CSV, XML, or equivalent).

3. Permit user-defined workflows and measurement sequences.

4. Include longitudinal study support tools such as repeated-measures tracking.

5. Operate on standard Windows-based laboratory computers.

6. Provide basic data-integrity safeguards consistent with research record-keeping requirements.

Deliverables The vendor shall provide:

1. TD-NMR analyzer meeting all previously listed specifications.

2. Probes/sample holders for both:

o Small tissue samples o Samples up to 650 g

3. Operational software.

4. All operating manuals, training materials, and maintenance documentation.

5. Technical support.

6. On-site setup and initial operational training for laboratory personnel.