C04_Statement of Work.pdf

PDF 170 KB Posted

Attached to
SoloVPE Plus Instrument Federal contract opportunity
Solicitation number
12639526Q0275
Issued by
Department of Agriculture Animal and Plant Health Inspection Service Marketing and Regulatory Programs Business Services

About this file

This is a Statement of Work for the acquisition of a SoloVPE Plus spectrophotometer by the USDA. The government seeks to procure an instrument capable of measuring highly concentrated samples without dilution, using slope-based measurements and requiring no reference standard for quantitation. The device must operate across a sample range from less than 20 µL to 2.5 mL, support multiple wavelengths and pathlengths through a single method, and achieve a maximum pathlength of 15.000 mm with slope repeatability of ≤2% RSD.

The instrument must incorporate a Fibrette Optical component constructed of fused silica and polyimide for UV-VIS measurements spanning wavelengths between 190 nm and 1100 nm, and must be GMP compliant. Buffer or background correction is required only in special circumstances. The contractor must deliver the equipment to USDA, 1920 Dayton Avenue, Ames, IA 50010, within 60 days of award. Evaluation will be based on technical capability using a Lowest Price Technically Acceptable approach. For technical inquiries, contractors should contact Anna Dash at 515-337-6838 or anna.c.dash@usda.gov. No government-furnished equipment is provided for this procurement.

View the file

Other files for this federal contract opportunity

Other files attached to SoloVPE Plus Instrument, newest first.
File Type Posted
A05_Justification for Under SAT_Final.pdf PDF
12639526Q0275.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Statement of Work Project: Acquire a SoloVPE Plus

GENERAL INFORMATION

1.0 Scope of Work: To acquire a SoloVPE Plus.

2.0 Background:

CONTRACTOR REQUIREMENTS

3.0 Technical Requirements/Tasks:

Minimum Must Have Specifications

1. Able to measure highly concentrated sample without dilution

2. Buffer/background correction is only needed in special circumstances

3. Sample range ˂20 µL to 2.5 mL

4. Slope-based measurements

5. No reference standard required for quantitation

6. Able to use different wavelengths and pathlengths using a single method

7. GMP compliant

8. Maximum pathlength: 15.000 mm

9. Slope repeatability: ≤2% RSD

10. Uses an Fibrette Optical component made of Fused silica and polyimide for UV-VIS measurements for wavelengths between 190 nm and 1100 nm

4.0 Government Furnished: N/A

5.0 Deliverables/Schedule : Must be delivered within 60 days of date of award

6.0 Evaluation Factors Technical capability, Lowest Price Technically Acceptable

7.0 Technical Point of Contact: Should there be technical questions, please contact Anna Dash, ph: 515-337-6838, email: anna.c.dash@usda.gov

8.0 Place of Delivery; USDA, 1920 Dayton Avenue, Ames, IA 50010 mailto:anna.c.dash@usda.gov

CONTRACTOR REQUIREMENTS
3.0 Technical Requirements/Tasks:

File details come from the government source that posted it. Updated .