This document is a specifications attachment for a federal procurement of scientific laboratory equipment. The USDA is seeking one Dynamic Vapor Sorption Analyzer to be delivered to the USDA ARS Southern Regional Research Center (SRRC) in New Orleans, Louisiana within 10 weeks.
The analyzer must measure vapor sorption characteristics for scientific samples, specifically cotton byproduct-based biocomposite materials. Minimum technical specifications include humidity generation capabilities ranging from 0% to 98% (±0.5%) RH at 5°C to 60°C, and 0% to 85% (±1%) RH at 60°C to 85°C, with temperature control from 5°C to 85°C (±0.2°C accuracy). The instrument must include an UltraBalance Low Mass Assembly with 1000mg capacity and ±150mg dynamic range, Mass Flow Controllers with high precision and stability, and a Vapor Mixing Manifold with independent sample and reference chambers. The purchase must include a Computer Kit with all Input/Output Cards installed and configured, control software for system operation and experimental design, installation and operation training, and all necessary consumables. An additional requirement specifies that the instrument purchase must include a warranty and/or service plan. Delivery should be made to Stephen Rogers at the specified SRRC address.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment II
SPECIFICATIONS
1.0 Specifications Description:
Dynamic Vapor Sorption Analyzer
Quantity: 1
Unit Type: Each
Minimum Requirements:
• A dynamic water vapor analyzer is required for the measurement of vapor sorption characteristics for scientific samples (for example, cotton byproduct-based biocomposite materials)
• The instrument should be able to provide humidity generation Range and accuracy of 0% to 98% (±0.5%) RH for 5°C to 60°C, and 0% to 85% (±1%) RH for 60°C to 85°C.
• The instrument should be able to measure sorption and desorption isotherms over a broad range of humilities and temperatures.
• The instrument must come with UltraBalance Low Mass Assy with capacity of 1000mg and dynamic range ±150mg.
• Instrument must include a system to provide a controlled and stable temperature environment for experiments in addition to providing anti-condensation protection.
• The control temperature range should be 5°C to 85°C with control accuracy of ±0.2°C.
• Instrument must include Mass flow controllers (MFCs) used to control and measure flow rates with high turn down ratio, control precision, and high long-term stability.
• Instrument must come with Stand and Vapor Mixing Manifold with Independent sample and reference chambers.
• Instrument purchase must include Computer Kit with All Input/Output Cards Installed and Configured.
• Instrument purchase must include the analyzer Control software which provide a complete system control, allows user to design, edit and run automatic or manual vapor sorption experiments.
• Instrument purchase must include installation and operation training, with all necessary consumables for operation
2.0 Place of Delivery
USDA ARS Southern Regional Research Center (SRRC) ATTN: Stephen Rogers 1100 Allen Toussaint Blvd New Orleans, LA 70124