This document is a specifications attachment for a federal procurement of laboratory equipment for the USDA Biologics Development Module (BDM) facility in Manhattan, Kansas.
The procurement requires one (1) cell imaging multimode plate reader specified as brand name or equal to AGILENT CYT9M-SN CYTATION 9 MULTIMODE READER. The system must perform multiple detection modes including UV-Vis absorbance, fluorescence intensity, luminescence, fluorescence polarization, time-resolved fluorescence, and alpha detection. The instrument must accommodate microplate types ranging from 6- to 1536-well plates depending on reading method and must feature both top and bottom reading options for solution-based and adherent cell-based assays. Key technical specifications include temperature control up to 65°C with condensation control and shaking capabilities, CO2 and O2 control options, fluorescence intensity detection wavelength ranges of 250–900 nm for monochromators and 200–850 nm for filters, luminescence detection wavelength of 300–700 nm, and fluorescence polarization detection wavelength range of 280–850 nm. The system must include integrated upright and inverted microscopes with imaging modes (fluorescence, phase contrast, brightfield variants, and color brightfield) and imaging methods (single color, multi-color, montage, time lapse, z-stacking).
The contractor must provide onsite installation, start-up, and software training with qualification verified through customer site acceptance testing. All deliverables must include operation manuals, software manuals, software installation procedures, required PC and software, component confirmation documentation, and regulatory compliance documentation. The purchase price must include all packaging, shipping, and shipping-insurance costs. Government Furnished Property is specified as none.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| C04 Attachment 1 Provisions and Clauses 1232SA26Q1161.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment 2 Specifications
Cell Imaging Multimode plate reader, Manhattan KS
Brand Name or Equal to AGILENT CYT9M-SN CYTATION 9 MULTIMODE READER
Background and Scope The USDA Biologics Development Module (BDM) requires multimode plate reader with capability to work in various detection modes including UV-Vis absorbance, fluorescence intensity, luminescence, fluorescence polarization and time-resolved fluorescence to accurately measure end point, kinetics, spectral scanning, and well-area scanning. The instrument must be compact that is equipped with automated microscope combined with multimode detection to capture quantitative data and phenotypic information in a single system. The system must be incorporated with upright microscope, inverted microscope, gas control option, temperature control, and shaking capabilities. It must also have software with automated image capturing, analyzing and processing in a single system.
Salient Characteristics - Applicable Documents and Standards
Multimode plate reader:
one (1) multimode plate reader is required for accurately reading UV-Vis absorbance, fluorescence intensity, luminescence, fluorescence polarization and time-resolved fluorescence of vaccine and diagnostic proteins for quality control, process monitoring, and research.
Salient Characteristics of the System:
1. must be able to work in detection mode of UV-Vis absorbance, fluorescence intensity, luminescence, fluorescence polarization, time-resolved fluorescence and alpha.
2. must have reading methods involving end point, kinetics, spectral scanning, and well-area scanning.
3. must have top reading option for solution-based assays and bottom reading options for adherent cell-based assays
4. Must be able to accommodate different microplate Types including Monochromator: 6- to 384-well plates, Filters: 6- to 1536-well plates, Imaging: 6- to 1536-well plates
5. Must have temperature control options (up to 65°C) with Condensation Control and Shaking capabilities
6. Must be easily operated with a PC using the software incorporated in the system
7. Must have CO2 and O2 Control options
8. Must have imaging mode including Fluorescence, phase contrast, brightfield, high-contrast brightfield, and color brightfield and imaging method including Single color, multi-color, montage, time lapse, z-stacking
9. must have fluorescence intensity detector for monochromator system and filter system;
monochromators (top/bottom), and filters (top) with Monochromators wavelength range: 250−900 nm and Filters: 200−850 nm
10. Must have luminescence detection wavelength 300-700 nm
11. Must have fluorescence polarization detector with wavelength range of 280-850 nm
12. must have Performance Validation software incorporated with system including notification of required maintenance in the software.
13. Must be compact in size about W42 × D51 × H45 cm
Testing and Verification Must have onsite installation, start-up and software training included.
Qualification via a customer site acceptance test will be the verification standard.
Government Furnished Property None.
Data Requirements The required documentation must include operation manuals, software manuals, software installation procedures. Required PC and software for cell imaging multimode plate reader and imaging must be included.
Documentation must also include confirmation of each component part, accessories, software, and regulatory compliance.
Packaging and Shipping All packaging, shipping, and shipping-insurance costs must be included with purchase price.