This is a Statement of Work (SOW) for microplastics analyses services. The contractor must provide laboratory analysis services using an Agilent 8700 Laser Direct Infrared (LDIR) Chemical Imaging System equipped with quantum cascade laser (QCL) imaging technology. This system detects microplastic particles ranging from 10 μm to 500 μm in size and determines their chemical composition. The contractor shall furnish all facilities, labor, and materials necessary to perform these non-personal services in accordance with the specifications outlined.

Quality assurance and control requirements mandate that the laboratory maintain a software microplastics library containing spectral data of non-microplastics (rubber, naturally occurring polyamide, chitin, cellulosic materials, carbonates, and stearates) to address potential interferences. Field blanks must be collected using deionized water, and procedural blanks must be analyzed during sample filtering, with both blank samples subject to identical preservation methods and holding times as environmental samples. Sample preservation requires storage at approximately 4°C in dark conditions, with samples analyzed within 7-14 days of collection; freezing is prohibited to prevent microplastic structural damage. The laboratory shall provide a comprehensive analytical report containing methodology description, quality control summary, total microplastics detected, microplastics size distribution charts, method blank results, laboratory control sample results, and both sample and method summaries.

View the file

Other files

File Type Posted
Sol_140G0226Q0039_Amd_0001.pdf PDF
Sol_140G0226Q0039.pdf PDF
B08_Attachement_B_-_Price_Schedule.xlsx XLSX spreadsheet

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

ATTACHEMENT A

Statement of Work

MICROPLASTICS ANALYSES

Analysis requires a laboratory equipped with an Agilent 8700 Laser Direct Infrared (LDIR) Chemical Imaging System. This system uses quantum cascade laser (QCL) imaging technology which is a new method for spectral analysis and chemical imaging that advances the ability to detect microplastic (MP) particles, determine their size, and acquire detailed information about their chemical composition. Compared to conventional spectroscopic or microscopic methods, LDIR enables efficient and accurate quantification and identification of MPs in sizes between 10 μm and 500 μm.

QUALITY ASSURANCE/QUALITY CONTROL

To address potential interferences by non-plastic particles from the samples, a software MP library should exist that contains various spectral data of non-microplastics (e.g., rubber, naturally occurring polyamide, chitin, cellulosic materials, carbonates, and stearates which are commonly present in environmental samples).

For field data collection, a field blank should be collected using deionized water to monitor for airborne and procedural contamination. The field blank should be prepared by rinsing clean sampling equipment and collecting the rinse water for analysis. The laboratory should analyze procedural blanks when filtering samples. Blank samples should have the same preservation methods and holding times as that of an environmental sample.

DATA REPORTING

A MP analytical report will be provided by the laboratory that contains methodology description, quality control summary, table of total number of MP detected, chart of MP size distribution, results of the method blank, laboratory control sample, sample summary, and method summary.

SAMPLING HANDLING

Sample Storage:

• Place glass bottles in a cooler box with ice packs (preferred method) at the field site to minimize biological activity. However, if an ice cooler box is not available, it is acceptable to transfer the samples directly into a refrigerator maintained at approximately 4°C upon returning to the laboratory.

• Avoid freezing to prevent MP structural damage.

• Store samples in the dark to avoid UV degradation.

Sample Preservation and Holding Time

• Preservation Method:

o Store samples in a cool (approximately 4°C) and dark environment after collection.

o Avoid exposure to direct sunlight or heat to prevent degradation or biofouling.

o Do not freeze water samples, as freezing and thawing may alter the physical characteristics of microplastics.

• Holding Time:

o Analyze samples within 7-14 days of collection for best integrity.

o If storage exceeds 14 days, validate through quality control whether degradation or changes have occurred.

o For long-term storage, samples may be filtered immediately in the lab and the filters dried and stored until analysis.

GENERAL REQUIREMENTS

The services provided by this order are Non-Personal Services. The contractor shall furnish all facilities, labor, and materials to provide goods/services in accordance with the terms and conditions herein and the specifications set forth above.