This document is a Solicitation for a Tracer Discharge Measurement System in accordance with the provided Salient Characteristics. The U.S. Geological Survey (USGS) Hydrologic Instrumentation Facility has a requirement for one such system to be delivered to the USGS Utah Water Science Center within 30 days of contract award.
The Solicitation, Number 140G0124Q0106, will be issued on or about March 19, 2024, with written quotes due by April 3, 2024 at 8:00 AM EDT. This is a total small business set-aside acquisition with a NAICS code of 334513 and a size standard of 750 employees. The anticipated award date is May 1, 2024. Vendors must be registered in the System for Award Management (SAM) to be eligible for the purchase order. All responsible sources may submit a quote, which will be considered by the agency. Questions must be submitted by March 22, 2024 at 8:00 AM EDT.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Sol_140G0124Q0106.pdf | ||
| B08_Attach_2_FISMA.pdf | ||
| B08_Attach_2_508.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics S24061
TRACER DISCHARGE MEASUREMENT SYSTEM
Salient Characteristics
1 ea. Tracer Discharge Measurement System
1. Shall consist of a portable, compact lightweight device that uses the sodium chloride (Salt) tracer dilution method for determining stream discharge.
2. Shall be suitable for use in fast flowing, turbulent waters.
3. Shall be capable of interfacing with the conductivity probe described below.
4. Shall include a user interface with an integral display.
5. Shall include at least 8 MB of internal memory.
6. Shall be capable of operating over a temperature range from at least -20° to 60°C.
7. Shall be capable of operating from rechargeable AA batteries.
8. Shall be capable of operating for at least 40 hours on a single battery charge, or fresh set of batteries.
9. Shall include a USB-C port.
10. Shall include Bluetooth communications.
11. Shall have a mechanical Ingress Protection (IP) rating of at least IP66.
12. Shall include all necessary calibration supplies, software, and documentation.
13. Software shall include features for displaying and graphing measurement data, guiding the user through the calibration process, post processing data, and exporting data in CSV or XML file formats.
3 ea. Conductivity Probe
14. Shall interface with the device described above.
15. Shall consist of a conductivity probe with integral cable of at least 30 ft. in length.
16. Shall include integrated temperature compensation.
17. Shall have an operating range from at least 0 to 2,000 uS/cm with a resolution of 0.1 uS/cm, or better.
18. Shall include weighting armor.
19. Shall include hand spool for cable.
2 ea. Transport Case
20. Shall include a transport case capable of housing all system components as well as calibration accessories.
21. Shall include protective inserts, a hinged cover, telescopic handle, and wheels.