B08_ATCH_1_SALIENTS_VegaPuls_c_23_Radar-UPDATED_1_22_25.pdf

PDF 707 KB Posted

Attached to
VEGAPULS C23 RADAR SENSOR OR EQUAL IAW SALIENTS
Solicitation number
140G0125Q0042
Issued by

About this file

This document is a Salient Characteristics specification for a Non-Contact Radar Water Level Sensor (Model VegaPuls C 23 or equal) required by the USGS Hydrologic Instrumentation Facility. The technical requirements specify an 80 GHz measurement frequency technology device capable of measuring water levels at data collection sites nationwide with a total measurement uncertainty of no more than 0.01 ft or 0.1% of indicated reading across a temperature range of -40° to 60°C.

The sensor must have specific capabilities including: measurement range up to 30m (98 feet), beam angle no larger than 4°, 12 VDC power source compatibility, SDI-12 interface, IP66/68 mechanical rating, integrated Bluetooth communication, and a minimum 15 ft integral SDI-12 communication/power cable. A mounting bracket option is required. This specification is related to solicitation 140G0125Q0042, which seeks to procure 125 units under a small business set-aside with NAICS code 334513, with quotes due January 30, 2025, at 8:00 AM ES and an anticipated award date of February 20, 2025.

View the file

Other files

File Type Posted
Sol_140G0125Q0042.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics

Non-Contact Radar (Model VegaPuls C 23 Radar Water Level Sensor or equal)

The USGS Hydrologic Instrumentation Facility needs to purchase non-contact radar water level sensors with 80 GHz measurement frequency technology. This equipment will be used for measuring water levels at data collections sites all over the country.

Non-Contact Radar

Shall meet USGS requirements for stage accuracy measurements and shall measure stage with a total measurement uncertainty of no more than 0.01 ft. or 0.1% of indicated reading, whichever is larger over a temperature range of at least -40° to 60°C.

Shall be capable of a measurement range up to at least 0 to 30m (0 to 98 feet).

Shall use 80 GHz measurement frequency technology.

Shall have a beam angle that is no larger than 4o.

Shall be capable of an operating temperature range of at least -40° to 60°C.

Shall be powered from a 12 VDC (nominal) power source.

Shall include an SDI-12 interface.

Shall have a mechanical IP rating of at least IP66/68.

Shall have an integrated Bluetooth communication interface.

Shall include an integral SDI-12 communication/power cable that is at least 15 ft. long.

Shall have an option for a mounting bracket.

File details come from the government source that posted it. Updated .