B08_ATCH_1_SALIENTS_S23101_Microbiology.pdf

PDF 181 KB Posted

Attached to
MICROBIOLOGY ANALYZER IAW SALIENTS
Solicitation number
140G0123Q0186
Issued by

View the file

Other files

File Type Posted
Sol_140G0123Q0186.pdf PDF
B08_ATCH_2_S23101_508.pdf PDF
B08_ATCH_3_S23101_FISMA.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics S23101

MICROBIOLOGY ANALYZER

SALIENT CHARACTERISTICS:

1 EA, Analyzer:

1. Shall consist of a portable, autonomous, self-contained, bacterial analysis system for analyzing viral contamination in public water supplies, water treatment plants, ground water, surface water, or other natural water bodies.

2. Shall be capable of analyzing the presence of E. Coli, Total Coliforms, Fecal Coliforms, or Enterococci.

3. Shall be fully portable, compact, no larger than 0.5 cubic ft., and weight no more than 10 lbs.

4. Shall have a response time no less than 2 hours.

5. Shall use reagents and be fully automated.

6. Shall have measurement range of at least 4 CFU-5 X 105 CFU/100 mL.

7. Shall be able to take at least 6 measurements on one charge.

Shall use GSM network for programing and data management.

i) Shall allow the user to start measurements via cell phone either manually or through pre-programming.

ii) Shall be capable of sending “alert” text messages via a cell phone.

8. Shall have a USB communication interface.

9. Shall be capable of being powered from a rechargeable battery, 12 VDC vehicle outlet, or an AC power source.

10. Shall be capable of operating over a temperature range from at least -10 to 40o C.

11. Shall include an internal, rechargeable, lithium-ion type battery and charger.

12. Shall have integrated internal antenna.

1 EA, Annual Web Based Service:

13. Shall allow for on-line management support services the analyzer system described herein.

1 EA, Disposable Reagent Cartridges:

14. Shall automatically inject/mix reagents and begin incubation and measurement.

15. Shall be compatible with E.Coli measurements.

1 EA, Extra Battery:

16. Shall be compatible with analyzer specs above.

17. Shall be a rechargeable lithium-ion battery and include charging module.

1 EA, Training:

18. Shall consist of at least 1 day of on-site training, or by video teleconferencing depending on travel restrictions.

1EA, Vials (set of 6)

19. Shall include a set of 6 vials.

20. Shall be made of glass with a flat bottom.

21. Shall include caps.

File details come from the government source that posted it. Updated .