Salient Characteristics S23029

RADON MONITOR

1 EA. Radon Monitor:

1. Shall consist of a portable, compact, electronic instrument for detecting and monitoring radon and thoron noble gases.

2. Shall be capable of continuous monitoring of radon in air, sniffing for radon and/or thoron, measuring radon in water, and measuring radon in soil gas.

3. Shall be capable of measuring radon concentrations from 0 to 270,000 pCi/L (picocuries per liter) with an accuracy of +/-5% or better.

4. Shall include a color, sunlight readable, touchscreen display capable of displaying data and graphs. Shall also include display navigation buttons.

5. Shall include Wi-Fi communications and 3 ea. USB ports.

6. Shall include collection sample input/output ports and LED status indicator lights.

7. Shall include internal data storage of at least 16 GB.

8. Shall include a built-in pump for drawing samples for analysis.

9. Shall be capable of operating from 12 VDC power and include a rechargeable lithium-ion battery capable of providing at least 3 days of continuous operation.

10. Instrument case shall have an Ingress Protection (IP) rating of IP67 or better and include a hinged cover with carry handle.

11. Shall include built-in barometer, relative humidity, accelerometer, and temperature sensors.

12. Shall be capable of interfacing to an external thermal printer.

13. Shall weigh no more 10 lbs.

14. Shall include applicable operational software to support system configuration, instrument setup, data transfers, real-time monitoring, graphing and statistical analysis, and data exporting.

15. Shall include all necessary accessories including, but not limited to, 12V adapter with plugs, USB cable for computer connection, desiccant, drying tubes, sample filters, vinyl tubing (at least 3 sections 3 ft.

long each).

1 EA. Radon Water Test Kit:

16. Shall be compatible with the Radon Monitor described above.

17. Shall consist of a device that interfaces to the Radon Monitor and allows the ability to measure radon in water.

18. Shall use no harmful chemicals or hazardous materials.

19. Shall include an adjustable clamp stand and clamp(s).

20. Shall include glass 40 ml (at least 12) and 250 ml (at least 64) vials with septum caps.

21. Shall include a drying and charcoal tube kit.

22. Shall come with a portable carrying case with protective inserts for housing all components of the test kit.