- GAS PYCNOMETER IN ACCORDANCE WITH SALIENTS
- Solicitation number
- 140G0124Q0216
About this file
This document contains the specifications and salient characteristics for a gas pycnometer required by the U.S. Geological Survey Geology, Energy, and Minerals Science Center (GEMSC). The key requirements include: a small tabletop unit less than 4 sq ft and 2 ft in height, temperature stabilization, touchscreen interface, network and USB connectivity, calibration to NIST standards, rapid measurement time of less than 15 minutes, and the capability to analyze a variety of sample types. The instrument must be new, include a one-year warranty, and additional training/consultation if requested.
The related federal contract opportunity is Solicitation Number 140G0124Q0216 for the "GAS PYCNOMETER IN ACCORDANCE WITH SALIENTS", which will be issued as a Request for Quote (RFQ) around 06/28/2024 with quotes due by 07/09/2024. This is a total small business set-aside with a NAICS code of 334516 and size standard of 1000 employees. The anticipated award date is 08/01/2024, and the delivery requirement is within 60 days of the contractor receiving the purchase order. Interested vendors must be registered in SAM.gov.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Sol_140G0124Q0216.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications/Salient Characteristics
Salient Characteristics that must be met are the following:
Gas pycnometer:
• Small tabletop unit less than 4 ft2 in area and less than 2 ft in height.
• Thermoelectric temperature stabilization of measurement cell within +/- 0.25°C of set temperature
• Hinged self-aligned lid for repeatable cell closure
• Touchscreen interface to provide standalone operation of instrument, including performing measurements, reviewing results, printing, reports, and defining methods.
• Laboratory network connectivity over TCP-IP through ethernet or WiFi connection
• USB connectivity for integration with lab balance or peripheral input devices
• Include calibration volume spheres, calibrated to relevant NIST tests, in accordance with ISO, and
ANSI certifications
• Determination of true volume and true density of solids, powders, and slurries, providing accuracy to 0.02% and repeatability of 0.01%
• Rapid measurement of less than 15 minutes per sample including multiple replicates to ensure data quality and assess measurement precision
• Cell volume for maximum sample volume must be between 95 cm3 and 140 cm3
• Pycnometer must connect seamlessly to gas tank source and cell must be purged
• Analysis must be automated upon closure of cell, including cell purging, replicate measurements and venting.
• Operate with free interoperable software loaded onto the instrument
• Capability to analyze oven dried or water saturated samples
• Capability to analyze lithified rock samples or disaggregated sediments
Miscellaneous:
• The instrument must be new and a model currently in production. Refurbished or demonstrator instruments are not acceptable.
• The instrument warranty must be one-year parts, labor, and travel.
• Must include at least 1 day of additional training or application consultation (either on-site of off-site) if requested.
Required Accessories:
• A durable power cord for standard North American wall outlets
• High quality, dual stage, stainless steel diaphragm gas regulator for inert gases must be provided, which meets delivery pressure range of 0-25 or 0-30 psig and includes North American CGA-580 fitting.