GAS PYCNOMETER IN ACCORDANCE WITH SALIENTS
Solicitation number
140G0124Q0216
Issued by

About this file

This document contains the specifications and salient characteristics for a gas pycnometer required by the U.S. Geological Survey Geology, Energy, and Minerals Science Center (GEMSC). The key requirements include: a small tabletop unit less than 4 sq ft and 2 ft in height, temperature stabilization, touchscreen interface, network and USB connectivity, calibration to NIST standards, rapid measurement time of less than 15 minutes, and the capability to analyze a variety of sample types. The instrument must be new, include a one-year warranty, and additional training/consultation if requested.

The related federal contract opportunity is Solicitation Number 140G0124Q0216 for the "GAS PYCNOMETER IN ACCORDANCE WITH SALIENTS", which will be issued as a Request for Quote (RFQ) around 06/28/2024 with quotes due by 07/09/2024. This is a total small business set-aside with a NAICS code of 334516 and size standard of 1000 employees. The anticipated award date is 08/01/2024, and the delivery requirement is within 60 days of the contractor receiving the purchase order. Interested vendors must be registered in SAM.gov.

View the file

Other files

File Type Posted
Sol_140G0124Q0216.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications/Salient Characteristics

Salient Characteristics that must be met are the following:

Gas pycnometer:

• Small tabletop unit less than 4 ft2 in area and less than 2 ft in height.

• Thermoelectric temperature stabilization of measurement cell within +/- 0.25°C of set temperature

• Hinged self-aligned lid for repeatable cell closure

• Touchscreen interface to provide standalone operation of instrument, including performing measurements, reviewing results, printing, reports, and defining methods.

• Laboratory network connectivity over TCP-IP through ethernet or WiFi connection

• USB connectivity for integration with lab balance or peripheral input devices

• Include calibration volume spheres, calibrated to relevant NIST tests, in accordance with ISO, and

ANSI certifications

• Determination of true volume and true density of solids, powders, and slurries, providing accuracy to 0.02% and repeatability of 0.01%

• Rapid measurement of less than 15 minutes per sample including multiple replicates to ensure data quality and assess measurement precision

• Cell volume for maximum sample volume must be between 95 cm3 and 140 cm3

• Pycnometer must connect seamlessly to gas tank source and cell must be purged

• Analysis must be automated upon closure of cell, including cell purging, replicate measurements and venting.

• Operate with free interoperable software loaded onto the instrument

• Capability to analyze oven dried or water saturated samples

• Capability to analyze lithified rock samples or disaggregated sediments

Miscellaneous:

• The instrument must be new and a model currently in production. Refurbished or demonstrator instruments are not acceptable.

• The instrument warranty must be one-year parts, labor, and travel.

• Must include at least 1 day of additional training or application consultation (either on-site of off-site) if requested.

Required Accessories:

• A durable power cord for standard North American wall outlets

• High quality, dual stage, stainless steel diaphragm gas regulator for inert gases must be provided, which meets delivery pressure range of 0-25 or 0-30 psig and includes North American CGA-580 fitting.