This document is a Salient Characteristics specification for a nitrate sensor procurement (S26031).

The requirement consists of two deliverables: one nitrate sensor unit and maintenance/calibration/repair services. The sensor must be a single-parameter optical UV-based device capable of continuous measurement in drinking water, wastewater, or activated sludge environments without requiring reagents. Technical specifications include a 2 mm path length, measuring range of at least 0.05 to 50.0 mg/L, accuracy of 5% of reading ±0.1 mg/L or better, minimum resolution of 0.01 mg/L, and operating temperature range from 2 to 40°C. The sensor housing must be stainless steel with an IP68 or better ingress protection rating, include sludge compensation features and a built-in self-cleaning wiper that requires no tools for replacement, and weigh no more than 12 lbs including a 10-meter polyurethane cable.

The maintenance and calibration component specifies services for Hach NT3100sc nitrate sensors, including two preventive maintenance services per year with full repair coverage for parts and labor. Additional requirements include unlimited priority technical support and troubleshooting, priority 5-day instrument turnaround time, and 2-day return shipping. The sensor must be compatible with a controller described elsewhere in the solicitation documents.

View the file

Other files

File Type Posted
B08_ATCH_3_FISMA.pdf PDF
Sol_140G0126Q0122.pdf PDF
B08_ATCH_2_508.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics S26031

NITRATE SENSOR

SALIENT CHARACTERISTICS:

1 EA. Nitrate Sensor:

1. Shall consist of a single-parameter sensor for measuring nitrate concentrations.

2. Shall be suitable for continuous measurement in drinking water, wastewater, or activated sludge environments.

3. Measurements shall be based on the optical UV measurement method and not require the use of reagents.

4. Shall be compatible with the controller described below.

5. Shall have a 2 mm path length.

6. Shall have a measuring range over a range of at least 0.05 to 50.0 mg/L.

7. Shall have an accuracy of 5% of reading +/-0.1 mg/L, or better.

8. Shall have a minimum resolution of 0.01 mg/L.

9. Shall be capable of operating over a temperature range from at least 2 to 40o C.

10. Housing shall be made of stainless steel and have an Ingress Protection (IP) rating of IP68 or better.

11. Shall include sludge compensation features.

12. Shall include a built-in self-cleaning wiper. Wiper shall not require tool(s) for replacement.

13. Shall include a polyurethane cable with a minimum length of 10m.

14. Shall weigh no more than 12 lbs. including a 10m cable.

1 EA. Maintenance, Calibration and Repair for Nitrate Sensor:

15. Shall include maintenance, calibration/certification, and repair services for Hach NT3100sc nitrate sensors.

16. Shall include 2 services per year consisting of preventive maintenance services and full coverage for any repairs (parts and labor).

17. Shall include unlimited, priority technical support and troubleshooting.

18. Shall include priority 5-day instrument turnaround time.

19. Shall include 2-day return shipping time.