- Refrigerated Water Samplers IAW Salients
- Solicitation number
- 140G0126Q0121
About this file
This is a Salient Characteristics document specifying requirements for refrigerated water samplers and associated equipment. The procurement calls for two (2) each of refrigerated water samplers, sampling bottles configurations, suction lines, and pump tubing.
The refrigerated water samplers must be automated units designed for outdoor, corrosive environments with corrosion-resistant and UV-resistant materials. Key operational specifications include 120 VAC, 60 Hz power supply, SDI-12 compatibility, at least 512 KB data logging memory, EPA-recommended sample preservation at 39°F (4°C), built-in freeze prevention, capacity for 24 polypropylene 1-liter bottles, and maximum dry weight of 165 lbs. The pump shall support vinyl suction tubing 3 to 25 feet long, provide automatic head height compensation, lift capacity of 25 feet, tubing life of at least 1,000,000 pump counts, repeatability of ±5 ml or ±5%, and sample delivery velocity of at least 2 ft./sec. The controller requires non-volatile ROM memory, operational temperature range of 32°F to 120°F, NEMA 4X, 6 (IP67) enclosure rating, clock accuracy of 1 minute per month, and capability for programmable composite samples from 1 to 999 samples. Operations specifications include programmable sample volumes of 10 to 9,990 ml in 1 ml increments, sampling modes for uniform time, non-uniform time, flow, and event, automatic suction line rinsing up to three times per collection, sample retry capability, up to 24 stop/resume commands, storage for at least five sampling programs, and diagnostic testing for RAM, ROM, pump, display, and distributor functions. Associated equipment includes sampling bottle configurations with racks, retaining rings, caps, discharge tubes, and minimum 24 polypropylene 1-liter bottles; 3/8" x 25 lg. vinyl suction lines with standard strainer and coupler; and pre-cut silicone rubber pump tubing with alignment bands (5 pieces cut to 32.25" length).
View the file
Other files
| File | Type | Posted |
|---|---|---|
| B08_ATCH_3_FISMA.pdf | ||
| Sol_140G0126Q0121.pdf | ||
| B08_ATCH_2_508.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics S26014
REFRIGERATED WATER SAMPLERS
SALIENT CHARACTERISTICS:
2 EA, Refrigerated Water Sampler:
1. Shall consist of a refrigerated style automated water sampler intended for use in outdoor, corrosive environments.
2. Shall be made of corrosion resistant materials and include UV resistant features.
3. Shall be powered from a 120 VAC, 60 Hz external power source.
4. Shall be SDI-12 compatible.
5. Shall have at least 512 KB of memory for data logging.
6. Shall be able to keep samples preserved at the EPA-recommended temperature of 39°F (4°C).
7. Shall include a built-in heater to prevent samples from freezing.
8. Shall be compatible with polypropylene 1-liter bottles and capable of holding 24 bottles for sample collection.
9. Shall weigh no more than 165 lbs. when dry without battery.
Pump:
10. Shall support intake suction tubing from 3 to at least 25 ft long made of vinyl with inside dimension no larger than 3/8 inch (1 cm).
11. Shall automatically compensate for changes in head heights.
12. Shall have a lifting capacity of at least 25 feet (7.6 m).
13. Shall have pump tubing life of at least 1,000,000 pump counts.
14. Shall have repeatability of at least ±5 ml or ±5% of the average volume in a set.
15. Shall have a tubing life indicator feature alerting when it’s time to change the tubing.
16. Shall have a sample delivery velocity of at least 2 ft./sec. at heads up to 25 ft.
Controller:
17. Shall have non-volatile ROM memory.
18. Shall have operational temperature range from 32° to at least 120°F (0° to 49°C).
19. Shall allow for programmable composite samples from 1 to at least 999 samples.
20. Shall have NEMA 4X, 6 (IP67) enclosure rating at a minimum.
21. Shall have clock accuracy of at least 1 minute per month, for real time clock.
Operations:
22. Shall sample programmable sample volumes from 10 to at least 9,990 ml in 1 ml increments.
23. Shall have sampling modes in uniform time, non-uniform time, flow, and event.
24. Shall have sample frequency of 1 minute to at least 99 hours and 59 minutes, in at least 1-minute increments non-uniform times in minutes or clock times 1 to at least 9,999 flow pulses.
25. Shall allow for automatic rinsing of suction line up to at least three (3) rinses for each sample collection.
26. Shall allow for sample retries, if sample is not detected and shall be user selectable.
27. Shall allow for sampling stop/resume with up to at least twenty-four (24) real time/date sample stop/resume commands.
28. Shall have program storage for at least five (5) sampling programs.
29. Shall be capable of running diagnostics test for RAM, ROM, pump, display, and the distributor.
Salient Characteristics S26014
2 EA, Sampling Bottles Configuration:
30. Shall be compatible with the sampler specs listed above.
31. Shall include bottle rack, retaining ring, caps, and minimum of two (2) discharge tubes.
32. Shall come in quantity of at least 24 polypropylene bottles.
33. Shall be capable of holding a minimum of 1 liter per bottle.
2 EA, Suction Line:
34. Shall be compatible with the sampler specs listed above.
35. Shall consist of 3/8” x 25 lg. vinyl suction line with standard strainer and coupler.
2 EA, Pump Tubing:
36. Shall be compatible with the sampler specs listed above.
37. Shall consist of pre-cut silicone rubber pump tubing with alignment bands.
38. Shall include a qty. of 5 ea. pieces cut to 32.25” in length.