This document is a Salient Characteristics specification for an XY Automated Sampler system (Reference S26012). The specification details requirements for one automated sampler unit and one sampling plate to be delivered.
The XY Automated Sampler must be an XY automated sampling system employing Fourier Transform Infrared Spectroscopy (FTIR) analytical techniques capable of analyzing multiple samples with both diffuse reflection and transmission sampling modes. The system shall accept standard 24-, 48-, or 96-well microplates and support custom mapping profiles with automated data collection. Technical specifications include gold-coated optics with an integrated Deuterated Triglycine Sulfate (DGTS) detector, elliptical optics with 3X beam demagnification at a 30° angle of incidence, and precision standards of ±5 µm repeatability, ±25 µm accuracy, and 1 µm resolution or better. The unit shall include USB connectivity, operate on 100–240 VAC 50/60 Hz power, feature a fully enclosed housing with purgeable optical path and front-loading tray, weigh no more than 10 lbs (4.5 kg), and measure no larger than 6.5" W x 13.5" D x 6" H (including micrometer). The sampler shall include in-compartment mounting compatibility with most FTIR spectrometers, standard purge tubes with connections, and all necessary accessories and software for setup, operation, and data collection. The accompanying sampling plate shall be a 24-well diffuse reflection plate designed for disposable cups and compatible with the sampler unit.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Sol_140G0126Q0035.pdf | ||
| B08_ATCH_2_508.pdf | ||
| B08_ATCH_3_FISMA.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Salient Characteristics
S26012
XY Automated Sampler
SALIENT CHARACTERISTICS:
1ea. Automated Sampler
1. Shall consist of an XY automated sampling system that employs Fourier Transform Infrared Spectroscopy (FTIR) analytical techniques and automates the process of analyzing multiple samples.
2. Shall include sampling modes for diffuse reflection and transmission.
3. Shall accept standard 24-, 48-, or 96-well microplates.
4. Shall support custom mapping profiles and automated data collection.
5. Shall include gold coated optics with integrated Deuterated Triglycine Sulfate (DGTS) detector.
6. Shall include a USB interface for computer connectivity.
7. Shall employ elliptical optics with 3X beam demagnification and an angle of incidence of 30°.
8. Shall have repeatability, accuracy, and resolution of +/- 5 µm, +/- 25 µm, and 1 µm, or better.
9. Shall include a removable 96-well diffuse reflection plate.
10. Shall be capable of being powered from a 100–240 VAC, 50/60 Hz, power source.
11. Housing shall be fully enclosed with purgeable optical path and front loading tray.
12. Shall include in-compartment mounting features compatible with most FTIR spectrometers.
13. Shall be no larger than 6.5” (165 mm) W x 13.5 (343 mm) D x 6” (152) mm H (including micrometer) and weigh no more than 10 lbs. (4.5 kg).
14. Shall include standard purge tubes and tube connections.
15. Shall include any accessories and applicable software, as required, for system setup, operation, and data collection.
1ea. Sampling Plate
16. Shall be compatible with the sampler described above.
17. Shall consist of a 24-well diffuse reflection sampling plate for disposable cups.