B08_ATCH_1_SALIENTS.pdf

PDF 79 KB Posted

Attached to
PORTABLE LIQUID SAMPLER IAW SALIENTS Federal contract opportunity
Solicitation number
140G0125Q0039
Issued by
Department of the Interior US Geological Survey Office of Acquisitions and Grants

About this file

This is a salient characteristics document specifying requirements for a Portable Liquid Sampler to be used for collecting water quality samples from data collection gaging stations near lakes, rivers, and streams across the US.

The required sampler must include a controller and pump, support both 24 1-liter polypropylene bottles and 24 350ml glass bottles, and be programmable for different sample frequencies and volumes. Key technical specifications include: built-in SDI-12 interface with window, adjustable air purge feature, 28-foot minimum lift capability, IP67 or higher enclosure rating, 512KB minimum memory, and 12 VDC external power operation. The unit must also include a liquid detection sensor, 3/8-inch tube intake and coupler, minimum 6-foot power cable, and bottle retaining ring. This solicitation (140G0125Q0039) is a small business set-aside under NAICS 334513, with quotes due January 27, 2025 at 11:00 AM ES. The US Geological Survey HIF requires delivery to Tuscaloosa, AL within 60 days of purchase order receipt, with an anticipated award date of March 1, 2025.

View the file

Other files for this federal contract opportunity

Other files attached to PORTABLE LIQUID SAMPLER IAW SALIENTS, newest first.
File Type Posted
Sol_140G0125Q0039.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics

Portable Liquid Sampler

This equipment will be used to collect water quality samples from various data collection gaging stations located near lakes, rivers, and streams all over the US.

Salient Features for Sampler

Shall include a controller and pump Shall support bottle configurations of 1-liter polypropylene bottles (24 each) and 350 ml glass bottles (24 each) Shall be programmable to allow for different sample frequencies and sample volumes Shall have a build-in SDI-12 interface with window Shall include an adjustable air purge feature Lift capability shall be at least 28 feet Enclosure shall have an IP rating of at least IP67 Shall include at least 512KB of memory Shall operate from external 12 VDC power Shall include a liquid detection sensor Shall include a 3/8-inch tube intake Shall include 3/8-inch tubing coupler Shall include cable for external 12 VDC power connection; min 6-foot long Shall include bottle retaining ring