About this file

This document outlines salient characteristics for refrigerated portable water samplers and accessories required under Solicitation Number 140G0123Q0268 by the Department of the Interior US Geological Survey Office of Acquisitions and Grants. The salient characteristics specify requirements for three wastewater samplers including pump capabilities, materials, power sources, bottle compatibility, and interfaces. Additional requirements are provided for the pump including lifting capability, liquid detection, tubing specifications, repeatability, velocity, and tubing life. The controller must include a user interface, memory, temperature range, and be SDI-12 compatible. The sampler must have storage for five sampling programs, allow sample frequencies from one minute to 99 hours and 59 minutes, support automatic rinsing, and be capable of running diagnostics tests. Bottles must be compatible with the sampler and consist of fourteen 950-ml polyethylene bottles with caps, distributor arm, discharge tubes, bottle carrier, and adapter.

View the file

Other files for this federal contract opportunity

Other files attached to BLZZRD Sampler for AC OR EQUAL IAW SALIENTS, newest first.
File Type Posted
Sol_140G0123Q0268_Amd_0004.pdf PDF
Sol_140G0123Q0268_Amd_0003.pdf PDF
Sol_140G0123Q0268_Amd_0002.pdf PDF
Sol_140G0123Q0268_Amd_0001.pdf PDF
Sol_140G0123Q0268.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics S23141

REFRIGERATED PORTABLE WATER SAMPLERS AND ACCESSORIES

SALIENT CHARACTERISTICS:

3 EA, Wastewater Sampler:

1. Sampler shall consist of a pump, controller, top cover, center section, base, distributor arm and two pump tubes and power cord. Does not include bottle configuration kit.

2. Shall be made of corrosion resistant, impact resistant, UV resistant materials.

3. Shall be capable of being powered from external 120 VAC, 60 Hz or 12VDC power sources.

4. Shall be compatible with polypropylene 950 ml bottles and capable of holding 14 bottles for sample collection.

5. Enclosure shall have a National Electrical Manufacturers Association (NEMA) 4X, 6 (IP67) enclosure rating, at a minimum.

6. Shall be capable of interfacing with a cell phone modem.

7. Shall be portable and include handles for transporting.

8. Shall include all necessary parts, cables, manual, etc., and applicable software for system operation.

Pump:

9. Shall have a lifting capability of at least 28 feet.

10. Shall include a non-contact type liquid detector device that detects when liquid sample reaches the pump to automatically compensate for changes in head heights.

11. Shall support intake suction tubing up to 99 feet (1 to 30 m) long and be made of vinyl with inside dimension 3/8 inch (1 cm).

12. Shall be repeatable within ±5 ml or ±5% of the average volume in a set, or better.

13. Shall have a sample delivery velocity of at least 2.7 FT./s at 15 ft. (0.83 m/s at 4.6 m)

14. Shall have a pump tubing life of at least 1,000,000 pump counts.

Controller:

15. Shall include a user interface control panel with backlit display and keypad.

16. Shall have non-volatile ROM memory.

17. Shall have operational temperature range from 32° to at least 120°F (0° to 49°C).

18. Shall be SDI-12 compatible.

Operations:

19. Shall have storage capability for at least five (5) sampling programs.

20. Shall have sample frequency of 1 minute to at least 99 hours and 59 minutes, in at least 1-minute increments.

21. Shall allow for automatic rinsing of suction line up to at least three (3) rinses for each sample collection.

22. Shall be capable of running diagnostics test for RAM, ROM, pump, display, and the distributor.

23. Shall sample programmable sample volumes from 10 to at least 9,990 ml in 1 ml increments.

3 EA, Bottles:

24. Shall be compatible with the sampler described above.

25. Shall consist of fourteen (14), 950-ml polyethylene bottles, with caps.

26. Shall include distributor arm, at least two (2) discharge tubes, bottle carrier and adapter.