United States Department of Agriculture Agricultural Research Service

STATEMENT OF WORK

INSTALL AUTOCLAVE ELECTRICAL

PHYSIOLOGY AND PATHOLOGY OF TREE FRUITS

RESEARCH UNIT

HOOD RIVER, OR

Date: February 21, 2023

STATEMENT OF WORK

1.0 CONTRACT PROJECT:

a. Project Number: TBD

b. Project Title: Install Autoclave Electrical

c. Project Location: Physiology and Pathology of Tree Fruits Research Unit Hood River Worksite

3005 Experiment Station Drive; Hood River, OR 97031

2.0 PROJECT CONTACT PERSONS:

a. CONTRACTING OFFICER (CO):

TBD

USDA-ARS-PWA

Email: Tel. No.:

b. CONTRACTING SPECIALIST (CS):

TBD

USDA-ARS-PWA

Email: Tel. No.:

c. LOCATION MONITOR (LM):

Rachel Leisso Physiology and Pathology of Tree Fruits Research Unit 3005 Experiment Station Drive; Hood River, OR 97031 Email: Rachel.Leisso@usda.gov Tel. No.: 541-561-1420

d. AREA SAFETY AND HEALTH MANAGER

Ken Cushman

USDA-ARS-PWA

800 Buchanan Street Albany, CA 94710 Email: ken.cushman@usda.gov Tel. No.: 510-529-5921

e. ENGINEERING PROJECT MANAGER (EPM) Peter Lane

USDA-ARS-PWA

800 Buchanan St. Albany, CA 94710

Email: peter.lane@usda.gov Tel. No.: 510-559-6340 mailto:Rachel.Leisso@usda.gov mailto:ken.cushman@usda.gov mailto:peter.lane@usda.gov

3.0 PROJECT DESCRIPTION:

Provide electrical connections for new autoclave equipment.

4.0 PROJECT REQUIREMENTS:

a. Do not disturb hazardous materials and report condition immediately to the Contracting Officer’s Representative (COR) regarding potentially hazardous conditions that are uncovered. Stop work in the area of the questionable material or condition until identification and direction is provided.

b. The Contractor’s work shall adhere to the criteria stated within these project requirements, drawings, and specifications. All construction shall comply with applicable federal, state, and local code requirements as well as current industry standards, (i.e., IBC, NEC, NSPC, NFPA, EPA, ASTM, GSA, AASHTO, etc.) and health and safety codes, in effect at the time of contract award.

c. Access the site between 8:00 am and 4:00 pm Monday through Friday excluding federal holidays. Contractor staff must sign-in on the first day of arrival and then sign-out on the final day of the project, upon completion.

4.1 Background & Site Description:

The USDA is leasing newly renovated space at for a postharvest physiology, quality, and pathology of sweet cherry and pear research program (Hood River Worksite). This program is administratively part of the USDA-ARS Tree Fruit Research Lab in Wenatchee, WA. We now require utility modifications to serve leased space for mission critical equipment. The equipment is an autoclave.

An autoclave is used to sterilize laboratory tools, glassware, and biological materials, as well as prepare sterile culture media for microorganisms and plants. An autoclave is essential to the plant pathology component of the postharvest sweet cherry and pear research program at the Hood River Worksite.

4.1 Bid Schedule:

Provide all labor, equipment and materials to perform this Statement of Work and Schedule of Items.

4.2 Description of Work:

• Provide emergency shut-off switches at outside of three exits where indicated by Government o Exit 1 is 19 ft from the autoclave o Exit 2 is 48 ft from the autoclave o Exit 3 is 25 ft from the autoclave

• The autoclave steam generator requires 240V, 3 phase (50/60 hz), with 72.2 amps/phase.

Provide dedicated circuit and connection for new autoclave steam generator o Provide one new 3-phase distribution panel with shunt trip 80-amp circuit breaker tied into shut-off switches.

o The panel is to be fed from existing service in the building. The existing panel is approximately 60 feet from the autoclave

• Provide equipment connection within 6 ft of the autoclave

• Provide a lockable and accessible (cannot be located behind the autoclave) high voltage disconnect in order to enable steam generator

• Provide all other work required to connect autoclave electrical. See attached diagram for further information on autoclave electrical requirements (Autoclave Lab 110 Series).

• Assist with final electrical connections and testing

4.3 Government Furnished Equipment\Services

a. Government will furnish a reasonable amount of electricity and water.

5.0 GENERAL REQUIREMENTS:

5.1 Permits: Where governing regulations and imposed codes and standards require notices, permits, licenses, inspections, tests, and similar items or actions in order to lawfully proceed with the required work, the Contractor shall obtain items and take those actions in accordance with the regulations of the governing authority. The costs of such permits, licenses, inspections, etc., are the obligation of the Contractor.

5.2 Existing Installations / Demolition: Contractor shall modify existing installations as needed to accommodate the new installation. Contractor shall coordinate modification of existing installations with Government. All disturbances of existing systems shall be thoroughly coordinated with the Government to prevent disturbances that might jeopardize on-going operations and research. Government reserves the right to retain any equipment / materials removed as part of the scope of work. Contractor shall coordinate any equipment / materials to be retained with Government prior to starting demolition operations.

Contractor shall legally dispose all any materials removed that will not be retained.

5.3 Installation: It shall be the responsibility of the Contractor to complete work in accordance with Federal, State, and Local Code Standards. Additional procedures and materials may be required than is indicated herein. If any item specified herein conflicts with the applicable codes and recommendations, the Contractor shall bring it to the attention of the Contracting Officer Representative for immediate resolution.

5.4 Submittals: Provide manufacturer’s descriptive literature/shop drawings for all materials to the Location Monitor for review. The descriptive literature shall be annotated/ highlighted to show its conformance with contract requirements. Submittals are required for new materials and will be stamped and/or signed by the general contractor prior to submission to the Government. Contractor is responsible for any costs incurred for equipment ordered without prior approval.

5.5 Quality Control: The work performed under this contract shall be subject to continuous audit by the Location Monitor. Quality control is the exclusive responsibility of the Contractor. The Contractor shall retain and pay for the services of an independent inspector or testing laboratory to perform all tests expressly required by this contract.

Rachel.Leisso Typewritten Text service and preventative maintenance

5.6 Protection of Work and Property/Safety Requirements:

a. The Contractor shall protect the work, the site, and all existing property and structures within the limits of the construction activities or that may be affected thereby until acceptance of the work. Any damage to property shall be repaired at the Contractor’s expense, to pre-damaged condition to the satisfaction of the Location Monitor.

b. Safety and health matters, as they relate to the work, are the exclusive responsibility of the Contractor. The Contractor shall furnish, erect, and maintain barricades, warning lights, signs, guards, or take other precautions as may be required by law or local authorities of the protection and security.

c. Contractor is responsible for verifying location of any underground utilities and is responsible for any damaged caused to such utilities.

5.7 Temporary Facilities:

a. Temporary utilities (water and electricity) may be furnished to the Contractor by the Government, if so requested. The Contractor shall maintain and protect all such utilities during the course of construction and shall repair or replace any items damaged through its own negligence.

b. Should the Contractor require any power or utility shutdowns during construction, the

Contractor shall obtain approval for such shutdown from the Location Monitor fourteen

(14) calendar days prior to the need.

5.8 Materials and Equipment:

a. The material and equipment to be furnished under these contract documents shall be the standard products of manufacturers regularly engaged in the production of such materials and equipment and shall be the manufacturer’s latest standard design.

b. All materials and equipment to be incorporated in the work shall be handled and stored by the manufacturer, fabricator, supplies and Contractor before during and after shipment in a manner to prevent any injury, theft or damage of any kind to the materials and equipment. All materials and equipment furnished by the Contractor shall be subject to inspection and approval by the CO/Location Monitor.

c. Government will not furnish any labor or equipment to unload or move materials or supplies.

5.9 Cleanup: The construction site shall be kept clean and free of debris at all times. Restore areas disturbed as nearly as possible to the original appearance and condition. Upon completion of the project, and prior to final inspection, the Contractor shall remove from the premises all unused material, trash, and debris resulting from the work to the satisfaction of the Location Monitor.

5.10 Warranty:

a. Submit data concerning the standard one year warranty for workmanship, including the warranty periods (dates), and warranty contacts with names, addresses and telephone numbers.

5.11 Final Payment/Contract Closeout: The following documents are REQUIRED PRIOR TO

FINAL PAYMENT (other documents may also be required):

a. General contractor’s minimum one (1) year warranty on work.

b. Release of liens

c. Final Invoice

6.0 Project Schedule:

The Contract time for this project is 60 calendar days from contract award.

7.0 Figures

Autoclave Lab 110 Series Manufacturer Information

Pencil

STERIS Corporation • 5960 Heisley Road • Mentor, Ohio 44060

STERIS Corporation Equipment Submittal

TRANSMITTAL

Total number of pages: 1 of 1 May 20, 2022

To: US Dept of Agriculture Address: 1104 N Western Ave Wenatchee, WA 98801

Attn: Brian Fien Ph: 509-335-3349 E-mail: brian.fien@usda.gov

PO#: 12905B22F0053

STERIS Order #: 14411505

From: Mailyn Mendoza

5960 Heisley Rd Mentor OH 44060

Ph: 440-392-7407

FX: 440-350-7077

E-mail: Mailyn_Mendoza@steris.com

STERIS IS TRANSMITTING:

Date Description Rev.

5/20/22 Equipment Submittals for Approval 0

Approval Needed For Reference Only Certified for Construction

Comments:

cc:

Important: Seismic installation price includes slab on grade method only. Any other method of seismic installation including but not limited to, “upper floor” or “through bolting” or “elevated slab” method is not included in the STERIS price or current scope of supply. Additional charges will apply if through bolting is required. These charges will be determined after specific site conditions are investigated. See original quote language: “Seismic Anchoring (Excludes x-raying the floor, through bolting, & associated hardware)”

APPROVAL FORM

Date: May 20, 2022

PROJECT: US DEPT OF AGRICULTURE

PO: 12905B22F0053

STERIS ORDER: 14411505

STERIS would like to thank you for your recent order. This equipment package relates to the items specified by your purchase order number referenced above. It is the intent of this package to promote a clear understanding (between all parties) of the space requirements for and mechanical details of the equipment to be furnished by STERIS Corporation.

The enclosed documents describe equipment only as it is routinely manufactured to conform to the standards cited in those documents and statutory requirements prevailing in most communities in which the equipment likely would be installed.

Unless specifically called for in your invitation for bids, request for quotations, contract, or order, the said equipment will not include any different or additional devices or appurtenances that might be required by specific local electrical, plumbing and other statutory requirements.

Key points to consider:

This copy then will be regarded as an integral part of your order or contract.

Fabrication of the equipment shall be in accordance with this signed, approved Equipment Submittal.

Manufacturing of the equipment will not begin until the signed submittal is returned, and all submittal comments are approved by STERIS.

Any and all notations made by you, the Customer, on this signed copy of the submittal will be reviewed by your STERIS Account Manager to determine any adjustments to the original contract price as well as original quoted lead-time. A revised quote and STERIS Change Acknowledgement Form will be submitted to you in order to confirm changes.

Sign and return one (1) copy of this submittal approval page by 6/3/2022 to your STERIS contact:

Mailyn Mendoza Sales Coordinator STERIS Corporation 5960 Heisley Rd.

Mentor, OH 44060 Ph: 440-392-7407

FX: 440-350-7077

E-mail: Mailyn_Mendoza@steris.com

Please check one of the boxes below:

Approved as is Approved as noted

• Resubmittal NOT required

Rejected Approved as noted

• Resubmittal required

Customer Approval Signature: Date:

Required On-Site Equipment Date:

APPROVAL FORM

Date: May 20, 2022

PROJECT: US DEPT OF AGRICULTURE

PO: 12905B22F0053

STERIS ORDER: 14411505

Comments & Clarifications - *Please utilize the below section to detail any comments, questions, or exceptions noted throughout this submittal package.

TAB #_________

ITEM #_________

COMMENT:

EQUIPMENT DETAIL

1 AMSCO® 110LS Small Steam Sterilizer

SF5113011240000000

16" x 16" x 26" (406 x 406 x 660 mm) Prevacuum Manual, Vertical Sliding, Single Door Cabinet Standard Piping: Brass and Copper Threaded Electric Steam Generator - Carbon Steel 208 Volt, 60 Hz

1 Seismic Kit

FV010018

Rachel.Leisso Underline

Rachel.Leisso Text Box 240 V, 60 Hz instead

SD1026 Rev D

Item ________________________ Location(s)___________________

AMSCO® 110LS AND AMSCO® 250LS

SMALL STERILIZERS – LIFE SCIENCES

APPLICATION

AMSCO 110LS and AMSCO 250LS Small Sterilizers are designed for use in laboratory and industrial applications.

Each is configured with pre-vacuum, liquid, and gravity cycles. A full list of standard and optional cycles begins on page 2.

DESCRIPTION

AMSCO 110LS and AMSCO 250LS Small Sterilizers for life science applications are the next advancement in the STERIS line of steam-jacketed sterilizers. Both sterilizers are equipped with the latest features in state of the art technology and ease of use.

Interior Chamber Nominal Dimensions and Capacity

• AMSCO 110LS – 16 x 16 x 26" (406 x 406 x 660 mm), 110L capacity

• AMSCO 250LS – 20 x 20 x 38" (508 x 508 x 965 mm), 250L capacity

Allen-Bradley MicroLogix™ Control System with enhanced functionality and user-friendly Allen-Bradley PanelView Plus™ 7 700 interface screen.*

• Touch-sensitive screen with 18-bit color graphic display

• Display features 640 x 480 resolution color-active matrix

MODEL

❑ AMSCO 110LS

❑ Prevacuum (LV-110) ❑ Isothermal (LI-110)

❑ AMSCO 250LS

❑ Prevacuum (LV-250) ❑ Isothermal (LI-250)

STEAM SOURCE

❑ Building Steam ❑ Stand Alone Electric Steam Generator

❑ STERIS Provided ❑ Customer Provided ❑ Integral Electric Steam Generator1

❑ Carbon Steel ❑ Stainless Steel Voltage Options

❑ 208 Volt, 60 Hz ❑ 240 Volt, 60 Hz ❑ 400 Volt, 50 Hz ❑ 480 Volt, 60 Hz ❑ 600 Volt, 60 Hz

❑ Integral Indirect Stainless-Steel Clean Steam Generator (SD589)10

❑ SINGLE DOOR

❑ Cabinet Enclosed/Freestanding ❑ Recessed

❑ DOUBLE DOOR

❑ Recessed through One Wall

❑ Sterile side ❑ Non-sterile side ❑ Recessed through Two Walls2

OPTIONS

❑ Power Door(s) ❑ Liquid Air Cool (w/vacuum)

❑ Effluent Decontamination Cycle11

❑ Pure Steam Piping to Chamber11

❑ Air Detector System ❑ Printer on Both Ends8

❑ Bioseal 3,11

❑ RTD Load Probe(s) and F0 Sterilization ❑ One Probe ❑ Two Probes (max. 2)

❑ Drain Line Reference Probe ❑ Air-Differential Seal (NOE) for Double Door Units11

❑ Back Panel for Single Door Cabinet Enclosed Unit ❑ 1" Chamber Penetration, Qty. 1

(AMSCO 110LS only)4

❑ One Additional Chamber Penetration

(AMSCO 250LS)4

❑ Vacuum Pump5

❑ 120 V, 1-Ph / 208 / 240 Vac, 3-Ph ❑ 120 V, 1 Ph / 240 Vac 3-Ph ❑ 120 V, 1-Ph / 480 Vac, 3-Ph ❑ 120 V, 1-Ph / 600 Vac, 3-Ph ❑ 230 V, 1 Ph / 400 Vac, 3-Ph (International) ❑ Other - Specify ___________________

❑ Dry Contacts ❑ Backflow Preventer ❑ Auto Flush for Steam Generator ❑ 0.2 Micron Bacterial Retentive Filter ❑ Green Gravity Water Saver System ❑ STERI-GREEN® Water Conservation System ❑ STERI-GREEN PLUS® Water Conservation System

(Chilled water required)

ACCESSORIES (for AMSCO 110LS)6

❑ One Intermediate Shelf ❑ Air Compressor, Portable, 115 Vac

ACCESSORIES (for AMSCO 250LS) ❑ One Intermediate Shelf ❑ Seismic Tie-Down Kit7,9

❑ Air Compressor, Portable, 115 Vac ❑ Loading Rack and Two Shelves

❑ Single Door ❑ Double Door ❑ Loading Car ❑ Transfer Carriage ❑ Chamber Track Assembly

❑ Single Door ❑ Double Door ❑ Loading Car, Transfer Carriage & Chamber Track

Assembly ❑ Single Door ❑ Double Door

Notes:

1. AMSCO 110LS double door sterilizers are not available with integral electric steam generator.

2. Available for AMSCO 250LS double door sterilizers only. Contact engineering if mounting through two walls is required for AMSCO 110LS.

3. Available on AMSCO 250LS double manual door sterilizers only.

4. One chamber penetration is standard on

AMSCO 250LS.

5. 1-Ph is for control, 3-Ph is for pump motor

6. Loading rack & 2 shelves standard on

AMSCO 110LS.

7. Based on CA requirements.

8. Printer at operating end is standard. Option adds additional printer to non-operating end.

9. Units with Bioseal require special seismic kit.

10. Integral indirect clean steam generator requires field certification for UL/CSA.

11. Compressed air is required for these options.

Selections Checked Below Apply To This Equipment

(Typical - details may vary.)

* Allen-Bradley MicroLogixTM and Allen-Bradley PanelView PlusTM are registered trademarks of Rockwell Automation, Inc.

mamendoz Highlight mamendoz Highlight mamendoz Highlight mamendoz Highlight mamendoz Highlight mamendoz Highlight mamendoz Typewritten Text

SF5113011240000000

mamendoz Highlight mamendoz Highlight

Rachel.Leisso Highlight

Rachel.Leisso Highlight

Rachel.Leisso Highlight 240 V, 60 hz, this is what we need.

• Display is designed with emphasis on human factors and user recognizable symbols.

• Noise level for units with water ejector 68Dba, units with vacuum pump 74Dba.

Cycle Programming and Flash Memory

• 20 cycles may be individually selected and programmed

• Help screens for control operation

• Program permanently stored in flash memory

• Variables permanently stored in flash memory

Vertical Sliding Door(s)

AMSCO LS configurations include choice of single or double door sterilizers and open or recessed mounting. The doors are manually operated. Door seals (1 per door) are non-lubricated, steam activated.

NOTE: Recess two wall mounting is not available for AMSCO 110LS double door sterilizers.

Modularized Vessel And Piping

Vessel and piping are designed for increased dependability and reduced service time.

• Reduced piping components increase reliability

• Vessel design allows higher operating temperature of 141°C (285°F)

• Non-clogging chamber drain line prevents media from plugging drain line

• Emergency manual exhaust valve

STANDARDS

Each sterilizer meets applicable requirements of the following listings and standards, and carries the appropriate symbols.

• Governing Directives for Affixing the CE Mark:

» Low-Voltage Directive (2014/35/EU)

» EMC Directive (2014/30/EU)

» Machinery Directive (2006/42/EEC)

• Pressure Equipment Directive (PED): 97/23/EC.

• UL/EN/CSA 61010-1: Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use

Part 1: General Requirements

• UL/EN/CSA 61010-2-040: Safety Requirements For Electrical Equipment For Measurement, Control, And Laboratory Use, Part 2-040: Particular Requirements For Sterilizers And Washer- Disinfectors Used To Treat Medical Materials

• ASME Code, Section VIII, Division 1 for unfired pressure vessels. The pressure vessel is so stamped; ASME Form U-1 is furnished. The shell and door are constructed to withstand a working pressure of 45 psig (3.1 bar).

CYCLE DESCRIPTIONS

Standard and optional cycles are as follows:

Gravity Cycle (standard) provided on prevacuum, and isothermal sterilizers, for sterilization of heat- and moisture-stable goods at 100°C to 141°C (212°F to 285°F), and decontamination of bagged non bio-hazardous laboratory wastes. Gravity cycle utilizes gravity air-displacement principle. However, the gravity cycle has a standard drying phase which uses a vacuum. The drying phase can be disabled.

Liquid Cycle (standard) provided on prevacuum, and isothermal sterilizers, for sterilization of liquids and media in vented borosilicate glass or metal containers at 100°C to 141°C (212°F to 285°F). Liquid cycle uses optimal solution cooling feature, during exhaust (cooling) phase, to control exhaust rate.

Prevacuum Cycle (standard) provided only on prevacuum sterilizer, for efficient, high-volume sterilization of porous, heat-and moisture-stable materials at 100°C to 141°C (212°F to 285°F). Prevacuum cycle utilizes a mechanical air-evacuation system.

USP 660 Cycle (standard) developed to assist meeting the requirements of USP 660 sterilization test cycle for glassware.

Load probe option is recommended when using this USP cycle.

Continuous Cycle (standard) This cycle allows for up to 9,999 cycles to be run consecutively without the need of an operator.

Cycle parameters can be set, along with the amount of time to lapse in between cycles. For this cycle to function, a power door is required to facilitate automatic opening and closing.

Waste Bag Cycle (standard) This cycle has been preconfigured with cycle parameters shown to be effective when processing lab waste in autoclavable bags. Through air removal assistance and a controlled exhaust, the internal temperature is increased more effectively and faster, while minimizing boil over of any liquids present. Parameters may need to be adjusted based on specific loads. Bags should not be completely sealed.

Isothermal Cycle (optional) provided only on isothermal sterilizer, for processing of heat-sensitive and heat-coagulable solutions in vented borosilicate glass or metal containers at 78°C to 110°C (170°F to 230°F). Isothermal cycle utilizes steam to enhance temperature control and prevent layering of steam and air within the chamber. Process maintains positive pressure in chamber to inhibit media boiling.

NOTE: Temperature control spread is greater on Isothermal units ±6°C.

Liquid Air Cool (Optional) provides water to jacket, and air pressure to chamber to improve exhaust time for liquid loads, and to reduce boil-over. Compressed air is required for this cycle.

ATF 1 and ATF 2 Cycles (Alternating Tangential Flow) are designed to sterilize the XCell™ ATF system. These can be used for other purposes that require rates and hold times in the pre-conditioning pulses and a cool down in exhaust by temperature.

Effluent Decontamination Cycle (Optional) is used for processing contaminated laboratory waste (BL-3 and BL-4). Condensate produced during processing cycle is decontaminated before discharge to floor drain. Steam is admitted through bottom of sterilizer chamber, and chamber is exhausted out top side of vessel. During purge and vacuum pulses, all purge and exhaust gases are vented through a

0.2 micron bacterial retentive filter. Filter housing is steam jacketed to prevent wetting of filter membrane. Available with fast exhaust or optimal solution cooling (slow exhaust) exhaust types. User is responsible for development of process parameters. Compressed air is required for this cycle.

Leak Test Cycle, for verification of door seal and piping system integrity. Cycle parameters are pre-programmed and fixed.

Acceptable maximum leak rate is 1.0 mm Hg/minute over a 10-minute period following a fixed stabilization time.

Daily Air Removal Test (Dart) Cycle, provided only on prevacuum sterilizer, for verification of effective removal of residual air in chamber and load during testing. Test cycle determines if even and rapid steam penetration into test load occurred. Cycle parameters are preprogrammed and fixed.

Bowie-Dick Test is available for 121°C (250°F) and 132°C (270°F) prevacuum cycles.

Healthcare Cycles - The LS Series sterilizers can be programmed to reproduce cycles that are used for healthcare applications, however, these sterilizers do not have FDA clearance and cannot be used for patient treatment applications

ENVIRONMENTAL SUSTAINABILITY

Automatic Utilities Startup/Shutdown

This feature permits slow cooling of the entire vessel and load.

Shutdown may be programmed to activate at the end of any designated cycle or time of day. When activated, control system automatically shuts off all utility valves, conserving steam and water usage. Sterilizer utilities can be restarted either by programmed time or manual operation. A different shutdown and restart time can be programmed for each day.

Green Mode

The Green Mode is a standard feature on the control that will shut off the steam to the jacket after the unit has sat idle for a specified period of time. The specific time frame is determined by the user and entered into the control during set up; it can be changed at any time.

Green Gravity Water Saver System

The Green Gravity Water Saver System provides additional water savings by collecting steam effluent and holding it in a cooling tank, reducing the amount of water required to cool the effluent.

STERI-GREEN® Water Conservation System

This system significantly reduces the consumption of potable water. The STERI-GREEN system utilizes a mixing tank and an air-cooled heat exchanger to cool and recycle vacuum pump water and steam effluent. Water temperature is constantly monitored to minimize the need to add fresh cool water to the mixing tank. The system can provide up to 79% water savings over the vacuum pump alone, or up to 87% over a water ejector.

When ordered, system includes a vacuum pump (3-phase power required).

STERI-GREEN PLUS® Water Conservation System

The STERI-GREEN PLUS system utilizes facility chilled water supply. The system utilizes a mixing tank and a series of heat exchangers, integrated with the a chilled water loop, to cool and recycle vacuum pump water and effluent. Water temperature is constantly monitored to minimize the need to add fresh cool water to the mixing tank. This system provides up to 99% water savings over water ejector or vacuum pump upgrade alone. When ordered, system includes a vacuum pump (3-phase power required).

STANDARD FEATURES

Hinged front cabinet panel opens for convenient access to sterilizer piping and control system.

Resistance Temperature Detectors (RTDs) are installed for sterilizer temperature control. The chamber drain line RTD senses and controls temperature variations within the sterilizer chamber. A jacket RTD provides temperature control within the jacket space. These RTD signals, converted into electrical impulses, provide accurate control inputs and readouts throughout entire cycle.

Software calibration is provided for all temperature and pressure inputs. Calibration is performed in the service mode, accessible through the touch screen displays, and accomplished using external or internal temperature and pressure sources. Control system provides a printed record of all calibration data for verification to current readings.

Cycle Data Records, recorded on the printer tape. They can also be retrieved for on-screen review, through the USB port located on the front fascia panel, or sent via e-mail if the system is enabled. The e-mail feature requires a network connection.

Optimal solution cooling is designed to safely cool various liquids in vented, borosilicate glass containers with minimum liquid loss due to boil-over, and to keep normal evaporation loss below 5%. Optimal solution cooling is an integral part of the factory-programmed liquid cycle. During the exhaust (cooling) phase, the control utilizes this feature to optimize the exhaust rate regardless of load size or container fill volume, up to one liter. During cooling, the initial rate, initial rate transition point, as well as the second rate and final vacuum level are adjustable.

Steam purge feature is provided to assist in air removal and preheat the load.

Automatic steam shutoff to jacket is provided for isothermal and liquid cycles. When activated for isothermal cycles, the jacket control conducts a timed jacket drain, automatically allowing for the operation of cycles at lower temperatures.

When activated for liquid cycles, steam supply to the jacket is turned off during exhaust phase, allowing load to cool more efficiently.

Insulation, one-inch thick, asbestos-free spin-glass (rated at 1000 °F [538 °C] continuous) encompasses the exterior of the sterilizer vessel and is sealed in an oil and water resistant outer jacket.

Lighted DIN connectors are installed on all steam, water, and exhaust valves for reliability and ease of maintenance.

Air Backup to seals is provided on all double door sterilizers, with either bioseal or air differential seals.

Visible Gauges are standard. Chamber and jacket pressure gauges are standard on the operating end and non-operating end.

CONTROL SYSTEM

Design Features

Together, the Allen-Bradley MicroLogix™ PLC control and PanelView Plus™ 7 700 display, monitor and control all sterilizer operations and functions. Control system is factory-programmed with standard sterilizing cycles, each adjustable to meet specific processing requirements. All control configuring is performed through touch screen displays.

Cycle values and operating features may be adjusted and verified prior to cycle operation. Once cycle is started, cycles and cycle values cannot be changed until cycle is complete.

On completion of cycle, timers reset to the previously selected values, eliminating the need to reset values between repeated cycles. If chamber temperature drops below the under temperature setting during the exposure phase, timer can be set to stop and automatically reset or resume once normal operating temperature is reached.

Critical control system components are housed within a sealed compartment to protect components from moisture and heat generated during sterilization process.

The control system is provided with features such as audit trail, password management and electronic signatures, which can facilitate compliance to 21 CFR Part 11 (Code of Federal Regulations Title 21)

Operator interface control panel, consisting of touch screen and thermal printer, is located on operating (load or nonsterile) end of sterilizer. If the sterilizer is equipped with double doors, an additional touch screen is provided on the non-operating (unload or sterile) end.

• Touch-Sensitive Screen features a color active matrix 18-bit graphics display. All sterilizer functions, including cycle initiation and cycle configuration, are operated by pressing the touch-sensitive areas on the display, referred to as buttons. Display indicates appropriate control buttons, operator prompts and status messages necessary to assist in sterilizer operation. All displayed messages are complete phrases with no codes to be cross-referenced. Display also indicates any abnormal conditions that may exist either in or out of a cycle.

• Thermal printer is a high resolution (8 dots per mm) printer.

It is fast and quiet, printing at 25 lines per minute on industrial grade thermal paper. The device provides an easy-to-read printed record of all pertinent cycle data on 2.25" (57 mm) wide paper. Data is automatically printed at the beginning and end of each cycle and at transition points during the cycle.Three paper tape rolls are furnished with each unit.

Printouts have a guaranteed lifetime of 25 years and are resistant from exposure to steam, alcohol, UV and visible light, oil, heat and water.

• Non-operating end (NOE) control panel (equipped on double-door sterilizers only) includes a touch-sensitive screen similar to the operating end screen. Preprogrammed cycles can be started from the NOE control panel. Display concurrently shows the same information as the operating end screen display.

• Cycle configuration is performed by accessing the change values menu on either end of a double door sterilizer. Utility shutdown parameters can only be changed on the OE display. In addition to adjustment of cycle values, the following operating parameters can also be changed through the change values menu:

» Time Display and Printout Units in standard AM/PM or 24-hour military (MIL) time.

» Selectable Cycle Name permits user to name each cycle with any combination of letters, numbers, blank spaces, and underscores, up to eight characters long.

» Print Interval permits adjustment of the time period between cycle-status printouts generated during the sterilize phase.

» Security access code is required to enter the operating mode (running cycles), supervisor mode (changing val-ues), and service mode. Operating the sterilizer or accessing change values menu causes display to request the entry of an access code. If access code is not properly entered, display returns to the standby screen, denying user access to the sterilizer or program-ming. Access to the sterilizer can be limited to 12 operators, each with a different access code.

» Alarm tones – The end of cycle and alarm tones are adjustable from low, medium and high. The end of cycle tone can also be set off.

» Temperature Display and Printout Units in Celsius (°C) or Fahrenheit (°F). Temperature is set, displayed, controlled, and printed to the nearest 0.1°. Recalibration is not required when changing temperature units from °C to °F and vice versa.

» Pressure/Vacuum Display and Printout Units in psi/In/Hg, Bar (Gauge and Absolute). Recalibration is not required when changing pressure units.

Mobile and PC Messaging (Standard)

This is a standard feature on the control that allows the user to receive text messages or emails alerts regarding the status of the autoclave. The operator can choose from a list of # possible alerts from which to be notified.

SCADA Ready Control Interface (Standard)

If requested, STERIS can provide the user the information to allow communication between the autoclave and the Building Management System. An ethernet connection is required.

Printer Data Storage (Standard)

All printer data is saved internally and can be retrieved and viewed on the display screen, exported to a USB drive (port located on the front fascia panel) or emailed to the User.

Cycle usage (Standard)

Utilization of the sterilizer can be monitored for each User, data can be exported via an Ethernet or USB port located on the front fascia panel.

SAFETY FEATURES

Control lockout switch (equipped on chamber door) senses when door seal is energized and tight against the door. Control prevents cycle from starting until the limit switch signal is received. If control loses appropriate signal during cycle, alarm activates, cycle aborts, and chamber safely vents with a controlled exhaust.

Typical Cycle Menu Display

Chamber float switch activates alarm, aborts cycle, and safely vents chamber with a controlled exhaust if excessive condensate is detected in the vessel chamber.

Door interlocks (double door units only) allow only one door to be opened at a time and, during processing, prevent the non-operating end (NOE) door from being opened until a satisfactory cycle is complete. If a cycle is aborted, the NOE door cannot be opened. The use of this feature may affect the door gaskets life expectancy unless an air differential or bio seal is provided.

Pressure relief valve limits the amount of pressure buildup so that the rated pressure in the vessel is not exceeded.

Emergency stop button (located on front of the sterilizer) is included on all sterilizers. A key is required to release the emergency stop (once pushed) before the unit can return to normal operation.

CONSTRUCTION

Shell Assembly

Two fabricated Type 316L stainless-steel shells, welded one within the other, form the sterilizer vessel. Type 316L stainless-steel end frame(s) is welded to door end. On single door units, back of chamber is fitted with welded, 316L stainless-steel formed head.

Sterilizer vessel is ASME rated at 50 psig (3.2 Bar) and insulated. The Vessel (for AMSCO 250LS only) includes one 1.0" (25 mm) NPT welded chamber bushing for Customer use.

Steam-supply opening inside the chamber is shielded by a Type 316L stainless-steel baffle.

Chamber Door(s)

Door is constructed of a single formed piece of Type 316L stainless steel. Door is insulated to reduce the surface temperature of the stainless-steel door cover.

During cycle operation, door is sealed by a steam-activated door seal. Door seal is constructed of a special long-life rubber compound. When sterilizer cycle is complete, the seal retracts under vacuum into a machined groove in sterilizer end frame.

Door seal can be manually retracted to open the door and remove critical load in an emergency situation (if loss of vacuum or loss of power occurs).

A handle is used to manually open (by lowering) and manually close (by raising) the door which is suspended by cables attached to a counterweight.

A long-life proximity switch is used, by the control, to determine if the door is closed. An additional seal pressure switch prevents inadvertent cycle initiation if door is not sealed.

The door assembly is equipped with a mechanical locking mechanism that ensures the door cannot be opened, as long as the seal is intact and energized, and more than 2.0 psi pressure is in the chamber. Door interlocks on double door sterilizers can be programmed to prevent inadvertent opening of door(s). Access code is required to override door interlocks.

NOTE: Bio-Seal option is available for AMSCO 250LS, double manual door sterilizers, as discussed on page 5.

Chamber Drain System

Drain system is designed to prevent pollutants from entering into the water-supply system and sterilizer. An optional backflow preventer is available. Water supply shutoff valve is located behind the front cabinet service panel under the chamber.

Automatic Drain Effluent Cooling

The piping system to the drain provides automatic condensing of chamber steam and disposal of condensate to waste.

Cooling water is added to ensure discharge temperature is discharged at or below 60°C (140°F). A separate resistance temperature detector (RTD) is included to limit the volume of water to only the amount required to achieve target temperature, thus conserving water.

Vacuum System

Chamber pressure is reduced during the conditioning phase and drying phase through the means of either a standard water ejector or an optional vacuum pump upgrade. Subsequent to the drying phase, the chamber is returned to atmospheric pressure by admitting air through a bacteria-retentive filter.

Steam Source

Sterilizers are piped valved, and trapped to receive building-supplied steam delivered at 50 to 80 psig (344.7 to 551.6 kPa) dynamic. If building steam source is not available, an electric carbon-steel steam generator or electric stainless-steel steam generator may be provided to supply steam to the sterilizer.

Steam piping is constructed of brass and includes a shutoff valve, steam strainer and a brass pressure regulator.

Optional stainless-steel indirect-type steam generator can be installed as a pure steam source. Pure steam reduces the probability of contamination which could adversely affect research, such as tissue culture and trace metals studies.

Steam feeds from the jacket to the chamber. A check valve is added between the jacket and chamber on sterilizers with decontamination cycle option.

Piping

All piping connections terminate within the confines of the sterilizer and are accessible from the front and side of sterilizer.

• Solenoid Valves in manifold with DIN connectors simplify sterilizer piping and can be serviced individually.

• Manual Shutoff Valves are pressure rated at 125 psig (862 kPa) for saturated steam. Valve handles are low-heat conducting.

MOUNTING ARRANGEMENT

Sterilizers are arranged for either freestanding or recessed installation, as specified. Each sterilizer is equipped with a height-adjustable, steel floor stand. Sterilizer subframe is equipped with a synthetic rubber gasket to ensure tight fit between the cabinet panels on freestanding units or between the front cabinet panel and wall partition on recessed units.

On freestanding units, stainless-steel side panels enclose the sterilizer body and piping.

Each AMSCO 110LS and AMSCO 250LS sterilizer has a lockable front service panel.

OPTIONS

Stainless-steel piping to chamber delivers steam generated from Customer purified water source to the chamber and its contents. All steam-to-chamber piping components are constructed of 300 series stainless steel. Option is provided with Pressure Reducing Valve. Compressed air required for pneumatic valves.

Integral indirect stainless-steel clean steam generator automatically produces clean steam using Customer-supplied steam and purified water. Generator is integrally connected to the clean steam-to-chamber piping system.

30 kW carbon-steel electric steam generator typically fed by a potable water source with hardness not to exceed 171 mg/L.

The generator is available for both single and double door sterilizers. The generator is mounted underneath both single and double door units. The generator option is not available for AMSCO 110LS double door units.

30 kW electric stainless-steel steam generator is electrically powered, automatically filled with water having 1 M·cm resistivity, and operates whenever the sterilizer power is on.

Generator is integrally connected to the clean steam-to-chamber piping system. The generator option is not available for AMSCO 110LS double door units.

Auto Flush for Steam Generator provides automatic flush of steam generator upon startup of sterilizer. Not required for SS generators.

RTD load probes and F0 sterilization (maximum of 2) automatically sense the load temperature during cycle operation. A single thermal load probe is sealed through the sterilizer vessel and manually placed in the product container within the chamber prior to cycle operation.

In conjunction with the load probe option, individual cycles can be set to start exposure phase according to chamber drain temperature or according to load temperature. Also, F0 set points are available for each cycle, allowing for exposure phase termination based on the calculated F0 value.

Bioseal (for AMSCO 250LS double manual door units only) is a 1/4” steel plate which is welded to the chamber and a 1/4” thick silicone gasket that extends between the plate and a carbon steel wall frame which is welded to wall imbeds.The bioseal is provided on the non-operating end of the sterilizer, prevents passage of airborne microorganisms from the space between the vessel body and the structural wall opening.Steam is the primary source of pressure behind the door seal. All sterilizers with bioseals have air back-up to maintain seal pressure when out of cycle or if the steam source is not available.

Air-differential seal (double door units only), provided on the non-operating end of the sterilizer, minimizes airflow between the dirty and clean sides of the barrier.

Back cabinet panel is provided on single door, freestanding units where the unit is accessible on all sides.

Air detector (integral factory piping option) is used to determine whether any air or non-condensible gas present in the chamber is sufficient to impair the sterilizing process.

Backflow preventer option can be installed on sterilizer piping to prevent the unwanted reverse flow of water or other substances into the potable water supply.

Power door provides automatic opening and closing by depressing a foot pedal. (Not available for sterilizers with Bio-Seal option).

Vacuum Pump Upgrade: Water ejector is replaced with a water ring vacuum pump.

Drain line reference probe automatically senses the drain line temperature during cycle operation.During the Sterilize phase, the chamber and reference probes are compared and if the difference is outside the allowable range, an alarm will occur.

This option replaces the standard drain RTD probe with a dual element RTD probe, in the same sheathing.

0.2 micron bacterial retentive filter provides sterile air during airbreak at end of cycle.

Additional chamber penetration: One 1" NPT capped chamber penetration port is located at the side of the vessel so as not interfere with other piping. The port provides for up to twelve (12) owner provided thermocouple probes or other test instrumentation.

One port is provided as standard on AMSCO 250LS.

Printer on both ends. An additional printer is provided on the non-operating end of the sterilizer.

Dry Contacts provide four (4) relays to communicate the following equipment status: door open, door closed, alarm, and unit on.

ACCESSORIES

Air Compressor, Portable, 115 Vac. This accessory is intended for pneumatic valves on sterilizers when an air utility is not provided by the facility. It may also be used for back-up pressure source for the door seal in bioseal applications.

This is a portable 1.5 Gallon compressor tank that delivers 48 LPM @ 689 KPa (1.7CFM @ 100 PSI). Refer to STERIS drawing no. 755718-038 for complete specifications.

NOTE: UL/CSA certified only.

Seismic tie-down kit conforms to current California Code of Regulations.

PREVENTIVE MAINTENANCE

A global network of skilled service specialists can provide periodic inspections and adjustments to help ensure low-cost, peak performance. STERIS representatives can provide information regarding annual maintenance agreements.

NOTES

1. The sterilizer is not supplied with a vacuum breaker or backflow preventer and, where required by local codes, installation of such a device in water line is not provided by

STERIS.

2. Pipe sizes shown indicate terminal outlets only. Building service lines, not provided by STERIS, must supply the specified pressures and flow rates.

3. Disconnect switches (with OFF position lockout only; not provided by STERIS) should be installed in electric supply lines near the equipment.

4. Access to the recessing area from the control end of the sterilizer is recommended.

5. Clearances shown are minimal for installing and servicing the equipment.

6. If loading car and carriage are to be used with a AMSCO 250LS sterilizer, front clearance should equal twice the length of the sterilizer. This will permit complete withdrawal of the loading car from the chamber and allow convenient maneuverability of the transfer assembly to and from the sterilizer.

7. Floor drain should be provided within confines of sterilizer framework.

UTILITY REQUIREMENTS

Refer to Equipment Drawing for complete information.

Sterilizer Using Facility Steam1

Steam

1/2" NPT, 50 to 80 psig (344.7 to 551.6 kPa) dynamic, 97 to 100% vapor quality

Drain

1-1/2" ODT drain terminal. (Floor drain capacity must handle peak water consumption; refer to engineering data.)

Electrical - Controls

120 Volt, 50/60 Hz, 1-phase, 3.0 Amp

International:

230 Volt, 50/60 Hz, 1-Ph, 1.5 Amp

Sterilizer Feed Water

1.0" NPT, 30 to 50 psig (206.8 to 344.7 kPa) dynamic. Water is used for ejectors, vacuum pumps, exhaust condensers, and trap cooling. Refer to Table 1 for recommended water quality. Use of feed water within the nominal conditions will optimize equipment performance and reduce maintenance.

NOTE: Backflow prevention is not standard on the unit, but a backflow preventer option can be ordered.

Sterilizer Equipped With Integral Carbon Steel Steam Generator

Drain

1-1/2" ODT drain terminal. (Floor drain capacity must handle peak water consumption; refer to engineering data.)

Generator Drain

1/2" ODT

Electrical - Controls

120 Volt, 50/60 Hz, 1-phase, 9.5 Amp

International:

230 Volt, 50/60 Hz, 1-Ph, 1.5 Amp

Electrical - Generator

208 Volt, 50/60 Hz, 3-phase

240 Volt, 50/60 Hz, 3-phase

480 Volt, 50/60 Hz, 3-phase

600 Volt, 60 Hz, 3-phase

International:

380/415 Volt, 50/60 Hz, 3-Ph, (Prevacuum Units)

Sterilizer Feed Water

1.0" NPT, 30 to 50 psig (206.8 to 344.7 kPa) dynamic.

Refer to Table 1 for water specification guidelines.

Steam Generator Feed Water

1/2" NPT, 20 to 50 psig (137.9 to 344.7 kPa) dynamic. Refer to Table 2 for required water quality. Use of feed water within the nominal conditions will optimize equipment performance and reduce maintenance.

NOTE: Backflow prevention is not standard on the unit, but a backflow preventer option can be ordered.

Sterilizer Equipped With Integral Stainless-steel Steam Generator

Sterilizer Feed Water

1.0" NPT, 30 to 50 psig (206.8 to 344.7 kPa) dynamic.

Refer to Table 1 for required water quality.

Steam Generator Feed Water

1/2" NPT, 20 to 50 psig (137.9 to 344.7 kPa) Refer to Table 3 for required water quality.

Drain

1-1/2" ODT generator drain terminal. (Floor drain capacity must handle peak water consumption; refer to engineering data).

Generator Drain

1/2" ODT

Electrical - Controls

120 Volt, 50/60 Hz, 1-phase, 9.5 Amp

International:

230 Volt, 50/60 Hz, 1-Ph, 1.5 Amp

Electrical - Generator

208 Volt, 50/60 Hz, 3-phase

240 Volt, 50/60 Hz, 3-phase

480 Volt, 50/60 Hz, 3-phase

600 Volt, 60 Hz, 3-phase

International:

380/415 Volt, 50/60 Hz, 3-Ph, (Prevacuum Units)

Refer to the Following Equipment Drawings for Installation Details

Equipment Drawing Number Equipment Drawing Title

AMSCO 110LS

387362-169 110LS, SINGLE DOOR, RECESSED, STEAM HEAT

387362-170 110LS, SINGLE DOOR, CABINET, STEAM HEAT

387362-171 110LS, SINGLE DOOR, RECESSED, ELECTRIC STEAM

387362-172 110LS, SINGLE DOOR, CABINET, ELECTRIC STEAM

387362-173 110LS, DOUBLE DOOR, RECESSED 1 WALL, STEAM HEAT

AMSCO 250LS

387362-159 250LS, SINGLE DOOR, RECESSED, STEAM HEAT

387362-160 250LS, SINGLE DOOR, CABINET, STEAM HEAT

387362-161 250LS, SINGLE DOOR, RECESSED, ELECTRIC STEAM

387362-162 250LS SINGLE DOOR, CABINET, ELECTRIC STEAM

387362-163 250LS, DOUBLE DOOR, RECESSED 1 WALL, STEAM HEAT

387362-164 250LS, DOUBLE DOOR, RECESSED 2 WALLS, STEAM HEAT

387362-165 250LS, DOUBLE DOOR, RECESSED 1 WALL, ELECTRIC STEAM

387362-166 250LS, DOUBLE DOOR, RECESSED 2 WALLS, ELECTRIC STEAM

1. External Supplied Steam (Facility Steam/Stand-Alone Steam Generator)

Table 1. Recommended Feed Water Quality for Sterilizers

Condition Nominal Conditions Maximum Conditions

Temperat…