Attachment G - OBAA Topic Solicitation - Criteria Questionnaire.pdf
PDF 288 KB Posted
- Attached to
- THOR Presolicitation Federal contract opportunity
- Solicitation number
- RWRD-21-0026
About this file
This document contains a desired capabilities questionnaire for a gamma detection device. The questionnaire lists 53 capabilities sought by the government customer at both threshold and objective levels. For each capability, offerors are asked to specify how their proposed solution will meet the criteria, including noting whether it will meet threshold or objective levels and whether demonstration will occur at PDR or CDR. Notable capabilities include gamma detection range, exposure rate measurement range and accuracy, safety alarms, identification of various shielded and unshielded nuclear materials, ability to identify multiple isotopes simultaneously, energy resolution, data storage and transmission capabilities, temperature operating range, and size and weight restrictions. Offerors are asked to fully justify any capabilities they indicate as alternative, partially meeting, or future capability.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| WP1_TlBr_Stability_FINAL.pdf | ||
| WP6_TlBr_References_FINAL.pdf | ||
| WP4B_TlBr_IP_CFG_FINAL.pdf | ||
| WP3_TlBr_Temp_Dependence_FINAL.pdf | ||
| WP4A_TlBr_IP_Pixelated_FINAL.pdf | ||
| Attachment F - THOR CTM.pdf | ||
| OBAA Topic Presolicitation (THOR).pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment G 70RWMD21R0000000X
1 | P a g e
Attachment G: Desired Capabilities Questionnaire
Instructions
The following questionnaire is designed to assist the Government in understanding how an Offerors proposed approach will incorporate the list of currently desired capabilities. Offerors are encouraged to evaluate their proposals accurately, however the Government reserves the right to make final determination of their assessment of the Offeror’s proposed system against the criteria based on the information provided, or not provided, in their proposal. For each capability, the Offeror is requested to also provide either a justification, to allow the Government a better understating of what capability the Offeror is proposing rather than just a simple response to the desired capabilities. Justifications may reference other portions of the proposal to allow for overall brevity.
For each desired capability, the Offeror is asked to provide one of the following responses:
• ‘Meet Threshold’ – Means that the proposed system will provide capability either meeting, or exceeding, the capability requested by the Government at the threshold level. Specify if this will be demonstrated by PDR or CDR.
• ‘Meet Objective’ – Means that the proposed system will provide capability either meeting, or exceeding, the capability requested by the Government at the Objective level. If a system will meet both the threshold and objective this should be stated by ‘Meet Objective’ designation. Specify if this will be demonstrated by PDR or CDR.
• ‘Partially Meet’ – Means that the proposed system will provide the desired capability in a partial manner, such that it meets most of the threshold level capability but not all the requested attributes. Provide an explanation of what part is met and if it will be met by PDR or CDR.
• ‘Future Capability/Does Not Meet’ – Means that the proposed system will not provide the capability during PDR or CDR as currently requested. A description of any capability in that capacity, or the limitation of providing that capability, should be provided in the justification section.
• ‘Alternative Capability’ – Means that the Offeror believes that they have a design which will provide service equal, or better, then the desired capability requested by the Government. Note that this statement will need strong justification within the proposal.
• ‘Not Applicable’ – Means that the Offerors proposal will not have the specified component or other item per the capability such that the requested capability is not applicable to their system. Provide an explanation on why the capability is not applicable.
2 | P a g e
Capability Worksheet.
# Descriptor Criteria Threshold Objective Offeror Evaluation Offeror Justification
NA Example Example (not a THOR Criteria)
Design will enable 8 hours of continuous use
Design will enable 12 hours of continuous use
Objective at CDR, Threshold at
PDR
Based on envisioned power design (see paragraph x in the proposal) the calculations support the threshold.
Additional optimization may enable achieving the objective at CDR.
CR-
Gamma Detection
Device, during scanning operations, shall alarm/provide an indication when the radiation field changes from ambient background to that with a non-background source of radiation present.
See Notes
Probability of detection (PD) ≥80% at a relative speed of
0.3 meters/second and distance of closest approach of 0.61 meters of 0.66 µR/hr of Co-57 and separately 7.53 µR/hr of Ba-133.
Probability of detection (PD) ≥80% at a relative speed of
0.61 meters/second and distance of closest approach of
1.22 meters of 0.66 µR/hr of Co-57 and separately 7.53 µR/hr of Ba-133.
CR-
Modify Threshold
The device shall allow modification of the detection alarm settings/thresholds.
User adjustable thresholds Objective = Threshold
Gamma Energy Range
Device shall detect gamma rays over a relevant range of energies.
25 keV-3000 keV 15 kev-8000 keV
CR-
Exposure Rate Range
The device shall measure exposure rates over a wide range.
0.1 uR/hr-50 mR/hr 0.1 uR/hr - 1 R/hr
CR-
Exposure Rate Accuracy
The device shall accurately display exposure rate or ambient dose equivalent rate readings.
Meet the accuracy requirements +/-10% for Cs-
Meet the accuracy requirements +/-30% for Am- 241 and Co-60 and +/-10% for Cs-137
Safety Alarm An alarm shall be provided to alert the user that the ambient radiation intensity is above a user-selected threshold level (typically 2mR/hr).
Distinct audible and visible alarm when the radiation field exceeds alarm setting.
Distinct audible, visible, and tactile alarm when the radiation field exceeds alarm setting.
Settable Safety Level
The device shall provide a user-definable safety alarm setting.
Settable threshold based on exposure rate/dose rate.
Objective =Threshold
3 | P a g e
Gamma False Alarm
Device shall have a low rate of gamma false detections.
≤1 false gamma alarm in 8 hours when operated in a stable background environment.
≤1 false gamma alarm in 12 hours when operated in a stable background environment.
Localize Device shall support the operator in localizing a radiation source.
Indicate rise and fall of counts using a time series plot with a maximum of 1 second time intervals.
In addition to the threshold provide directionality to a gamma source.
Radionuclide Library
The device shall have a library of radionuclides/reactions that can be identified
The Library includes radionuclides/reactions listed as Priority 1, 2 and 3.
Library includes radionuclides/reactions listed as Priority 1, 2, 3, 4 and 5.
PID
Unshielded
The device shall correctly identify with a Probability of Identification (PID) ≥ 80% the present radionuclides that are unshielded and produce 50 μR/hr above background at the reference point of the gamma detector, with the exception of NORM (Ra-
226. Th-232, K-40) that are to be 20 μR/hr above background.
Radionuclides/reactions listed as Priority 1, 2 and 3.
Radionuclides/reactions listed
PID shielded The device shall correctly identify, with a Probability of Identification (PID) ≥ 80%, the present radionuclides that are shielded* and produce 50 μR/hr above background at the reference point of the gamma detector, with the exception of NORM (Ra-226. Th-232, K-40) that are to be 20 μR/hr above background.
Radionuclides/reactions listed as Priority 1 and 2.
Radionuclides/reactions listed
4 | P a g e
PID HEU Device shall correctly identify, with a Probability of Identification (PID) ≥ 80%, 70 grams of High Enriched Uranium (HEU) metal at a distance of 25 cm in the following configurations:
i. Unshielded
ii. Shielded by 6.0 cm of 304 stainless steel (8.0g/cc)
iii. Shielded by 15cm of graphite (2.25g/cc)
Objective = Threshold
CR-
PID WGPu Device shall correctly identify, with a Probability of Identification (PID) ≥ 80%, 70 grams of Weapons Grade Plutonium (WGPu) metal at a distance of 25 cm in the following configurations:
i. Unshielded
ii. Shielded by 1.0 cm of tungsten (19.25g/cc)
iii. Shielded by 6.0 cm of 304 stainless steel (8.0g/cc)
iv. Shielded by 15cm of graphite (2.25g/cc)
Objective = Threshold
CR-
PID DU The device shall correctly identify, with a Probability of Identification (PID) ≥ 80%, 3.9 Kilograms of Depleted Uranium (DU) metal at a distance of 25 cm in the following configurations:
i. Unshielded
ii. Shielded by 1.0 cm of tungsten (19.25g/cc)
Objective = Threshold
CR-
PID Multiple Isotopes
The device shall be able to identify multiple radionuclides simultaneously with a probability of correct identification ≥ 80% when each source is producing 50 μR/hr above background at the reference point of the gamma detector.
All cases of two radionuclides. The following are the minimum to be demonstrated for compliance.
a)Cs-137 + DU
b) Tc-99m+ HEU
c) Tl-201+ HEU
d) Ga-67+ HEU
e) I-131+ WGPu
f) NORM + HEU
g) NORM + WGPu
Threshold plus three or more non-NORM radionuclides. The following are the minimum to be demonstrated for compliance.
a) WGPu + Co-60 + Ba-133
b) HEU + Ir-192 + Co-60
c) Cs-137 + Co-60 + Ba-133
Unidentified An indication shall be displayed (e.g., “not identified”) if the device indicates the presence of a source but is not able to identify a radionuclide.
Display not identified for a radionuclide having an exposure rate 50 uR/hr above background after 2 minutes exposure that is not in the device library; where the radionuclide may be removed from the radionuclide library for the test.
Objective = Threshold
5 | P a g e
ID Within 2 Min
The device shall identify present radionuclides with a minimal spectrum acquisition duration.
≤ 2 minutes ≤ 30 seconds
CR-
ID Confidence Device shall provide indication as to the confidence of the radionuclides identified.
Display of a numerical indication (e.g., 1 to 10, 0 to 1)
Objective = Threshold
CR-
Energy Resolution
Device shall have high resolution.
Better than 4% energy resolution as measured on the Cs-137 662 keV peak.
Better than 2% energy resolution as measured on the Cs-137 662 keV peak.
Store Files The device shall have the ability to internally store complete identification data files
Store 100 complete data files.
Store 1000 complete data.
CR-
Transmit Files The device shall transmit complete identification data files in the ANSI N42.42 data format.
Threshold = Criteria Objective = Threshold
CR-
Data Record Info
Data record shall contain pertinent information.
a) RIID manufacturer name, model, and serial number
b) Software version
c) Gamma detector kind (e.g., NaI, polyvinyl toluene [PVT])
d) Date and time of measurement
e) Measured gamma-ray radiation level (e.g., exposure rate)
f) Background spectrum
g) Live time and real time for background spectrum
h) Measured spectrum
i) Live time and real time for measured spectrum
j) Energy calibration for the background and measured spectrum
k) Radionuclide identification results
l) Confidence indication
Objective = Threshold
CR-
Transfer to Ancillary Device
The device shall transfer data to ancillary devices.
1) Computers using Windows 10 or higher.
2) Mobile devices iOS 9.3 and Android OS 9.0 and higher
6 | P a g e
Transmit Method
The device shall be able to transmit data using all of the listed interfaces.
1) USB 2.0 or higher
2) Wi-Fi 5 (802.11ac)
3) Bluetooth security mode 3
4) Removable media
5) Cellular
6) Bluetooth security mode 4
Display The display shall provide operationally relevant indications and information
Display the following, when appropriate:
a) Gamma alarm
b) Neutron alarm, if applicable
c) Radionuclide cannot be identified (e.g., “not identified,” “unknown”)
d) Gamma and neutron count rate readings
e) Count rate is too high for radionuclide identification
f) Radiological over-range conditions (e.g., “over-range” or “high counts”)
g) Radionuclide categorization, identification, and confidence indicator
h) Access mode (e.g., routine or restricted)
i) Battery status
j) Operational status (e.g., normal, calibration needed, stabilization needed)
k) Radiation protection alarm
l) High or low detector count rate conditions indicative of a detector failure
m) Energy stabilization or calibration invalid or not acceptable
n) Gamma exposure rate or dose rate
o) Gamma count rate -time series plot (localization)
Objective = Threshold
CR-
Size Device shall have small physical dimensions.
≤ 33 cm x 23 cm x 16 cm ≤ 16 cm x 10 cm x 8 cm
CR-
Weight Device (including battery etc.) shall have an operational weight of 5 lbs or less.
≤ 2.27 Kg (5 pounds) ≤ 1.6 Kg (3.5 pounds)
7 | P a g e
Temperature Ramp
Device shall operate without degradation in the operational environment in extreme temperature.
-30° C to +50° C -40° C to 50° C
CR-
Self Calibrate Device shall have the ability to self-calibrate.
See Notes
Using a radioactive source that does not exceed the 10CFR3071 schedule B exempt quantity and does not have special shipping constraints.
Does not use a radioactive source (e.g. uses background