Surgical Lights
| Item | Description | Fargo-OR 1-4 | Fargo-OR 5 | St. Cloud-OR 1-2 | St. Cloud-OR 3 | Central Iowa-OR 2 | Central Iowa-OR 3-5 | Black Hills-OR 1-3 | Iowa City | Total | ||
| 1 | LIGHT SOURCE, TYPE | No minimum requirement | No minimum requirement | LED | LED | LED | LED | |||||
| 1a | Number of lightheads | 8 | 3 | 6 | 3 | 1 | 9 | 6 | 20 | 56 | ||
| 1b | Number of bulbs or elements | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | Greater than or equal to 50 LEDs per light | |||
| 1c | Volts | 110/120 | 110/120 | 110/120 | 110/120 | 110/120 | 110/120 | 110/120 | ||||
| 1d | Life, hr | Greater than 30,000 LED | Greater than 30,000 LED | Greater than 30,000 LED | Greater than 30,000 LED | Greater than 30,000 LED | Greater than 30,000 LED | Greater than 30,000 LED | Greater than 30,000 LED | |||
| 2 | COLOR TEMPERATURE, K | Adjustable from 3500-5000 | Adjustable from 3500-5000 | Adjustable from 3500-5000 | Adjustable from 3500-5000 | Adjustable from 3500-5000 | Adjustable from 3500-5000 | Adjustable from 3500-5000 | Adjustable from 3500-5000 | |||
| 2a | Adjustable | Required-Touch at light and control display | Required-Touch at light and control display | Required-Touch at light and control display | Required-Touch at light and control display | Required-Touch at light and control display | Required-Touch at light and control display | Required-Touch at light and control display | Required-touch as light and control display | |||
| 3 | COLOR RENDERING INDEX | >90 | >90 | >90 | >90 | >90 | >90 | >90 | ≥97 | |||
| 3a | R9 value | minimum 90 | minimum 90 | minimum 90 | minimum 90 | minimum 90 | minimum 90 | minimum 90 | minimum 90 | |||
| 4 | FIELD SIZE, cm (in) | |||||||||||
| 4a | Diameter | Minimum of 2 choices, min, size 5" Max. size 10" | Minimum of 2 choices, min, size 5" Max. size 10" | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | 19 - 31 cm | |||
| 5 | FOCAL LENGTH, cm (in) | ≥65 | ≥65 | ≥65 | ≥65 | ≥65 | ≥65 | ≥65 | ≥65 | |||
| 6 | ILLUMINATION LEVEL, maximum lux at 1 m | ≥160,000 | ≥160,000 | ≥160,000 | ≥160,000 | ≥160,000 | ≥160,000 | ≥160,000 | ≥160,000 | |||
| 7 | MOUNTING | Boom only | Boom only | Boom only | Boom only | Boom only | Boom only | Boom only | Boom only | |||
| 8 | CONTROLS | |||||||||||
| 8a | Dimmer | Required, At light and control display. 5 -100% | Required, At light and control display. 5 -100% | Required, At light and control display. 5 -100% | Required, At light and control display. 5 -100% | Required, At light and control display. 5 -100% | Required, At light and control display. 5 -100% | Required, At light and control display. 5 -100% | Required, At light and control display. 5 -100% | |||
| 8b | Focus | At light and control display | At light and control display | Required, At light and control display | Required, At light and control display | Required, At light and control display | Required, At light and control display | Required, At light and control display | Required, At light and control display | |||
| 8bi | Automatic | Required | Required | Required | Required | Required | Required | Required | Required | |||
| 8c | Field size | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |||
| 8d | On sterile handle | Required | Required | Required | Required | Required | Required | Required | Required | |||
| 9 | ROTATION, ° | The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location. | The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location. | The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location. | The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location. | The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location. | The Surgical Lights must be of a design that allows for 360 degrees of rotation at any yoke or support location. | ≥420 degrees at the lamp unlimited at the suspension | ||||
| 10 | VERTICAL ADJUSTMENT RANGE, cm (in) | > than 30" | > than 30" | ≥80 cm | ≥80 cm | ≥80 cm | ≥80 cm | ≥80 cm | ≥125 cm | |||
| 11 | HEAT | |||||||||||
| 11a | Maximum irradiance at 1 m, W/m2 | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |||
| 11b | Heat-to-light ratio, mW/m2.lux | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |||
| 12 | LIGHTHEAD MATERIAL | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |||
| 13 | LENS COVER MATERIAL | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | Polycarbonate? Why? Material strength is better so will prevent being broken. | |||
| 14 | FILTER MATERIAL | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |||
| 15 | REFLECTOR MATERIAL | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |||
| 16 | SHADOW CONTROL | |||||||||||
| 16a | Automatic shadow dilution adjustment | Automatically by in-light control | Automatically by in-light control | Automatically by in-light control | Automatically by in-light control | Automatically by in-light control | Automatically by in-light control | Automatically by in-light control | Automatically by in-light control | |||
| 17 | SATELLITES | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |||
| 18 | AMBIENT LIGHTING | Required | Required | Required | Required | Required | Required | Required | Required | |||
| 19 | BATTERY OR EMERGENCY BACKUP | Required | Required | Required | Required | Required | Required | Required | Required | |||
| 20 | STERILIZABLE HANDLE | Required | Required | Required | Required | Required | Required | Required | Required | |||
| 21 | VIDEO CAMERA | |||||||||||
| 21a | Compatible (type) or integrated-separate video source incorporated into one of the surgical lights brought to one of the surgical fields streaming live video | Yes, HD minimum | Yes, HD minimum | Yes, HD minimum | Yes, HD minimum | Yes, HD minimum | Yes, HD minimum | Yes, HD minimum | Yes, HD minimum | |||
| 21b | Location on lighthead | Light | Light | Light | Light | Light | Light | Light | Light | |||
| 22 | MINIMUM CEILING HEIGHT, cm (in) | 9' 11.5" | 9' 6" | 9' 4" | 9' 4" | Curent ceiling height 12'-11.5" | Curent ceiling height 12'-11.5" | 10 ft | OR 2-6=120 inch |
OR 1=122 inch OR 8 =107 inch ICU rooms =108 inch CT room 2 =108 inch CT room 1 =101 inch GI rooms=94 inch
| 23 | DIAMETER, cm (in) | |||||||||
| 23a | Lighthead | No minimum requirement | No minimum requirement | =>30" | =>30" | No minimum requirement | No minimum requirement | CURRENTLY 24", NOT REQURED BUT SOMETHING CLOSE | No minimum requirements | |
| 24 | WEIGHT, kg (lb) | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |
| 25 | SUPPLY VOLTAGE, VAC | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |
| 26 | POWER, W | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements | |
| 27 | PURCHASE INFORMATION | |||||||||
| 27a | Warranty | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | |
| 28 | ADDITIONAL SPECIFICATIONS | |||||||||
| 28a | Ability to turn light on/off at light | Required | Required | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirements | No minimum requirements |
Booms
| Item | Description | Fargo-OR 1-4 | Fargo-OR 5 | St. Cloud OR 1-3 | Central Iowa-Cysto Room | Central Iowa-OR 2-5 | Black Hiills-OR 1-3 | Iowa City-OR 1 | Iowa City-OR 2 | Iowa City-OR 3 | Iowa City-OR 4 | Iowa City-OR 5 | Iowa City-OR 6 | Iowa City-GI 1 | Iowa City-GI 2 | Iowa City-GI 3 | Iowa City-CT 1 | Iowa City-CT 2 | Iowa City-ICU 1 | Iowa City-ICU 2 | Iowa City-ICU 3 | Iowa City-ICU 4 | Iowa City-ICU 5 | Iowa City-ICU 6 | Iowa City-ICU 7 | Iowa City-ICU 8 | Iowa City-ICU 9 | Iowa City-ICU 10 | Total | |
| 1 | Number of Anesthesia Booms Per Room | 4 | 1 | 3 | 0 | 0 | 3 | 1 | 1 | 1 | 1 | 1 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 17 | |
| 2 | Number of Equipment Booms Per Room | 4 | 1 | 3 | 1 | 4 | 3 | 2 | 2 | 1 | 2 | 2 | 2 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 42 | |
| 3 | SWIVEL RADIUS, cm (in) | 330 degrees | 330 degrees | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 4 | HEIGHT ADJUSTMENT, cm (in) | 24" | 24" | Powered Vertical Articulation | No minimum requirement | No minimum requirement | No minimum requirement | 24" | 24" | 24" | 24" | 24" | 24" | No minimum requirement | No minimum requirement | No minimum requirement | 24" | 24" | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 5 | ELECTRICAL FEATURES | |||||||||||||||||||||||||||||
| 5a | Circuit breaker | No minimum requirement at boom | No minimum requirement at boom | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 5b | Electrical power receptacles | 5ea. Emer. duplex, 2ea. normal duplex | 5ea. Emer. duplex, 2ea. normal duplex | Greater or equal =>8 and Anesthesia boom shall have 8 duplex power receptacles. | No minimum requirement | No minimum requirement | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 10 | 4 | 4 | 12 | 12 | 12 | 12 | 12 | 12 | 12 | 12 | 12 | 12 | ||
| 5c | Equipotential ground bus | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 5d | Grounding jacks | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 5e | Telephone jack | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | ||
| 5f | Isolated power system | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | Required | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 5g | Code alarm button | No minimum requirement | No minimum requirement | Required | No minimum requirement | No minimum requirement | Not required | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 1 | 1 | 0 | 0 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | ||
| 5h | Nurse call system | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | Not required | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | ||
| 5i | Elapsed timer/clock | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | Not required | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 6 | MEDICAL GAS SERVICES | |||||||||||||||||||||||||||||
| 6a | Oxygen outlets | 2ea. | 2ea. | 2 | No minimum requirement | No minimum requirement | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 1 | 1 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | ||
| 6b | Outlet type | Ohio | Ohio | DISS | No minimum requirement | No minimum requirement | Diamond | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | CHEMETRON | ||
| 6c | Vacuum inlets | 2ea. Ohio style | 2ea. Ohio style | 2 | No minimum requirement | No minimum requirement | 3 | 1 | 1 | 1 | 1 | 1 | 1 | 2 | 2 | 2 | 2 | 2 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | ||
| 6d | Medical compressed air | 2ea. Ohio style | 2ea. Ohio style | 2 | No minimum requirement | No minimum requirement | 1 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 0 | 0 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | ||
| 6e | Nitrous oxide | No minimum requirement | No minimum requirement | 1 | No minimum requirement | No minimum requirement | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | ||
| 6f | WAG evacuation | No minimum requirement | No minimum requirement | 1 | No minimum requirement | No minimum requirement | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 0 | 0 | 0 | 1 | 1 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | ||
| 6g | Medical gas piping | Conductive- meets NFPA 99 standard | Conductive- meets NFPA 99 standard | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | Meets NFPA 99 standard | ||
| 6h | Vacuum bottle slides | slides or comparative | slides or comparitive | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | ||
| 6i | Vacuum bottle storage | Not required | Not required | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | 3 | ||
| 7 | ACCESSORIES | |||||||||||||||||||||||||||||
| 7a | Monitor bracket | bracket or arm for monitor if requested | bracket or arm for monitor if requested | Required | No minimum requirement | No minimum requirement | REQUIRED | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 1 | 1 | 0 | 0 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | ||
| 7b | Bed bumper | No minimum requirement | No minimum requirement | Required | No minimum requirement | No minimum requirement | No minimum requirement | No | No | No | No | No | No | Yes | Yes | Yes | No | No | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | Yes | ||
| 7c | IV supports | Required | Required | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 1 | 1 | 1 | 1 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | ||
| 7d | Catheter basket | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 1 | 1 | 1 | 1 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | 2 | ||
| 8 | H x W x D, cm (in) | No minimum requirement | No minimum requirement | Horizontal Anesthesia Boom | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 9 | WEIGHT, kg (lb) | No minimum requirement | No minimum requirement | >=100 lbs per shelf | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 10 | LOAD CAPACITY, kg (lb) | 350 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 250 lbs or greater per arm | 250 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 350 lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | 150lbs or greater per arm | ||
| 11 | PURCHASE INFORMATION | |||||||||||||||||||||||||||||
| 11a | Warranty | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | Minimum of 1 year | ||
| 12 | ADDITIONAL SPECIFICATIONS | |||||||||||||||||||||||||||||
| 12a | Nitrogen | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | Must have regulated Nitrogen outlet/w gauge | ||
| 12b | Network Jacks | 4 plex or more in each room | 4 plex or more | Anesthesia boom shall have 5 network jacks | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | 4 plex or more | ||
| 12c | Keyboard/mouse | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | 1 | ||
| 12d | Adustable Equipment Shelves | Required | Required | Required. Equipment boom shall have 4 adjustable shelves with => 100 lb load capacity per shelf | Required | Required | Required | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 4 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | ||
| 12e | Carbon Dioxide on equipment boom | Required | Required | Required | Required | Required | Required | 1 DISS | 1 DISS | 1 DISS | 1 DISS | 1 DISS | 2 DISS | 2 DISS | 2 DISS | 2 DISS | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | ||
| 12f | Medical Gas Connections | Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration. |
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging Each room shall have the foolowing Medical Gas connections available. Quantities to be determined by each room configuration.
-Oxygen -Medical Air -Vacuum -Nitrogen
-CO2
-Waste Gas Scavenging
| 12g | Universal connections for video inputs | •Equipment boom shall have 6 universal connections for video inputs | ||||||||||||||||||||||||||||
| •Anesthesia boom shall have 3 universal connections for video inputs | •Equipment boom shall have 6 universal connections for video inputs | |||||||||||||||||||||||||||||
| •Anesthesia boom shall have 3 universal connections for video inputs | •Equipment boom shall have 6 universal connections for video inputs | |||||||||||||||||||||||||||||
| •Anesthesia boom shall have 3 universal connections for video inputs | •Equipment boom shall have 6 universal connections for video inputs | |||||||||||||||||||||||||||||
| •Anesthesia boom shall have 3 universal connections for video inputs | •Equipment boom shall have 6 universal connections for video inputs | |||||||||||||||||||||||||||||
| •Anesthesia boom shall have 3 universal connections for video inputs | •Equipment boom shall have 6 universal connections for video inputs | |||||||||||||||||||||||||||||
| •Anesthesia boom shall have 3 universal connections for video inputs | Equipment boom must have ≥ 5 video/image capture inputs and Anesthesia boom must have ≥ 4 video/capture inputs. | Equipment boom must have ≥ 5 video/image capture inputs and Anesthesia boom must have ≥ 4 video/capture inputs. | Equipment boom must have ≥ 5 video/image capture inputs and Anesthesia boom must have ≥ 4 video/capture inputs. | Equipment boom must have ≥ 5 video/image capture inputs and Anesthesia boom must have ≥ 4 video/capture inputs. | Equipment boom must have ≥ 5 video/image capture inputs and Anesthesia boom must have ≥ 4 video/capture inputs. | Equipment boom must have ≥ 5 video/image capture inputs and Anesthesia boom must have ≥ 4 video/capture inputs. | Equipment boom must have ≥ 5 video/image capture inputs. | Equipment boom must have ≥ 5 video/image capture inputs. | Equipment boom must have ≥ 5 video/image capture inputs. | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | |||||||||
| 12h | Video integration Foot Switch / Sync cable | No minimum requirement | No minimum requirement | Equipment boom shall have 1 video integration Foot Switch / Sync cable | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 12i | Smoke Evacuator | No minimum requirement | No minimum requirement | Equipment boom shall have a smoke evacuator | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 12j | Storage space for ESU, electrosurgical unit, foot switches | No minimum requirement | No minimum requirement | Equipment boom shall have storage space for ESU foot switches in additional to shelving | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 12k | Mounts for display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | Must have mounts for at least 3 display monitors | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 12l | CO2 Cylinder mount | No minimum requirement | No minimum requirement | A CO2 cylinder mount for a backup CO2 tank on equipment boom | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | ||
| 12m | Arm Support | a) From the ceiling hub location, the arms supporting the devices must be able to reach a minimum length of 9 feet. | ||||||||||||||||||||||||||||
| b) The Equipment carriers have an optional side support to help protect the devices from collision damage. | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement | No minimum requirement |