Attachment B Statement of Need AMSCO 400 SERIES MEDIUM STEAM STERILIZERS.pdf

PDF 165 KB Posted

Attached to
6515--Equipment: MEDIUM STEAM STERILIZERS Federal contract opportunity
Solicitation number
36C24823Q0421
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 8

About this file

This document outlines requirements for medium steam sterilizers. The North Florida/South Georgia VA Healthcare Systems requires medium steam sterilizers that can process up to 12 trays weighing 25 pounds each, with features including an electric vacuum pump, sensors to minimize cooling water needs, and dimensions of 26 x 37.5 x 48 inches. The sterilizers must have single hand door hinges, loading docking stations, safe and easy front and left side access to components, remote monitoring via ProConnect, multi-color touchscreen displays, and safety controls. Installation is required by February 3, 2023 at the Lake City VA Medical Center in Lake City, Florida, and must be performed by licensed teams with healthcare assembly experience. The related federal contract opportunity is solicitation number 36C24823Q0421 for these medium steam sterilizers through the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 8.

View the file

Other files for this federal contract opportunity

Other files attached to 6515--Equipment: MEDIUM STEAM STERILIZERS, newest first.
File Type Posted
Drawings.pdf PDF
36C24823Q0421 0001 SF-30 with Questions and Answers.pdf PDF
36C24823Q0421 0001.docx DOCX document
Attachment C DELIVERY SCHEDULE.pdf PDF
Attachment A Combined Synopsis Solicitation - Medium Steam Sterilizers.pdf PDF
36C24823Q0421.docx DOCX document
Attachment D VA Limitations on Subcontracting.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Attachment B 36C24823Q0421

Statement of Need

Background: The North Florida/ South Georgia VA Healthcare Systems has a requirement for - Medium Steam

Sterilizers.

Scope: Scope: The Contractor shall provide all new equipment, accessories, deliver, install new equipment, and perform all required operational testing with the SPS Department. Contractor shall also, deinstall specified equipment. All equipment and installation must meet Original Equipment Manufacturers (OEM) and VA specifications. All new equipment must be compatible with current operating equipment in place; (previously purchased from

Steris).

Salient Features Include:

Features, Fit, And Form:

o Must be able to process up to 12 trays weighing 25lbs each.

o Must come with an electric vacuum pump o Must contain sensors that minimize cooling water needs o Must have wall space 26 x 37.5 x 48" (660 x 950 x 1220 mm) o Must have single hand door hinge o Must have loading docking station o Must have safe and easy access to all components from the front and left side of sterilizer o Must have real-time remote monitoring of sterilizer function and cycle parameters with ProConnect o Must have operator interface control panels o Must have multi-color touch-sensitive screen graphics 30-line x 40-character display o Must have Ink-On-Paper Impact Printer o Must have safety control lockout switch o Must have chamber float switch o Must have safety pressure relief valve o Must have power multi-door safety features o Must Non-Operating End (NOE) Control Panel o Must have Lighted DIN connectors o Must contain ProConnect® Technical Support Services o Prevac Cycles (4-minute exposure) - 270°F (132°C) o Prevac Cycle (3-minute exposure) - 275°F (135°C) o Gravity Cycle (30-minute exposure) - 250°F (121°C) o Liquid Cycle (45-minute exposure) - 250°F (121°C)

• Water Supply o Sterilizer Cold Water Supply Temperature Requirement 70°F (21°C) Max o 20-50 PSIG Dynamic (1.38 - 3.45 BAR) o 1” NPT o Peak – 13 GPM (49.2 LPM) o Consumption Average – 26 x 37.5 x 48 – 120 GPH (8.0 LPM)

Attachment B 36C24823Q0421

• Drain o 2" ODT drain terminal (Floor Drain Capacity Must Handle Peak Water Consumption)

• Steam o 1" NPT, 50 to 80 psig (3.5 to 5.6 bar) dynamic, 97% to 100% vapor quality

Peak – 26 x 37.5 x 48 – 340 LB/HR (154 KG/HR)

Average - 26 x 37.5 x 48 – 148 LB/HR (67 KG/HR)

• Sterilizer Feed Water o 1" NPT, 20 to 50 psig (1.4 to 3.5 bar) dynamic. Water is used for vacuum pump, heat exchanger and trap cooling

• Electrical - Controls o 120 Volt, 50/60 Hz, 1-phase, 2.0 Amp (3A Circuit Breaker Recommend)

• Electrical – Vacuum Pump o 208/240 Vac, 50/60 Hz, 3-phase, 6 Amp per phase (8A Circuit Breaker Recommended) o 380/415 Vac, 50/60 Hz, 3-phase, 3 A per phase o 480 Vac, 50/60 Hz, 3-phase, 3 Amp per phase (4A Circuit Breaker Recommended)

Facilities Management Requirements:

o Deinstall current Steris AMSCO® Century V-148H Prevac Steam Sterilizers.

o Do all plumbing and provide mixing valves.

o Provide water supply

Delivery Date of Equipment: 2/03/2023 or within 30 days after contract is awarded.

Warranty of Equipment: Contractor must provide manufacturer(s) warranty on all provided equipment/supplies.

Place of Performance: Lake City VA Medical Center, 619 S. Marion Ave., Lake City, Florida 32025

Note: Complete onsite assembly and installation Services: Installation shall be performed by licensed, bonded, vaccinated healthcare storage assembly teams with ample experience working under the time restricted needs of the OR and MSD.