Attachment B Salient Characteristics_Visionix Rentinomax K_5 or Equivalent.docx

DOCX document 16 KB Posted

Attached to
6515--691-Hand-Held Autorefractors
Solicitation number
36C26225Q1358
Issued by

About this file

This document is an Attachment B Salient Characteristics specification for a Visionix Retinomax K+5 or Equivalent autorefractor and keratometer. The Department of Veterans Affairs (VA) is seeking to procure two units through a Service-Disabled Veteran-Owned Small Business (SDVOSB) set-aside solicitation under NAICS code 339112. The medical device must be an FDA-registered, handheld, portable autorefractor capable of taking measurements on patients in various positions, with specific technical requirements including a rechargeable battery, LCD display, and measurement capabilities for refractive errors ranging from -20.00 to +20.00 diopters, with cylinder ranges up to +/- 10 diopters and a minimal pupil size measurement of 3.0mm. The solicitation (36C26225Q1358) will result in a single Firm-Fixed Price (FFP) award and is being conducted using simplified acquisition procedures.

View the file

Other files

File Type Posted
36C26225Q1358_1.docx DOCX document
Copy of Attachment A - 36C26225Q1358 Visionix Rentinomax Brand Name or Equivalent.xlsx XLSX spreadsheet
36C26225Q1358.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Visionix Retinomax K+5 or Equivalent

a) The unit must be an FDA registered autorefractor (refractometer) and keratometer. *An autorefractor is a portable medical device used to accurately measure an individual's refractive error, which is the basis for glasses or contact lens prescriptions. *A keratometer (ophthalmometer), is an instrument used to measure the curvature of the cornea that assists with fitting contact lenses.

· Unit must offer refractive and keratotomy evaluations

· Unit must be handheld and portable.

· Unit must be capable of taking measurements on a patient who is partially or fully reclined (supine)

· Unit must function on a rechargeable battery and have an LCD display.

· Unit must offer autorefraction in plus or minus cylinder format with a range on spherical component -20.00 diopters to +20.00 diopters, cylinder range up to +/-10 diopters

· Unit must offer autokeratometry with range up to +/- 10 diopters

· Unit must provide a minimal pupil size measurement of 3.0mm

· Unit must offer astigmatism detection. *A condition where the eye is not perfectly round.

File details come from the government source that posted it. Updated .