This is a salient characteristics document (Attachment B) detailing the technical requirements for a Pulsed Electric Field (PEF) Ablation System for the VA North Florida/South Georgia Veterans Health System in Gainesville, FL. The document specifies requirements for a non-thermal pulmonary soft tissue ablation system for use in bronchoscopy procedures.
The system must be FDA approved for endoscopic lung nodule/cancer treatment and include a generator, cardiac monitor with synchronization capabilities, user manual, and specific electrodes (5-pack), flex needles (5-pack), and 19G needles (5-pack). Key technical requirements include high-voltage short-duration electrical energy delivery, preservation of surrounding tissue, constant ablation size, 4-lead cardiac synchronization, LCD touchscreen display, and various connection ports for power, foot switch, and electrodes. The system requires a minimum 1-year manufacturer warranty and includes on-site staff education for clinical personnel.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Attachment D VA Limitations on Subcontracting.pdf | ||
| 36C24825Q0455.docx | DOCX document | |
| Attachment A Combined Synopsis Solicitation.docx | DOCX document | |
| Attachment C Price Delivery Schedule.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
36C24825Q0455 Attachment B – Salient Characteristics
The Pulsed Electric Field (PEF) Ablation System shall meet or exceed the following characteristics:
· Shall be able to provide one (1) PEF Ablation System with consumables that supports non-thermal pulmonary soft tissue ablation procedures in the Bronchoscopy suite setting.
· Shall be FDA approved for endoscopic treatment of lung nodules / cancer intervention.
· Deliver high voltage in short duration electrical energy locally that alters the transmembrane potential of a cell by disrupting cellular homeostasis and inducing non-thermal programmed cell death.
· Able to preserve surrounding tissue.
· Provide potential immune activation.
· Provides a constant ablation size for a predictable ablation zone.
· Shall provide a compatible cardiac monitor for synchronization of the delivery of energy with the patient’s cardiac cycle.
· 4-lead configuration that detects the R-wave of the EKG/ECG signal, then sends a trigger signal to the generator for exact synchronization of energy output.
· Generator includes software with algorithm that detects trigger signals from cardiac monitor.
· PEF system shall include:
· Generator
· Cardiac Monitor with Synchronization
· User manual
· Electrodes (system specific, 5 pack)
· Flex Needles (system specific, 5 pack)
· 19G Needles (system specific, 5 pack)
· Display: LCD front panel display with touch screen capable of displaying graphics, messages, activation information, and receiving touch screen input from operator.
· Provide displays of power output and current output.
· Provide connections for:
· AC line power
· Foot switch
· Synchronization input
· Ablation device connector
· Dispersive electrode connections
· Manufacturer warranty: minimum 1 year.
· Provide staff education on-site to all clinical staff utilizing equipment.