PRODUCT DESCRIPTION

For

NON-EXPENDABLE EQUIPMENT

VERATHON PRIME PLUS BRAND NAME OR EQUAL

BLADDER SCANNERS

Prepared By:

Strategic Acquisition Center Department of Veterans Affairs 10300 Spotsylvania Avenue, Suite 400 Fredericksburg, VA 22408

1. SCOPE OF WORK:

The Veterans Health Administration (VHA) Non-Expendable Equipment National Program has identified the Bladder Scanner product line as a candidate for a VA-wide (otherwise referred to as “national”) single Requirements contract award.

Bladder scanners are ultrasound-based devices primarily used to identify patients who have urinary retention of various etiologies. Bladder Scanners (also referred to as bladder ultrasound devices) are specifically designed to non-invasively assess urinary bladder volume, urinary retention, and post-void residual bladder volume. These devices may be used in a variety of clinical settings including acute care hospitals for assessing urinary retention, free-standing urology practices for assessing overall urinary function and urodynamic studies, and emergency departments for assessing obstructive uropathy.

The Vendor must demonstrate the ability to meet all requirements of the solicitation. The objective is to provide bladder scanners to be used by clinicians throughout the VA medical centers and facilities. The period of performance is for one 12-month base period with two 12-month option periods.

2. REQUIREMENT

This requirement will be IAW FAR clause 52.211-6, “Brand Name or Equal” which requires the vendor indicate that each product being offered as an “equal” product to the Verathon Prime Plus bladder scanner. For each equal product, the vendor must include a description reflecting the salient characteristics (SC) and level of quality that will satisfy the salient physical, functional, or performance characteristics of the equal product(s) specified in the solicitation. The vendor must also clearly identify the item by brand name (if any) and make/model number. Finally, the vendor must include descriptive literature, such as illustrations, drawings and clearly describe any modifications it plans to make to a product to make it conform to the solicitation requirements.

Contract Line Items
Manufacturer
Part Number
Description
0001
Verathon
0270-0870
Bladderscan Prime Plus System
0002
Verathon
0800-0532
Bladderscan Prime Mobile Cart
0003
Verathon
0270-0868
Bladderscan Prime Thermal Printer
0004
Verathon
0800-0540
Bladderscan Prime Carrying Case

The Department of Veterans Affairs (VA) is seeking vendors who can provide Verathon Prime Plus Bladder Scanners and associated equipment as listed above or equal commodities which meet all of the following salient characteristics(SC):

CLIN 0001

Bladderscan Prime Plus System Part Number 0270-0870

SC
Salient Characteristics
Method of Evaluation
Location in Literature
1
Shall compute bladder volume via 3D ultrasonography scan
Literature
2
Shall have the capability of scan modes for male and female patients.
Literature
3
Shall have a volume measurement range of at least 1 - 999 ml.
Literature
4
Shall have a maximum measurement accuracy of at least +/-15 ml or +/- 15%.
Literature
5
Shall have a Pre-scan function.
Literature
6
Shall have B-mode-slice or B-mode scan capability.
Literature
7
Shall be powered by Lithium Ion rechargeable batteries and 110-240V 50-60Hz AC power adapter.
Literature
8
Shall be equipped with a console display screen.
Literature