Attachment B .docx
DOCX document 17 KB Posted
- Attached to
- AJ91--QuantStudio 3 Real-Time PCR System Federal contract opportunity
- Solicitation number
- 36C26220Q1273
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| 36C26220Q1273 0001.docx | DOCX document | |
| Attachment A.xlsx | XLSX spreadsheet | |
| 36C26220Q1273.docx | DOCX document |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment B Brand Name or Equal Description/Requirements:
The Brand Name for the below is Thermo Fisher. See Attachment A for associated quantities.
| PRODUCT DESCRIPTION |
| PRODUCT # |
| QuantStudio 3 Real-Time PCR System |
| A28572 |
| Optical Adhesive Covers |
| 4360954 |
| Taqman RNAse P 96-well |
| 4432382 |
| 96-well Hard Shell Plates |
| 4483354 |
| SmartStart QuantStudio 3, Half-Day Training |
| A28806 |
| AB Assurance, QuantStudio 3 |
| ZG11SCQS3STD |
Salient characteristics –
1. Instrument UI:
a. Preferably able to run from touchscreen
b. Must be able to run from attached desktop computer (incl instrument software & at least 100GB free space on hard drive for data) attached
2. Block format / sample capacity / reaction volume:
a. 96-well, 0.2mL
3. Excitation source:
a. Fixed source, not shuttle configuration; no time delay
4. Optical detection & range:
a. 450-640nm
b. at least 3 filters
5. Multiplexing:
a. at least 3 targets
b. Maximum & average block ramp rate: max at least 4C/sec, avg at least 3C/sec
6. Temperature uniformity:
a. no more than 0.5C variation well-to-well
7. Temperature accuracy:
a. no more than 0.3C from programmed target
8. Programmable temperature zones
9. Factory-compatible with variety of commercial dyes
File details come from the government source that posted it. Updated .