ATTACHMENT A - SALIENT CHARACTERISTICS - ROTARY MICROTOME.docx

DOCX document 22 KB Posted

Attached to
6640--AMENDMENT 0001 - MICROTOMES INSTRUMENT
Solicitation number
36C26226Q0813
Issued by

About this file

This file is an attachment detailing the salient characteristics and product specifications for a rotary microtome procurement by the VA San Diego Healthcare System Pathology and Laboratory Department. The specific equipment required is the Leica Microsystems HistoCore MULTICUT (Configuration 149MULTI0C1) with a USA 15-AMP power cord (Product #14041157009), with one unit needed for in-house laboratory sectioning of paraffin-embedded samples for examination and diagnosis purposes.

The equipment must meet nine critical salient characteristics: include an antistatic waste tray; operate as semi-motorized with manual operation capability; feature a coarse feed wheel with selectable turn direction; move the object head between 200μm/s and 1800μm/s; include a user-adjustable force balance system; accommodate both low- and high-profile blades; provide individually replaceable blade holder components; not exceed dimensions of 477mm width × 620mm depth × 295mm height; and include a digital display. The corresponding federal solicitation (36C26226Q0813) is issued as a Request for Quotation on a one-time basis for the VA Greater Los Angeles Path and Lab Department and is set-aside for service-disabled veteran-owned small business (SDVOSB) concerns under the Department of Veterans Affairs Veterans Health Administration, Veterans Integrated Service Network 22.

View the file

Other files

File Type Posted
36C26226Q0813 0001_2.docx DOCX document
RFQ AMENDMENT - 36C26226Q0813 0001 - MICROTOMES.docx DOCX document
RFQ - 36C26226Q0813 v3 - MICROTOMES.docx DOCX document
RFQ 36C26226Q0813 - ROTARY MICROTOME.docx DOCX document
36C26226Q0813.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

VA San Diego Healthcare System (VASDHCS) Rotary Microtome

1. BRAND NAME OR EQUAL

BRAND
PRODUCT DESCRIPTION
PRODUCT #
QTY

Leica Microsystems

HistoCore MULTICUT – Configuration

149MULTI0C1

Leica Microsystems

Power cord USA - 15 AMP

14041157009

2. SALIENT CHARACTERISTICS

The salient characteristics describe the requirements for a high through-put rotary microtome needed by VASDHCS Pathology and Laboratory department utilized to maintain an in-house laboratory sectioning of paraffin embedded samples as necessary preparation for their examination and subsequent diagnosis.

1. Must have an antistatic waste tray

2. Must be semi-motorized with capability for manual operation.

3. Must have coarse feed wheel with selectable turn direction.

4. Must have the capability to move object head between 200um/s and 1800um/s.

5. Must have user adjustable force balance system a.

6. Must be capable of being used with both low- and high-profile blades..

7. Must have interchangeable parts that allow for the replacement of blade holder components individually.

8. The dimensions of the equipment must not exceed the following, width 477mm x depth 620mm x height 295mm.

9. Must have a digital display. Comment by Carabon, Kahlfanie K.: Is a display needed? If yes, does it need to be digital? Comment by Dominguez, Catalina: Yes, we need it be digital.

File details come from the government source that posted it. Updated .