ATTACHMENT A - SALIENT CHARACTERISTICS - ROTARY MICROTOME.docx
DOCX document 22 KB Posted
- Attached to
- 6640--AMENDMENT 0001 - MICROTOMES INSTRUMENT
- Solicitation number
- 36C26226Q0813
About this file
This file is an attachment detailing the salient characteristics and product specifications for a rotary microtome procurement by the VA San Diego Healthcare System Pathology and Laboratory Department. The specific equipment required is the Leica Microsystems HistoCore MULTICUT (Configuration 149MULTI0C1) with a USA 15-AMP power cord (Product #14041157009), with one unit needed for in-house laboratory sectioning of paraffin-embedded samples for examination and diagnosis purposes.
The equipment must meet nine critical salient characteristics: include an antistatic waste tray; operate as semi-motorized with manual operation capability; feature a coarse feed wheel with selectable turn direction; move the object head between 200μm/s and 1800μm/s; include a user-adjustable force balance system; accommodate both low- and high-profile blades; provide individually replaceable blade holder components; not exceed dimensions of 477mm width × 620mm depth × 295mm height; and include a digital display. The corresponding federal solicitation (36C26226Q0813) is issued as a Request for Quotation on a one-time basis for the VA Greater Los Angeles Path and Lab Department and is set-aside for service-disabled veteran-owned small business (SDVOSB) concerns under the Department of Veterans Affairs Veterans Health Administration, Veterans Integrated Service Network 22.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| 36C26226Q0813 0001_2.docx | DOCX document | |
| RFQ AMENDMENT - 36C26226Q0813 0001 - MICROTOMES.docx | DOCX document | |
| RFQ - 36C26226Q0813 v3 - MICROTOMES.docx | DOCX document | |
| RFQ 36C26226Q0813 - ROTARY MICROTOME.docx | DOCX document | |
| 36C26226Q0813.docx | DOCX document |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
VA San Diego Healthcare System (VASDHCS) Rotary Microtome
1. BRAND NAME OR EQUAL
| BRAND |
| PRODUCT DESCRIPTION |
| PRODUCT # |
| QTY |
Leica Microsystems
HistoCore MULTICUT – Configuration
149MULTI0C1
Leica Microsystems
Power cord USA - 15 AMP
14041157009
2. SALIENT CHARACTERISTICS
The salient characteristics describe the requirements for a high through-put rotary microtome needed by VASDHCS Pathology and Laboratory department utilized to maintain an in-house laboratory sectioning of paraffin embedded samples as necessary preparation for their examination and subsequent diagnosis.
1. Must have an antistatic waste tray
2. Must be semi-motorized with capability for manual operation.
3. Must have coarse feed wheel with selectable turn direction.
4. Must have the capability to move object head between 200um/s and 1800um/s.
5. Must have user adjustable force balance system a.
6. Must be capable of being used with both low- and high-profile blades..
7. Must have interchangeable parts that allow for the replacement of blade holder components individually.
8. The dimensions of the equipment must not exceed the following, width 477mm x depth 620mm x height 295mm.
9. Must have a digital display. Comment by Carabon, Kahlfanie K.: Is a display needed? If yes, does it need to be digital? Comment by Dominguez, Catalina: Yes, we need it be digital.