The attachment A - Salient Characteristics document details technical specifications for the Philips IntelliVue Multi-Measurement Module X3, a critical care patient monitoring device for the Department of Veterans Affairs Veterans Health Administration. The device is intended to replace 194 existing Philips IntelliVue X2 series units across GLA facilities, with specific requirements including a 5-hour battery life, wireless connectivity, 15.5 cm size with a color LCD touchscreen, and comprehensive monitoring capabilities for ECG, SpO2, respiration, blood pressure, and multiple other critical patient parameters.

Key technical specifications mandate the device be constructed from durable, antimicrobial hard plastic with Corning Gorilla Glass, designed for hospital settings with the ability to withstand drops up to 1 meter. The monitor must interface with existing Philips PIIC iX Central Station and be capable of monitoring critical care patients during transport and at bedside, with customizable screen configurations. The opportunity is a pre-solicitation Request for Information (RFI) under solicitation number 36C26225Q0490, issued by Veterans Integrated Service Network 22, seeking to upgrade their existing patient monitoring equipment.

View the file

Other files

File Type Posted
RFI_SSN Description - Transport Patient Monitors - 36C26225Q0490.docx DOCX document
36C26225Q0490.docx DOCX document

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

ATTACHMENT A – SALIENT CHARACTERISTICS

a.
Common Nomenclature (commercial description):
IntelliVue Multi-Measurement Module X3
b.
Kind of material (i.e., type, grade, alternatives, etc.):
Must be made out of durable hard plastic and antimicrobial Corning Gorilla Glass. Built from chemical-resistant materials to withstand disinfectants
c.
Electrical Data (i.e., 110 v, 60 Hz, 3 Ph, etc.):
Battery life must be 5 hours or more (5 hours of run time on a single charge). Must be able to connect wirelessly (via existing 802.11 network or IntelliVue Smart-Hopping system, to IntelliVue central stations as well as host-system devices and/or remote display screens.
d.
Dimensions, size, capacity:
Must be 15.5 cms (6.1 inches in size) with an active matrix color LCD display with a capacitive multi-touch screen (resolution 1024 x 480)
e.
Principles of operation:
Must be able to monitor ECG, SpO2 (Philips FAST-SpO2, Nellcor OxiMax SpO2, Masimo Rainbow SET SpO2), respiration, NBP, 2 invasive pressures – CO2 and second SpO2, cardiac output (right-heart and PiCCO), Covidien Microstream CO2, Respironics mainstream Co2 and/or LoFlo Sidestream CO2.
f.
Restrictive environmental conditions:
For hospital settings only (must survive a drop of up to 1 meter) and must be ergonomic to handle.
g.
Intended use:
Must be able to monitor critical care patients during transport and at the bedside. Reduces a number of steps to prepare patients for transport. Each screen can be customized to user needs
h.
Equipment with which the item is used with

(N/A if not applicable):

Must interface with existing Philips PIIC iX Central Station

i.
End item application:
Replacing the Philips IntelliVue X2 series (194 units)
j.
Original equipment manufacturer part number:
867030 (IntelliVue Multi-Measurement Module X3)
k.
Other pertinent information that describes the item, material or service required:
Note: All 194 existing Philips IntelliVue X2 devices are approaching end of life. These devices are integrated into our hospital operations in monitoring critical care patients. The Philips IntelliVue X3 are the upgrade devices to replace the 194 systems through GLA and its facilities.