Attachment A_Bill of Materials for RFQSSPR474450.pdf
PDF 952 KB Posted
- Solicitation number
- RFQSSPR474450
- Issued by
-
Department of Energy
About this file
This is a Bill of Materials for federal procurement reference PR474450, detailing a comprehensive Atomic Force Microscopy (AFM) system manufactured by Nanosurf. The document specifies 16 line items totaling a complete DriveAFM system with integrated accessories and services for laboratory research applications.
The core system comprises the Nanosurf DriveAFM scanhead (Item 1) featuring a 100 μm XY scan range, 20 μm Z range, 840 nm low-coherence SLD light source with motorized alignment, and advanced photodetector with noise specifications below 3 pm DC and 20 fm/√Hz AC. Supporting equipment includes the DriveAFM Control Unit (Item 2) with CX Controller offering 16 output channels and 13 input channels at varying resolutions, HV amplifier (-40V to +160V), rack-mount PC with Intel Core i7, 16GB RAM, and Windows 10 Pro, and a 21:9 ultra-wide screen monitor. Additional components specified are the motorized XY stage with 20 mm range and 1 μm resolution (Item 3), Isostage 300 active vibration isolation platform (Item 4), glove-box integration kit with feedthrough connectors (Item 5), AFM cantilever starter package with 30 cantilevers across three modes (Item 6), and DriveAFM Revision 2 scanhead with updated camera (Item 7). Accessory items include sample holders, temperature controllers, electrical isolation holders, environmental options, and advanced humidifier. The bill of materials concludes with Installation and Training (Item 15) and Shipping and Freight Charges (Item 16). FAR 52.211-6 Brand Name or Equal provision applies to the procurement, allowing equivalent products to be offered in lieu of the specified Nanosurf equipment.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Item Country of Origin Manufacturer Quantity Part Number Description
1 Nanosurf 1 Nanosurf DriveAFM
Nanosurf DriveAFM
- Max Scan Range o XY scan range: 100 μm o Z range: 20 μm
- Light Source o Wavelength: 840 nm low-coherence SLD o Diameter: 10 μm o Alignment: Motorized, Automated
- Photodetector o Bandwidth: >8 MHz o Noise ▪DC noise: <3 pm (RMS, 0.1 Hz - 1 kHz) ▪AC noise: <20 fm/√Hz above 100 kHz
- Measurement Noise o Dynamic Mode: <15 pm rms o Static Mode: <10 pm rms
- Position Sensor Noise o XY: 125 pm rms (0.1 Hz – 1 kHz) o Z: 55 pm rms (0.1 Hz – 1 kHz)
- Drive Resolution:
o XY: 0.4 pm o Z: 0.06 pm
- Approach: 10 mm motorized, parallel DriveAFM CleanDrive
- Photothermal Excitation light source: 785 nm laser
- Alignment: Motorized, Automated DriveAFM Top-View Camera
- USB3 5 MP color sensor, 2.2 μm pixel size, 15 FPS
- Field of view: 1100 x 850 μm2
- Magnification: 5.1x (air) / 5.4x (water)
- Resolution: <1.7 μm DriveAFM Conductive Cantilever Holder
- Grooveless - Conductive
Bill of Materials for PR474450
2 Nanosurf 1 DriveAFM Control Unit
DriveAFM Control Unit CX Controller
- Outputs (16 total):
o high resolution: 16x 28 bit @ 1 MHz (8 system, 4 user) o high speed: 4x 16 bit @ 100 MHz (3 system, 1 user)
- Inputs (13 total):
o high resolution: 10x 20 bit @ 1 MHz (6 system, 4 user) o high speed: 3x 16 bit @ 100 MHz (2 system, 1 user)
- Modes:
o Imaging Modes:
▪Static Force and Lateral Force Microscopy ▪Dynamic Force w/ Phase Contrast ▪MFM/EFM w/ Contour Following ▪Force-Modulation ▪Torsional Resonance Microscopy ▪WaveMode o Spectroscopy Modes:
▪Force-Distance ▪Amplitude-Distance ▪Stop-by-Value o Lithography Modes:
▪Standard o Modules:
▪Cantilever Calibration ▪Scripting Interface ▪Stage Control ▪ViewPort
HV Amplifier
- Range: -40 V to +160 V
- Noise: 26 μV rms (0.1 Hz to 1 kHz) Rack-mount PC and Monitor / - Intel Core i7 CPU, - 16 GB RAM, - 250 GB SSD, - 2 TB HDD, - 2 Ethernet ports, - Windows 10 Pro 64bit, - 21:9 wide screen monitor (3440x1440 pix)
3 Nanosurf 1 DriveAFM Motorized Stage
DriveAFM Motorized Stage
- Max XY range: 20 mm, closed loop
- Max Sample Diameter: 100 mm
- Max Sample Height: 26 mm
- Resolution: 1 μm
- Repositioning Accuracy: <2 μm
4 Nanosurf 1 Noise Isolation/Mitigation Noise Isolation/Mitigation Isostage 300 active vibration isolation platform
- Vibration Isolation: 23 dB (93%) at 5 Hz, 40 dB (99%) above 15 Hz
5 Nanosurf 1 Glove-Box Integration Kit
Glove-Box Integration Kit Feedthrough connectors for DriveAFM - these feedthroughs must be mounted in a panel that is mounted in the wall of the glovebox:
- Extension cables and adapters o DB37 (scan head) o DB15 (HV) o DB26HD (motorized stage) o DB25 (isostage)
6 Nanosurf 1 AFM Cantilever Starter Package (30 ct)
AFM Cantilever Starter Package (30 ct) includes:
o Dyn190 10ct (Dynamic Mode / Phase Contrast) o Mulit75 10ct (Torsional Resonance Microscopy) o WM0.8PTD 10ct (WaveMode)
7 nanosurf 1 Nanosurf DriveAFM Revision 2
Nanosurf DriveAFM Revision 2- Max Scan Range o XY scan range: 100 μm o Z range: 20 μm
- Light Source o Wavelength: 840 nm low-coherence SLD o Diameter: 10 μm o Alignment: Motorized, Automated
- Photodetector o Bandwidth: >8 MHz o Noise ▪DC noise: <3 pm (RMS, 0.1 Hz - 1 kHz) ▪AC noise: <20 fm/√Hz above 100 kHz
- Measurement Noise o Dynamic Mode: <15 pm rms o Static Mode: <10 pm rms
- Position Sensor Noise o XY: 125 pm rms (0.1 Hz – 1 kHz) o Z: 55 pm rms (0.1 Hz – 1 kHz)
- Drive Resolution:
o XY: 0.4 pm o Z: 0.06 pm
- Approach: 10 mm motorized, parallel DriveAFM Top-View Camera Revision 2
- USB3 5 MP color sensor, 2.2 μm pixel size, 15 FPS
- Field of view: 1100 x 850 μm2
- Magnification: 5.1x (air) / 5.4x (water)
- Resolution: <1.7 μm
8 Nanosurf 1 Upgrade Upgrade Existing DriveAFM Scanhead with Direct QPD Connection
9 Nanosurf 1 BT06647 BT06647 - Heater Sample Holder
10 Nanosurf 1 BT06952 BT06952 - Temperature Controller for all temperature stages
11 Nanosurf 1 BT10230 BT10230 - Electrically Isolated Holder w/ SMA connection
12 Nanosurf 1 BT08532 BT08532 - Tip-Access Module
13 Nanosurf 1 BT07066 BT07066 - Environmental Option
14 Nanosurf 1 BT08993 BT08993 - Advanced Humidifier
15 Nanosurf 1 Installation and Training Installation and Training
16 Nanosurf 1 Shipping and Freight Charges Shipping and Freight Charges
FAR 52.211-6 Brand Name or Equal shall apply
File details come from the government source that posted it. Updated .