This is a specifications document detailing the salient characteristics required for an Operating Room (OR) surgical table with integrated patient warming capabilities for the Cheyenne Veterans Affairs Medical Center (VAMC). The table must support a weight capacity of 1,200 lbs when centered and 1,000 lbs in articulated positions, with dimensions of approximately 84 inches in length and 20-22 inches in width. The table shall feature comprehensive articulation capabilities including slide articulation up to 18 inches for C-arm compatibility, Trendelenburg/reverse Trendelenburg adjustment of 30 degrees, lateral tilt of 20 degrees, and multiple adjustable sections (back, leg, head, and flex/reflex) to enable various surgical positioning including supine, prone, lateral, lithotomy, and Fowler's positions.

The table must include operational features such as hand-held pendant control with pre-programmed positions, battery operation with rechargeable batteries for power outage continuity, manual override capability, and auto-level functionality. Construction materials shall consist of stainless steel and radiolucent composite materials to support imaging compatibility with C-arm and x-ray equipment while maintaining infection control standards. The table must allow 360-degree rotation, feature removable head, arm, shoulder, and leg sections for surgical modifications, include accessory mounting for a broad range of surgical attachments, incorporate a patient warming system with minimal footprint integration, and be designed with sealed surfaces to facilitate easy cleaning and withstand repeated sterilization processes.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics: OR Surgical Table

· Shall offer a weight capacity up to 1,200 lbs. (544 kg) centered; 1,000 lbs. (454 kg) in articulated positions.

· Shall offer Length: approx. 84 in (213 cm) with headrest and foot section

· Shall offer Width: approx. 20 in (51 cm) (without side rails), 22 in (56 cm) (with side rails)

· Shall offer height adjustment from 26 in (66 cm) (lowest) to 45 in (114 cm) (highest)

· Shall offer articulations:

· Slide articulation: up to 18 in (46 cm) for improved imaging access (C-arm compatibility)

· Trendelenburg/Reverse Trendelenburg ±30 degrees

· Lateral Tilt: ±20 degrees

· Back Section: +90 to -40 degrees

· Leg Section: +20 to -105 degrees; Removable

· Flex/Reflex: 140 degrees / 100 degrees

· Head Section: +45 to -90 degrees; Removable and can be exchanged for accessories

· Shall offer Patient Positioning in the supine, prone, lateral, lithotomy, fowler’s, etc.

· Shall offer hand-held pendant controller with pre-programmed positions

· Shall offer Imaging Capability with carbon fiber table section to allow for unobstructed, radiolucent access for imaging (C-Arm, x-ray)

· Shall offer battery operation in addition to lined power, rechargeable batteries to allow continued operation in power outages

· Shall offer manual override provision in case of power failure

· Shall offer Auto-Level function

· Shall offer accessory mounting for capability with a broad range of surgical attachments

· Shall be constructed with stainless steel and radiolucent composite materials.

· Shall be easily cleaned and capable of withstanding repeated cleanings, and contained sealed surfaces to support infection control and easy maintenance

· Shall allow removable head, arm, shoulder, and leg sections to allow for various surgical modifications and positioning

· Shall allow 360-degree rotation capabilities for patient positioning

· Shall offer a patient warming system with simple integration and operate with a minimal footprint within the surgical area, thereby reducing excess surface area that could contribute to potential contamination or impede accessibility to patient care