Attachment 2_Specifications_1232SA26Q1516.pdf

PDF 230 KB Posted

Attached to
Bruker Brand Time-Domain Nuclear Magnetic Resonance Analyzer (TD-NMR)
Solicitation number
1232SA26Q1516
Issued by

About this file

This is a Specifications document for a federal procurement requirement by the Research, Education, and Economics (REE) Mission Area, Agricultural Research Service (ARS), National Center for Agricultural Utilization Research. The requirement is for a Bruker Brand Minispec mq20 Upgradeable Variable Temperature Time-Domain Nuclear Magnetic Resonance Analyzer (TD-NMR) to be delivered to the National Center for Agricultural Utilization Research located at 1815 North University Street, Peoria, Illinois 61604.

The contractor must provide a complete TD-NMR system meeting specific technical specifications including a 20 MHz transmitter, EMI-shielded and CE-conformed closed housing, programmer frequency range of 5-65 MHz with 1 Hz stepping, pulse programmer with 20 nanosecond timing resolution, and phase resolution better than 0.2 degrees. The system must include a 20 MHz probe assembly with 10 mm ratio probe featuring variable temperature capability from -5°C to 65°C, a thermostat bath (-20 to +150°C), pre-calibrated Solid Fat Content (SFC) calibration samples (0% to 70%), and an automated sample changer system compatible with existing TC3/TC6 baths. The deliverable must include a standalone Windows 11 computer with Minispec software (Automation Release 2.1 compatible), support for international standard methods (AOCS Cd 16b-93, AOCS Cd 16-81, ISO 8292, and IUPAC 2.150), and onsite installation, training, testing, and verification. The contractor must also provide shipping and import costs, with a minimum one-year warranty from completed installation.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Specifications

Background

The Research, Education, and Economics (REE) Mission Area, Agricultural Research Service (ARS), National Center for Agricultural Utilization Research has a requirement for a Bruker Brand Minispec mq20 Upgradeable Variable Temperature Time-Domain Nuclear Magnetic Resonance Analyzer (TD-

NMR).

The contractor must provide a TD-NMR analyzer in accordance with the following salient characteristics.

• 20 MHz transmitter

• EMI-shielded closed housing

• CE-conformed closed housing

• Programmer Frequency range of 5-65 MHz, with 1 Hz stepping and resolution 1 ppm or better.

• Pulse programmer of 20 ns timing resolution

• Phase resolution of better than 0.2°

• Receiver/Analog Digital converter:

• Must be compatible with Minispec Software or provide and upgrade format compatible with Windows 11 (current Minispec software is Automation Release 2.1; Microsoft NET Framework 3.0) for data acquisition

• 20 MHz probe assembly with 10 mm ratio probe with variable temperature capability ranging from at least -5 °C to 65°C with provided cryostat bath and matching daily check samples and 20 MHz pre-amplifier

• PC/keyboard/Monitor/mouse with Microsoft Windows 11 for running Minispec software.

NOTE: Computer not connected to USDA network (standalone computer).

• Certified Solid Fat Content (SFC) Calibration samples ranging from 0% to 70% Solid Fat Content.

• All types of SFC methods including direct/indirect and parallel/serial methods are supported.

• Automated sample changer system compatible with existing TC3/TC6 baths (10 mm probe and sample vial size), and minispec software.

• Following international standard methods: AOCS Cd 16b-93, AOCS Cd 16-81, ISO 8292 and IUPAC 2.150.

• Thermostat bath (115 V/60 Hz), -20 to +150 °C; Bath opening (W x D)/bath depth: 13 x 15 cm / 15 cm.

• Shipping/Import Costs/Handling

• 1+ year warranty from completed installation

• Onsite installation, training, testing, verification.