Attachment 2_Specifications_1232SA26Q0750.pdf

PDF 382 KB Posted

Attached to
Bio-Rad ChemiDoc Go Imaging System and Image Lab Touch Software (Brand Name or Equal) Federal contract opportunity
Solicitation number
1232SA26Q0750
Issued by
Department of Agriculture Agricultural Research Service

About this file

This document is a purchase specification for laboratory imaging equipment issued by the USDA/ARS/Produce Safety & Microbiology division in Albany, California.

The procurement requires the purchase of a ChemiDoc Go Imaging System with Image Lab Touch Software (brand name or equal). The system must meet specific technical requirements to support critical laboratory workflows for protein and nucleic acid analysis. Key mandatory specifications include: a high-resolution detector of at least 20 MP CMOS with deep cooling for chemiluminescence detection; all-LED internal illumination (UV, Blue, Amber, Green, and White) to eliminate bulb replacement needs; Smart Tray Technology that automatically recognizes application-specific trays and adjusts imaging parameters accordingly; a large imaging area of at least 21 x 14 cm capable of imaging up to four mini-gels or blots simultaneously; an onboard multi-touch screen of 9.7 inches or larger running integrated software for image capture and analysis without requiring an external PC; and automated imaging capabilities including pre-calibrated autofocus and auto-exposure algorithms. The delivery location is USDA/ARS/Produce Safety & Microbiology at 800 Buchanan Street, Albany, CA 94710, with a required delivery schedule of 60 days after receipt of order (ARO).

View the file

Other files for this federal contract opportunity

Other files attached to Bio-Rad ChemiDoc Go Imaging System and Image Lab Touch Software (Brand Name or Equal), newest first.
File Type Posted
Attachment 1_Solicitation Terms and Conditions_1232SA26Q0750.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Purchase Bio-Rad ChemiDoc Go Imaging System with Image Lab Touch

Software (Brand Name or Equal)

1.0 Specifications

The ChemiDoc Go Imaging System is required to support critical laboratory workflows for protein and nucleic acid analysis. This system is uniquely qualified due to the following technical necessities:

Sensitivity & Resolution: The 20 MP CMOS detector with deep cooling is essential for detecting low-abundance proteins in chemiluminescence blots, providing sensitivity comparable to traditional film without the environmental and time costs of a darkroom.

Operational Efficiency: The "Smart Tray Technology" and automated pre-calibrated autofocus/auto-exposure algorithms minimize user error and variability between experiments, which is crucial for high-throughput diagnostic platform development.

Internal Illumination: All-LED internal provides a broad range of excitation wavelengths for various stains while eliminating the recurring maintenance costs and downtime associated with bulb replacements.

Benchtop Integration: The large imaging area (21 x 14 cm) combined with an onboard 9.7" multi-touch screen eliminates the requirement for a dedicated external PC, optimizing valuable lab bench space and allowing for rapid, standalone image acquisition and analysis.

Must meet the following specifications:

Must have a high-resolution detector (at least 20 MP CMOS) with deep cooling to provide sensitivity for chemiluminescence film.

Must have all-LED internal illumination (UV, Blue, Amber, Green, and White) to eliminate the need for bulb replacements and provide precise excitation for various stains.

Must have "Smart Tray Technology" that automatically recognizes the application-specific tray and adjusts imaging parameters and software options accordingly.

Required to have a large imaging area (at least 21 x 14 cm) capable of imaging up to four mini-gels or blots simultaneously.

Required to have onboard, multi-touch screen operation (9.7" or larger) running integrated software for image capture, eliminating the need for an external PC at the bench.

Required to have automated imaging capabilities, including pre-calibrated autofocus and auto-exposure algorithms for rapid and optimal image acquisition.

2.0 Place of Delivery

USDA/ARS/Produce Safety & Microbiology 800 Buchanan St.

Albany, CA 94710

3.0 Delivery Schedule

60 days ARO