Attachment 2_Specifications_1232SA26Q0338.pdf
PDF 252 KB Posted
- Attached to
- Vibratome Instrument Federal contract opportunity
- Solicitation number
- 1232SA26Q0338
About this file
This document is a specifications attachment for a vibratome procurement. The vibratome must be capable of cutting unfixed, live tissues with minimal distortion, no tearing, and reduced mechanical artifacts to preserve specimen integrity for downstream analysis. The unit must operate as a standalone device fitting within a standard biosafety cabinet with maximum dimensions of 20 inches in any direction.
The equipment must accommodate tissue samples up to 2 centimeters in diameter and 2 centimeters in length on a fully stabilized sample platform. Required performance specifications include automated sectioning capability, blade oscillation of at least 1 millimeter, adjustable blade advance speed up to 4 millimeters per second, minimal z-axis deflection near zero, and adjustable thickness settings ranging from less than 10 micrometers to 750 micrometers. The vibratome requires 110V AC power and must be capable of complete decontamination and cleaning between uses to maintain biosafety compliance.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Attachment 1_Solicitation Terms and Conditions_1232SA26Q0338.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Attachment 2 - Specifications
Vibratome specs
Salient Characteristics
The vibratome must be able to cut a variety of unfixed, live tissues with minimal distortion of natural conformation and morphology, no tearing, and minimized mechanical artifacts that could negatively impact downstream results. A larger blade oscillation results in cleaner cuts. The unit should be stand alone and able to fit within a standard biosafety cabinet (no more than 20 inches in any dimension).
Specifications Capacity:
Must be able to accommodate tissues up to 2cm in diameter and 2cm in length.
Fully stabilized sample platform
Utilities:
110V AC power connection
Performance:
Automated sectioning Blade oscillation of at least 1mm.
Adjustable blade advance speed up to 4mm/s Minimal, near 0 z-axis deflection Adjustable thickness from less than 10um to 750um.
Able to be decontaminated/fully cleaned between uses.
File details come from the government source that posted it. Updated .