Attachment 2 Salient Characteristics Spreadsheet .xlsx

XLSX spreadsheet 14 KB Posted

Attached to
Brand-Name or Equal: Thermo Scientific Orbitrap Exploris 480 Basic System 2/IC Federal contract opportunity
Solicitation number
75N95024Q00173
Issued by
Department of Health and Human Services National Institutes of Health National Institute on Drug Abuse

About this file

This document is a Combined Synopsis/Solicitation for a brand name or equal Thermo Fisher Mass Spectrometer-Liquid Chromatography system for the National Center for Advancing Translational Sciences (NCATS) at the National Institutes of Health (NIH). The solicitation number is 75N95024Q00173, and responses are due by 11:00 am EST on May 17, 2024. The anticipated award date is May 27, 2024. This is a full and open competition with no set-aside restrictions under NAICS code 334516. The government intends to award a purchase order based on an evaluation of technical approach, past performance, and price, with technical and past performance being significantly more important than price. The solicitation includes a detailed list of salient characteristics the system must meet, such as mass resolution, scan speed, mass accuracy, and dynamic range. Respondents proposing an equal product must provide documentation demonstrating how their product meets the salient characteristics. Installation, training, and optional extended warranty are also required.

View the file

Other files for this federal contract opportunity

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Sheet1

Solicitation # 75N95024Q00173
Salient CharacteristicsMeets Salinet Characteristic (Y/N)Please Identify where in your submission this criteria is met (Page and Section)Additional Notes
SENSITIVITY
1MS/MS: 50 fg reserpine on column yields S/N 100:1 or better
2SIM: 50 fg reserpine on column yields S/N 150:1 or better
Isolation and Fragmentation
3The instrument must be able to isolate ions stepless between
40.4 Da and full mass range (between m/z 50 and m/z 400)
50.7 Da and full mass range (between m/z 400 and m/z 700)
61 Da and full mass range (between m/z 700 and m/z 1000)
71.5 Da and full mass range (between m/z 1000 and m/z 1500)
82 Da and full mass range (between m/z 1500 and m/z 2000)
93 Da and full mass range (between m/z 2000 and 2500)
10Must have a straight ion routing multipole with axial field for Higher Energy Collisional Dissociation (HCD).
COLLISIONAL DISSOCIATION (HCD)
1Mass Resolution
2Maximum resolution of 480,000 (FWHM) at m/z 200 for all scan modes and both polarities.
3Minimum resolution of 7,500 (FWHM) at m/z 200 for all scan modes and both polarities.
SCAN SPEED
1Highest scan speed is 40 scans per second (R=7,500 at m/z 200) in FullScan or SIM mode
2Scan speed for MS/MS is up to 40 MS/MS scans per sec at high resolution (R=7,500 at m/z 200) in targeted or data dependent mode.
MASS ACCURACY
1< 3 ppm RMS – with external calibration over 24 h and without any kind of mass recalibration1
2< 1 ppm RMS – with internal calibration
DYNAMIC RANGE
15000:1 within one spectrum (single transient acquisition)
MASS RANGE
1The mass spectrometer must operate in the mass range of 50 – 6000 Da
SCAN RANGE
1The scan range must be at least: last mass = 15 x first mass
POLARITY SWITCHING
1The mass spectrometer must be capable of fast polarity switching acquiring one spectrum in positive and one in negative ionization mode (one full cycle) at > 1.4 Hz at a resolution of 60,000 (@ m/z 200). In polarity switching mode the instrument must perform within the defined mass accuracy specifications above.
SCAN FUNCTIONS
1The mass spectrometer instrument set up must allow scheduled set up of different events using a graphical user interface
2Must acquire and display Full Scan mass spectra
3Must acquire and display Selected Ion Monitoring (SIM) scan data for monitoring selected ions for target compound analysis
4Must acquire and display Full Scan MS/MS spectra
5Must acquire and display Selected Reaction Monitoring/Multiple Reaction Monitoring (SRM/MRM) like data sets Parallel Reaction Monitoring PRM
6Must acquire and display multiplexed SIM and multiplexed MS/MS mass spectra of up to 20 simultanously detected precursor ions.
7All-Ion-Fragmentation” mode, fragmenting all ions of a defined mass range with high resolution accurate mass acquisition
8Neutral-loss triggered data dependent MS/MS methodology, selecting predominantly precursor ions for MS/MS fragmentation, which show a predefined neutral loss under fragmenting conditions.
9Timed SIM for scheduled data acquisition of target compounds
10Timed MS/MS for scheduled data acquisition of target compound fragment spectra
11Must acquire “Data Independent Acquisition” (DIA) data sets
12Must acquire Boxcar Scan in DDA and DIA mode
13Must acquire Turbo TMT scan and SDM technology
ADVANCED FUNCTIONS
1Must have Automatic Gain ControlTM (AGC) to automatically control ion injection time for optimized ion current across the entire chromatographic peak
2The instrument must have a Predictive AGC algorithm to determine the ion injection time without carrying out a prescan event to increase the instrument scan speed/decrease the instrument cycle time.
3Must have Dynamic ExclusionTM capability to temporarily exclude an ion, while allowing MS/MS spectra acquisition of minor abundant precursors;
4Must have Normalized Collision EnergyTM capability to automatically compensate for the mass dependent energy deposition and provide instrument-to-instrument reproducibility
5Must have Stepped Normalized Collision Energy to allow multiple activation energies to be applied to the precursor before detection product ions.
6Must have an Apex triggering algorithm, which enables the user to trigger data dependent events close to the Apex of an eluting chromatographic peak.
7Must have Dynamic Retention Time window correction for PRTC peptides in tMS2 (PRM) scan function
8On the fly charge state deconvolution for intelligent ddMS2 on intact proteins applying Smart HCD in the ion routing multipole
HARDWARE
TRANSFER ION OPTICS
1Bent square RF only quadrupole device with axial field for most robust operation and highest transmission
2The transfer ion optics must be designed for high stability and ion transmission efficiency
3Must have a gas-free multipole ion transfer optics
4Must have gas filled C-Trap as intermediate storage device
5Must have segmented Hyperbolic Quad mass filter, providing variable and step-less precursor isolation width selection from 0.4 Da to 1960 Da
VACUUM SYSTEM
1The vacuum system must have no more than one turbomolecular pumps
2The mass spectrometer must have a differentially pumped vacuum system with pressures ranging from atmospheric pressure to 5E-10 mBar in the orbitrap analyzer
3The mass spectrometer must have a high vacuum aluminum manifold
Detection System
4The detection system must have digital electronic noise discrimination
ORBITAL TRAP
1Must have an electrostatic orbital trap
2Must have very low electronic capacitance for high ion signal to noise
Additional Items
13 days LSMS Onsite Support
2Trade-In Serve Deinstallation Labor
Extended Warranty
1Extended Warranty Exploris 480 – Product Number 701-056936
2Vanq Horizon Split Sampl - Product Number 701-054211
3Vanq Horizon Column Cop - Product Number 701-054215
4Vanq Horizon Binary Pump – Product Number 701-054209
Warranty Characteristics
1Pre-scheduled, on-site preventive maintenance (PM) and PM parts kit

(1/system/year)

2Priority on-site response time target within 3 business days
3On-site corrective maintenance (Includes factory-certified parts, labor, and travel)
4Unlimited access to remote support engineers through our latest digital

and augmented reality tools

5Priority remote diagnosis and repair, when possible.
6Software and firmware updates during preventive and corrective maintenance visits, upon request
7Replacement of malfunctioning computers purchased from Thermo Fisher Scientific and required for operation of the instrument
8Priority status technical support with targeted immediate phone response (Mon- Friday business hours)
Submitted by: (Name of Quoter)
Date:

File details come from the government source that posted it. Updated .