Nitric Oxide delivery system SALIENT CHARACTERISTICS
Nitric Oxide delivery system: Equipment is required for severely critically ill infants.
Characteristics:
| Unit must have FDA clearance |
| Contractor must provide 2 units in house |
Contractor shall supply equipment needed to run the device
| Nitric Oxide Gas tanks will be house in facility and switched out quarterly by vendor |
| User Friendly interface |
| Patient safety features and alarms designed to help mitigate and manage the risks of device-related rebound pulmonary hypertension. |
| Validated with over 70 ventilators and modes of ventilation. |
| Delivery systems includes the ability to use in the MRI suite. |
System must have the ability to effectively transport neonates for seamless delivery of Nitric Oxide therapy.
| Be validated with non-invasive systems; |
| Babi.Plus Bubble PAP Valve and Pressure Relief Manifold |
| Fisher & Paykel Healthcare Bubble CPAP |
| Fisher & Paykel Healthcare RT329 Breathing Circuit & BC Nasal Cannula |
Fisher & Paykel Healthcare RT202 Breathing Circuit & Optiflow™ Adult Interface Standard Neonatal Nasal Cannula
| Vapotherm® Precision Flow® |
| Vyaire™ Infant Flow® Nasal CPAP |
Vyaire™ Infant Flow® SiPAP™ Validated with mechanical ventilation systems, including high-frequency oscillatory and high- frequency jet systems