Attachment 1A. Salient Characteristics (COVID Vaccine – Freezers).pdf
PDF 550 KB Posted
- Attached to
- Covid Vaccine - Freezers Federal contract opportunity
- Solicitation number
- FA524021S0002
About this file
This document includes an attachment with specifications for freezers required for a Covid vaccine opportunity and the related federal contract solicitation. The attachment provides details on the quantities, capacities, and features of two models of medical freezers including interior and exterior dimensions, voltage requirements, temperature ranges, certification standards, and material specifications. The solicitation is for Covid Vaccine - Freezers through the Department of the Air Force Pacific Air Forces. Interested parties should review the attachments for requirements and specifications in order to respond to this pre-solicitation notice for medical freezer equipment to support Covid vaccine storage and distribution.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Sources Sought - COVID Vaccine - Freezers.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Qty 63 Qty 14
451 / 434 747 / 720
Vials 30000 (2ml) 50000 (2ml)
Cryoboxes 345 (model H50) 575 (model H50)
Plasma Bags 300 (350ml) 500 (350ml)
1 h 18 min. 1 h 18 min.
1988 x 699 x 1039 1988 x 992 x 1039
1370 x 447 x 713 1370 x 740 x 713
250 297
230 - 220 230 - 220
50 - 60 50 - 60
2000 - 2000 2000 - 2000
9.1 - 12.5 11.9 -13.5
326 - 405 426 - 511
46 - 44 50 - 55
49 - 49 48 - 48
Inner body / door
Outer body / door
Separate interiod doors
Shelf
7’’ full touchscreen display
Ergonomic handle open/close - heavy duty door lock mechanism
Integrated pressure release valve
Separate interior doors in order to minimize the loss of cold air - Qty 2
ST-Shelf - Qty 3
Ethernet interface
Potential-free alarm contact in case of system failure
Integrated port for external sensor
Smooth castors with stabilizers - Qty 4
Door hinge Right | Left cCSAus certificate
GMP - Clean room classification
Freezers
General Equipment
Door insulation (polyurethane)
Cabinet insulation (polyurethane)
European Medical Device Regulation
EMC directive
Low voltage directive
FDA regulation | Medical Device
Compressor Running Time (%)
Noise Level (dB(A))
(at 1m height & 1m distance)
Regualtion Sensor
Display Sensor in reference body / 100ml DOW corning 200-5CST (Silicon Oil)
Accu data / fucntion time of the control panel when system faillure
Relative humidity at +32°C
G e n e r a l
F e a t u r e s Net weight with standard equipment (kg)
Supply Voltage (V)
Hold over time (-80°C to -60°C)
External dimensions H x W x D (mm)
Inner dimensions H x W x D (mm)
Gross / Net volume (l)
Class ISO 5 / EC GMP A
E n e r g
S a v e r
S p e c i f i c a t i o n s
Frequency (Hz)
Power (W)
Energy Consumption (kWh/24h)
Heat Emission (Kcal/h)
Stainless steel (1.4301)
MDR (EU) 2017/745, Class IIa (models: 230 V - 50 Hz)
2014 / 30 / EU (models: 230 V - 50 Hz)
2014 / 35 / EU (models: 230 V - 50 Hz)
21CFR Part 864.9700, Class II (models: 220 V - 60 Hz)
Yes (models: 220 V - 60 Hz)
12 V - 7 AH / 48h
≤ 75%
65-82 mm PU with 20 mm VIP
77-82 mm PU with 20-30 mm VIP
Stainless steel (1.4301)
DX51D + Z100 coated Material
Anodized aluminium
Storage Capacity
PT1000, from -80°C to -180°C at 0°C ± 0.1°C
PT1000, from -80°C to -180°C at 0°C ± 0.1°C
File details come from the government source that posted it. Updated .