Attachment 1 Testing Procedure for Cleaning Validation Study.docx

DOCX document 16 KB Posted

Attached to
Formlabs Surgical Guide Resin Cleaning Validation Federal contract opportunity
Solicitation number
36C25226Q0713
Issued by
Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 12

About this file

This is a Testing Procedure document that outlines the standardized methodology for conducting a cleaning validation study on Formlabs Surgical Guide Resin. The procedure establishes the specific testing parameters and protocols that the contractor must follow to validate the effectiveness of cleaning processes for the resin material used in surgical guide manufacturing. The document serves as the technical specification and quality assurance framework for the one-time cleaning validation service required by the Department of Veterans Affairs Veterans Health Administration Veterans Integrated Service Network 12 for the Clement J Zablocki VA Medical Center in Milwaukee, Wisconsin.

The testing procedure defines the step-by-step process, acceptance criteria, and measurement standards that must be applied during the validation study. The contractor performing the cleaning validation study must execute the procedures exactly as specified in the document and complete all testing activities within the 120-day performance period following contract award. This document functions as a critical attachment to the solicitation that ensures the contractor understands the precise technical requirements and testing methodology necessary to successfully complete the cleaning validation for the Formlabs Surgical Guide Resin, ultimately supporting the VA's quality and safety standards for surgical guide manufacturing.

View the file

Other files for this federal contract opportunity

Other files attached to Formlabs Surgical Guide Resin Cleaning Validation, newest first.
File Type Posted
S02 - Solicitation_36C25226Q0713.docx DOCX document

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Attachment 1: Testing Procedure for Cleaning Validation Study

· Prepare ENZOL® solution per manufacturer’s instructions and immerse dental guides in the solution for one minute.

· Agitate dental guides in solution so all air bubbles are removed and all surfaces are in contact with the solution. Use small syringe or brush to flush solution into small areas if needed.

· While immersed, use a lint-free wipe or a soft bristled brush to gently wipe or brush all surfaces for one minute, including the small openings.

· Remove dental guides from ENZOL® solution.

· Rinse each dental guide under lukewarm running tap water for a minimum of one minute.

· After the rinse, use a syringe filled with lukewarm tap water to aspirate water through all channels, slots, small openings, or crevices to remove ENZOL® solution.

· Soak dental guide in DI water for a minimum of one minute.

· Agitate dental guide for 30 seconds in the DI water to ensure complete rinsing.

· Repeat two more times.

· Wipe dry with lint-free wipe.