The document is a detailed technical specification sheet for the Boonton Model 4500C, a high-precision RF measurement device with advanced capabilities. The system offers comprehensive sensor inputs with a frequency range of 30 MHz to 40 GHz, measurement ranges from -60 to +20 dBm, and a video bandwidth of 125 MHz. Key features include Real-Time Power Processing, ultrafast trace acquisition, automated power and time measurements, and multiple display and statistical analysis modes.
The technical specifications highlight the device's sophisticated measurement techniques, including random interleaved and real-time sampling with a maximum sampling rate of 100 MS/s across four channels simultaneously. Optional statistical acquisition modes provide capabilities like Cumulative Distribution Function (CDF), Complementary Cumulative Distribution Function (CCDF), and various display options. The system includes a multi-level calibrator, precise time base with 100 ps resolution, and flexible trigger and pulse measurement capabilities, making it suitable for complex RF and pulse analysis applications.
View the file
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Key Specifications:
Frequency range 30 MHz to 40 GHz
Measurement range -60 dBm to +20 dBm
RF channel video bandwidth 125 MHz
RF channel rise-time < 5 ns
Overall accuracy 0.2 dB
Time Resolution 100 ys
Min Pulse Width / Max PRF 6 ns / 50 MHz
Channels 2 RF and 2 trigger
Key Features
- Real-Time Power Processing™
- Ultrafast trace acquisition and refresh rate
- O ne button push for 15 automated power and time measurements
- Automatic peak-to-peak, delay-by-time and delay-by-events triggering
- Multi-level, multi-function calibrator
- C continuous statistical analysis (optional)
- Includes gated PDF, CDF, and CCDF
- Displays up to 4 measurement, 2 memory, and 1 math channel simultaneously
Specifications detail:
Sensor Inputs
RF Frequency Range 30 MHz to 40 GHz
Pulse Measurement Range -50 to +20 dBm
Modulated Measurement Range -60 to +20 dBm
Relative Offset Range }100.00 dB
Logarithmic Vertical Scale
0.1 to 50 dBm/div in 1-2-5 sequence
0.1 to 50 dBV/div in 1-2-5 sequence
0.1 to 50 dBmV/div in 1-2-5 sequence
0.1 to 50 dBuV/div in 1-2-5 sequence
Linear Vertical Scale
1 nW/div to 50 MW/div in 1-2-5 sequence
1 mV/div to 50 kV/div in 1-2-5 sequence
Video Bandwidth 125 MHz
Rise-Time <5 ns
Single-Shot Bandwidth 35 MHz
Pulse Repetition Rate 50 MHz max
Minimum Pulse Width 6 ns
Time Base
Time Base Range 5 ns/div to 1 hr/div
Time Base Accuracy +/- 4 ppm peak
Time Base Resolution 100 ps
Time Base Display Sweeping or roll mode
Statistical X-Axis (optional)
Scale Linear or logarithmic, 1 to 7 cycles
Linear Ranges 0.1%/div to 10%/div
Linear Offset 0 to 99.9%, 0.1% resolution
Log Range 1e-9% to 100%
Calibration Source
Operating Modes CW, internal or external pulse
Frequency 1.024 GHz } 0.01%
Level Range -50 to +20 dBm
Resolution 0.1 dB
Output VSWR 1.10 maximum
Absolute Accuracy }0.04 dB ( }0.9%) at 0 dBm
Accuracy vs Level add }0.02 dB per 5 dB increment from 0 dBm
Preset Internal Pulse Period 0.1 or 1 or 10 ms
Preset Internal Pulse Duty Cycle 10% to 90% in 10% increments
Variable Pulse On Time 7 μsec to 65.535 ms in 1 μsec steps
Variable Pulse Period 28 μsec to 131.070 ms in 2 μsec steps Off-time limits - within 7 μsec to 65.535 ms
Pulse Polarity + or –
RF Connector Precision Type N
External Pulse Input Rear panel BNC, TTL level compatible
Auto-Calibration
The calibrator is used to automatically generate linearity calibration data for peak power sensors.
Measurement System
Sensor Inputs Two peak power sensor measurement channels and two trigger/oscilloscope channels.
Measurement Technique
Random interleaved and real-time sampling system that provides pre- and post- trigger data, as well as statistical histogram accumulation.
Maximum Sampling Rate 100 MS/s on up to four channels simultaneously.
(Equivalent effective sampling rate of 10 GS/s with Random Interleaved Sampling)
Waveform Averaging 1 to 16,384 samples per data point (time domain measurement)
Number of Histogram Bins 16,384
Size of Sample Bins 32-bits (4,096 mega-samples)
Bin Power Resolution <0.02 dB
Statistical Acquisition (optional)
Modes Continuous or gated by pulse mode time markers
Sampling Rate 100 MS/s on all channels simultaneously
Limit Count Adjustable, 2 to 4096 Megasamples
Limit Time 3600 seconds (approx. 41 s at full sample rate)
Terminal Action Stop, flush and restart or decimate
System Displays
Display Type
Power versus time (pulse mode), Power versus time (modulated mode), External trigger versus time (pulse mode), Auto-measure text
(all modes), Help text (all modes), Reports (sensors, configuration, calibrator, files, stored waveforms, GPIB commands, GPIB buffers)
Statistical Display Type (optional)
Cumulative Distribution Function (CDF), Complementary Cumulative
Distribution Function (CCDF), Split screen, gated CCDF and power versus time (pulse mode,CCDF), Distribution function (histogram), External trigger statistical (statistical mode), Auto-measure text
(statistical mode)