REVISED 29 March 2023

STATEMENT OF WORK

SPRING TESTER

1. SCOPE. This Statement of Work (SOW) defines the requirement for one (1) hardware and software upgrade of an existing automatic spring tester at Anniston Army Depot (ANAD), Anniston, AL, and one (1) new automatic spring tester. The spring tester upgrade and new spring tester shall meet the requirements herein.

2. APPLICABLE DOCUMENTS. There are no documents referenced herein.

3. REQUIREMENTS

3.1 Upgrade of Existing Spring Tester. The existing Larson Systems Inc. FLASH 11

Model No. 050-0000-0121-00B SN 2775012608 automatic spring tester shall be upgraded to the latest shipping revision and/or version. The upgrade shall include the following.

3.1.1 Standard Service / Calibration. The existing tester shall be serviced and calibrated off-site at the Contractor’s facility. Service shall include cleaning, performance of any OEM recommended service and evaluation of the tester. Any repairs shall be quoted separately, and no repairs shall be performed without written authorization from the Contract Officer.

Calibration shall include the tester and up to two (2) load cells. A calibration certificate shall be provided to ANAD.

3.1.2 Software Update. The existing tester’s FLASH testing software shall be updated to the latest shipping version. Any test procedures, settings/preferences, and results unique to the tester shall be retained after the software update. Retaining this data will allow the tester to be put back into our existing production process without disruption. This requirement does not apply to data that is corrupted or otherwise unreadable due to the condition of the existing hard drive.

3.1.3 Windows 10 Upgrade. The existing tester’s operating system shall be updated to

Windows 10. Windows 10 shall be fully activated and licensed to ANAD. Tester hardware shall be replaced as required to be compatible with the new operating system.

3.2 New Spring Tester. A new automatic spring tester shall be provided to ANAD. The tester shall be new and the latest shipping revision and/or version. The tester shall be a tabletop model with a 1,000 lb. servo drive motor, Windows 10 touch screen interface, quick interchangeable load cells and clear guarding. In addition, the tester shall meet the following requirements.

3.2.1 Compatibility. The new tester shall be compatible with the existing Larson Systems

Inc. FLASH 11 Model No. 050-0000-0121-00B automatic spring tester after that tester is upgraded per Paragraph 3.1. Spare parts and test software specification files shall be interchangeable between the new and existing testers. Spring tester operation and graphical user interface (GUI) shall be the same as the existing spring tester after upgrade. The purpose of compatibility is to lower the new spring tester acquisition cost, reduce overall maintenance and operational cost of the spring testers, and provide redundant spring testing capability.

3.2.2 Testing Capability. Testing shall include measurement of spring rate and load at height, comparison of these measurements to a specific range and visually indicating pass if the measurement is withing the range and fail if it is not. These measurements shall have an error of

0.5% of reading or less. The new spring tester shall include everything required to test the following compression springs.

Spring #

Spring Characteristics Spring Requirements

Free Height ID OD Rate Height at

Load Load

1 0.688 ~0.265 0.5625

Maximum ? 0.625 140 to 160

2 1.245 to 1.255 ~0.338 ~0.470

44.1 to

45.9

0.984 12

3 ~1.667 ~0.194 ~0.435 ? ? ?

4 1.577 min ~0.334 ~0.484 ? 1.11

32.7 +/-

0.9

2.000 nominal

1.9530 to 2.0470

0.525 nominal

~0.808 ? 1.80 50 to 55

1.5150 to

1.6090 1.178 nominal

1.391 to 1.421 ? 1.40

6.00 to 6.50

7 ~1.935 ~0.608 ~0.928 ?

1.848 1.592

127 to 153 192 to 224

0.867 nominal

0.780 to 0.954

~0.113 0.183 to 0.191 ? 0.760 5.3 to 5.7

9 2.18 to 2.66 ~1.018 ~1.432 ? 2.014

64.0 to

71.6

10 1.400 nominal ~0.200 ~0.288 ? ? ?

11 ~1.233 ~0.947 ~1.116 ? ? ?

12 ~1.483 ~0.419 ~0.530 ? ? ?

1.431 nominal

1.420 min 0.330 nominal

0.480 nominal ? 1.11

22.5 +/- 0.5

2.419 nominal

2.500 to 2.558

0.676 to 0.686

0.983 nominal ?

2.105 1.830

54 to 65.4

96.2 to 106.2

2.548 nominal

2.18 to 2.66 ~1.064 1.388 to 1.404 ? 2.014

64.0 to 71.6

16 1.400 nominal ~0.200 ~0.288 ? ? ?

17 1.89 nominal 0.88 nominal

1.13 nominal 90 1.56 25 to 35

3.2.3 Load Cells. The tester shall utilize quick interchangeable load cells. The load cell(s) shall have capacity to test the springs described in this SOW. Testing of these springs shall be accomplished using two or less load cell(s).

3.2.4 Accessories. A thru-rod kit suitable for the springs described herein and a dust cover shall be provided with the tester.

3.3 Shipping. ANAD will package and ship the existing spring tester to the service facility at the Governments expense. When the new tester arrives at ANAD, its shipping crate will be reused to ship the existing tester to the Contractor facility. The Contractor shall inform ANAD when they are ready to receive the existing tester, so that the time required to upgrade it is minimized. Return shipping of the upgraded tester and any accessories and shipping of the new tester and accessories to ANAD shall be at the Contractor’s expense and included in the overall contract cost.

3.4 Documentation. Documentation shall be provided with the upgraded existing spring tester and new spring tester. The documentation shall be that which the Contractor customarily provides to its customers.

3.5 Delivery. Contractor shall ship the new spring tester to ANAD within four (4) weeks of contact award date. Contractor shall ship the upgraded existing spring tester to ANAD within six

(6) weeks of receipt at their facility.

3.6 Installation / Training. Onsite installation of the new spring tester only and training shall be quoted as an option to the Government. Travel and meal expenses shall be included in the quoted price. The installation and training shall be of a duration that is normally provided by the Contractor to its customers. Installation and training shall not be scheduled or performed without written approval of the Contracting Officer.

3.6 Warranty. The upgraded spring tester and new spring tester shall include the

Contractor’s standard commercial warranty and terms.

4.0 SCOPE CHANGE. The Contractor shall not proceed with any verbal or written change(s) to the scope of work from sources other than the Contracting Officer, who will provide changes in writing.