These specifications detail requirements for a Walk In Plant Growth Chamber designed for cultivating plants under precisely controlled environmental conditions. The chamber must provide a minimum of 100 sq. ft. of shelf area with 36 inches of vertical growth space, featuring advanced environmental control systems including LED lighting at 1200 µmoles/m2/s with dimmable intensity, temperature control ranging from 10-44°C, a remote air-cooled condensing unit, and a humidifier capable of maintaining relative humidity up to 85%. Additional specifications include an onboard programmable digital control system for parameter monitoring and installation by the manufacturer's factory team, ensuring comprehensive environmental regulation for scientific plant cultivation.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFQ_1232SA25Q0982 Terms and Conditions.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications
Instrument: Walk In Plant Growth Chamber
For cultivation of plants on under controlled environmental conditions (temperature, humidity, light intensity, photoperiod length).
Requirements:
1. Shelf area: minimum total shelf area of 100 sq. ft.
2. Vertical space for plant growth height: 36 inches.
3. Lighting: LED lighting, 1200 µmoles/m2/s at 100% intensity, dimmable from 10-100% at 1% intervals.
4. Temperature control: set point of 10 - 44°C
5. Refrigeration: remote air-cooled condensing unit.
6. Humification: humidifier with relative humidity sensor required, set point up to 85%
RH.
7. Controls: On board programmable digital control and parameter monitoring.
8. Installation by factory installation team