This document is a set of specifications for a growth chamber to be used in conducting experiments on legume plants and bacterial organisms at the Yurgel lab in Pullman, Washington. The key requirements include temperature ranges, multi-tier adjustable shelving, broad spectrum LED lighting with user controls, uniform refrigeration, and an intuitive touchscreen interface with alarm notifications. The chamber must fit through a standard commercial door, have an interior volume of 27.6 ft3, and not exceed 120V electrical requirements.
The specifications are part of a Federal contract opportunity issued by the Department of Agriculture's Agricultural Research Service, Pacific West Area, as a Request for Quotation (RFQ) for a "Growth Chamber for Pullman, WA". The solicitation is unrestricted with a NAICS code of 335416 and a small business size standard of 1,000 employees. Offers are due by July 30, 2024 at 1:00 pm Pacific Time, and the delivery location is the PWA Grain Legume Genetics Physiology office in Pullman, WA. The contract terms include FAR Clause 52.212-4 and FAR Clause 52.212-5.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combined Syn-SOL 12905B24Q0173.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Growth Chamber for Yurgel lab in Pullman: Specifications
Extended Description of Product: This product is for conducting experiments on legume plants and bacterial organisms. Multiple experiments using different containers and varying conditions will be conducted, requiring a chamber with adjustable features that are user friendly. The product must have the following attribute and capabilities:
• Temperature must go to 2°C lights off and 5°C lights on
• Multi-tier shelving that is adjustable
• Broad spectrum lighting with ability to be user adjusted
• LED lighting
• Energy saving lighting
• Uniform refrigeration conditions
• Intuitive touchscreen for controlling and monitoring conditions
• Audio, visual and email notifications of alarms
• Fit through standard commercial door
• Chamber must be reach-in
• Chamber must include observational window
• Chambers electrical requirements must not exceed 120 V
• Chamber exterior dimensions should be no larger than 41 x 32 ½ x 77 (inches)
• Interior volume must be 27.6ft3
• Extended 5-year warranty option