This document is a Specifications attachment for a federal contract opportunity related to the procurement of an analytical balance. The key requirements are: the balance must have a readability of 0.01 mg, a capacity of at least 120 g with 0.1 mg accuracy and 41 g with 0.01 mg accuracy, seamless integration with an existing DSC3 instrument, anti-static protection, indicator lights for balance status and issues, a leveling indicator, and the ability to detect static on samples.
The associated federal contract opportunity is a Request for Quotation (RFQ) solicitation for a "Balance XPR105DR with EQPac and Extended Care" for delivery to Albany, CA. The solicitation number is 12905B24Q0140 and it is an unrestricted competition under NAICS code 334516 with a small business size standard of 1,000 employees. Offers are due by July 19, 2024 at 1:00pm Pacific Time.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| Combined Syn-SOL 12905B24Q0140.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Specifications Project: Analytical Balance
Page 1 of 1 ANALYTICAL BALANCE | SPECIFICATIONS, June 24, 2024`
Recommended Instrument:
Mettler-Toledo Balance XPR105DR with EQPac Comprehensive Qualification and Extended Care Packages
Must meet the following specifications:
1. Must have readability of 0.01 mg (5 decimal place metric balance that displays and measures in 0.01 mg increments.
2. Must have at least a 120 g capacity with accuracy to 0.1 mg.
3. Must have at least a 41 g capacity with accuracy to 0.01 mg.
4. Must have seamless connection between the balance and the existing Mettler-Toledo
DSC3 allowing transfer of weight data, including pan weight (tare) and sample+pan weight data to reduce or eliminate operator error when entering data.
5. Must include installation and integration with existing DSC3 instrument and operator training.
6. Must have an anti-static protection device inside balance chamber to eliminate weight shifts due to static.
7. Must have an indicator light that shows when the balance is ready for use.
8. Must have an indicator light that shows when the balance has an issue or needs service.
9. Must have a leveling indicator for leveling the balance that also shows how to correct the level on the balance.
10. Must have an indicator for detecting static on the sample and or the container.
11. Must have available factory-trained field service technicians that have diagnostic tools for installations, preventive maintenance and repairs.