Item Description:
Capillary Gel Electrophoresis Unit for Protein Analysis (Instrument & Software) – Abby System - Brand
Name Or Equal Product.
Product Specifications:
• Bench-top footprint with dimensions not exceeding 36 cm H x 36 cm W x 57 cm D
• Weight not to exceed 23 kg
• Fully operable requiring only a single universal voltage (100VAC-240VAC, 50/60 Hz) power output connection
• Fully operable in a humidity range of 20-60% relative, non-condensing
• Fully operable in a temperature range of 18-24oC
• Processes a maximum of 25 samples per run
• Run time <3 hours, <5 with RePlex
• Uses capillaries with a length of 5 cm
• Uses capillaries with a volume capacity of 400 nL
• Uses capillaries with a 100 μM inner diameter
Sample Input Specifications
• Process a maximum of 25 samples per run
• Store samples at room temperature
• Analyze the following samples: protein lysates from cultured mammalian cell lines, white blood cells, FACS-sorted cells, plasma, and crude cell culture supernatants
Reaction Vessel Specifications
• Fused silica capillaries that are 5 cm in length, 100 μM inner diameter and 400 nL volume capacity
• Perform protein capture using proprietary inner capillary chemistry coating that is UV light activated to cause covalent immobilization of all proteins to the inner capillary wall
Throughput Specifications
• Process up to 25 individual samples in a microtiter plate with space for separation and stacking matrices, primary and secondary antibodies, and peroxide and luminol chemiluminescent substrate which are required to carry out 25 individual immunoassays.
• Complete 25 nano-immunoassays within a cycle time in three hours
• Complete 2 rounds of Immunoprobing and detection in a single capillary (2 serial immunoassays or immunoassay + total protein assay) in five hours using RePlex.
• Robotic automated system which prepares capillaries for sample uptake
• Robotic automated system which volumetrically loads 40 nL sample and ~360 nL of buffer
• Perform automated size-based separation
• Wash capillaries, remove matrices, and perform automated immunoassay steps incubating of primary and secondary antibodies and protein blocking steps
• Automatically introduce chemiluminescent/fluorescent reagents for immunoassay detection results
Separation Specifications
• Molecular weight size separation power supply able to produce 0-3,000 volts
• Constant voltage output
Immobilization Specifications
• Software to control the UV lamp
• Contains a low-pressure mercury lamp
Detection Specifications
• Automated detection of fluorescently labeled protein size standards
• Molecular weight ladder ranges from 2 to 440 kDa
• Ability to distinctly resolve biotinylated protein standards
• Chemiluminescent detector
• CCD Camera that is cooled and with 1.4 megapixel and 16-bit resolution
Software Specifications
• Compass software to set-up assay protocols and analyze run data
• Specify up to 25 samples per run
• Specify primary and secondary antibodies per sample
• Specify the molecular weight per sample
• Specify chemiluminescence collection time
• Perform relative quantitation of identified peaks by total area or peak height