About this file
This document appears to be the "Salient Characteristics" attachment for a federal solicitation for Ultraviolet (UV) Disinfecting Robots. The key details are:
The robots must meet the following requirements: operate on 120V electrical, be an FDA-authorized medical device, have a minimum pulse rate of 67Hz, utilize high-intensity broad spectrum UV light (240-315nm), include motion sensors to automatically stop the device when movement is detected, have four locking casters and two stainless steel push handles, provide cloud-based reporting of key metrics with individual user logins, require no warm-up or cool-down delays, incorporate applicable safety signage and features, have a maximum push weight of 5 lbs, and provide a minimum 1-year warranty and training. The robots must also be capable of deactivating 100% of VRE, 99.9% of MRSA, 99.9% of SARS-CoV-2, and 95% of C. Diff spores. The robots must meet specified safety standards such as ANSI/AAMI and IEC requirements.
The related federal contract opportunity is Solicitation #36C24624Q0871 issued by the Department of Veterans Affairs Veterans Integrated Service Network 6 for Ultraviolet Disinfecting Robots. The solicitation seeks to acquire these UV disinfecting robots per the salient characteristics provided.
View the file
Other files
| File | Type | Posted |
|---|---|---|
| RFQ 36C24624Q0871 UV DISINFECTING ROBOTS.pdf |
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
SALIENT CHARACTERISTICS
Mobile automated disinfec on robot(s)
120 Volt, Conven onal Outlet
FDA authorized medical device
Pule Rate 67Hz minimum
High Intensity light, intense broad spectrum UV light range (240 – 315 nm) minimum
Start bu on
Mo on Sensors stop the device when mo on is detected
Unit includes four locking casters
Two stainless steel push handles
Automa c lamp retrac on for transporta on, extended height 72” maximum
Cloud‐based repor ng of key metrics, provide monthly reports, individual user logins and passwords
No warm‐up or cool0down delays
Must incorporated signage on unit with all applicable safety features
Push weight maximum of 5 lbs
Minimum of 1 year warranty and service
Training
No network connec on required; cellular based repor ng
Must deac vate the environment of the following pathogens:
o VRE 100% o MRSA 99.9% o SARS‐ VoV‐2 99.9% o C Diff spores 95%
Safety standards o ANSI AAMI ES60601‐1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and
A2:2010/(R)2012 o IEC 62471 First Edi on 2006‐07 o IEC 60730‐1:2013, AMD1:2015,AMD2:2020