Attachment 1 - Minimum Specifications.pdf

PDF 77 KB Posted

Attached to
Direct shear and direct simple shear test system Federal contract opportunity
Solicitation number
1232SA25Q0152
Issued by
Department of Agriculture Agricultural Research Service

About this file

The document is an Attachment 1 - Minimum Specifications for a Fully Automated Direct Shear and Direct Simple Shear System. The specifications require a system with displacement rate control accuracy of 0.0000002 to 1.3 in/min, a shear frame capacity of at least 11 kN, and hardware modules for automated collection, processing, and reporting of Direct Shear and Direct Simple Shear tests. Additional requirements include NIST Traceable Calibration Certificates for all equipment and shipping to a project office in Stillwater, OK.

This specification is part of a small business set-aside solicitation (Number 1232SA25Q0152) issued by the USDA Agricultural Research Service. The solicitation is for one Direct Shear and Direct Simple Shear Test System, with quotes due by 06/19/2025 at 2 PM ET. Delivery is required no later than 09/30/2025, with the system to be delivered FOB Destination to the USDA ARS facility in Stillwater, OK. The associated NAICS code is 334516, with a small business size standard of 1,000 employees.

View the file

Other files for this federal contract opportunity

Other files attached to Direct shear and direct simple shear test system, newest first.
File Type Posted
Clauses and Provisions - 1232SA25Q0152.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Minimum Specifications

Salient characteristics –

Fully Automated Direct Shear and Direct Simple Shear System

Displacement rate control accuracy to meet or exceed 0.0000002 to 1.3 in/min (0.000006 to 33 mm/min).

Shear frame capacity to meet or exceed 11 kN

Hardware and modules for fully automated collection, processing, and reporting of Direct Shear and Direct Simple Shear tests

Calibration of all equipment with NIST Traceable Calibration Certificate

Shipping of all equipment to project office in Stillwater, OK.