Attachment 1 - Ion Chromatograph Specifications.pdf
PDF 445 KB Posted
- Attached to
- Ion Chromatograph Federal contract opportunity
- Solicitation number
- FA9101-24-Q-B051
About this file
This document contains the specifications for an Ion Chromatograph (IC) required by the Department of the Air Force Materiel Command Test Center (AFTC) at Arnold Air Force Base, Tennessee. The key requirements include a fully functional RFIC system with features such as eluent generation, electrolytically regenerated suppression, and high-pressure tolerance. The solicitation also specifies required components like an autosampler, column heater, trap column, and anion-exchange column. The procurement is being conducted through a Request for Quote (RFQ) with solicitation number FA9101-24-Q-B051. No refurbished or remanufactured parts are authorized for this requirement. The contractor will be responsible for providing all materials, labor, and equipment necessary to deliver the fully functional Ion Chromatograph system.
View the file
Other files for this federal contract opportunity
| File | Type | Posted |
|---|---|---|
| Attachment 2 - FA9101-24-Q-B051 Redacted - single_source_justification.pdf | ||
| FA9101-24-Q-B051 - Ion Chromatograph- Combined Synopsis Solicitation-Final.pdf | ||
| FA910124QB051 - Model Contract.pdf |
On GovTribe
Work with this file on GovTribe
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
IC Specifica ons
Specifications for Ion Chromatograph (IC) at Arnold Air Force Base, TN;
Arnold Engineering Development Complex (AEDC)
Provide all materials, labor, and equipment necessary to provide a new fully functional Ion Chromatograph. Refurbished or remanufactured parts are not authorized for this requirement.
1) Full Reagent-Free IC (RFIC) support including Eluent Generation and Electrolytically
Regenerated Suppression
- Integrated vacuum degas for in-line degassing of eluents
- Chemically inert, metal-free PEEK flow path
- High-pressure tolerant system (0-35 MPa)
- Support for continuously regenerated chemical suppression
- Serial dual-reciprocating piston pump
- 6-port, 2-position injection valve
- Conductivity detector and cell with thermal control and output range up to 18,000 µS/cm
- Eluent shut-off valve
- Leak sensor
2) AS-DV Autosampler
3) Eluent Monitor Kit
4) Thermo Scientific Dionex Inuvion Column Heater (or equivalent)
- Precise temperature control over a wide temperature range
- Heater supports 2mm to 5mm guard and separator columns
- Temp range 10 to 60 °C, minimum 5 °C above ambient
5) PEEK Viper Fittings Kit
6) Dionex CR-ATC 600 Trap Column (or equivalent)
- Continuously regenerated anion trip column
7) Dionex EGC 500 (Potassium Hydroxide) KOH Gererator Cartridge (or equivalent)
8) ADRS600, 2mm
9) Dionex CRD 200 (2mm) Carbonate Removal Device (or equivalent)
10) Dionex IonPac AS17-C is a hydroxide-selective anion-exchange column (or equivalent)
11) Dionex IonPac AG17-C guard column (or equivalent)
12) CM License
13) Service/installation
14) Chromeleon Single Instrument Controller, Class 1 instrument license
15) Power Cords, 4,
16) Assembly Sample Loop, 200UL
17) Year 2 and 3 IC/SP Parts only Warranty