Attachment 1 Brand Name or Equal Specifications.pdf

PDF 137 KB Posted

Attached to
Total Reflection X-Ray Fluorescence Spectrometer - USDA ARS Parlier, CA Federal contract opportunity
Solicitation number
12905B23Q0238
Issued by
Department of Agriculture Agricultural Research Service Field Research Implementation and Information Delivery Pacific West Area

About this file

This document contains an attachment specifying brand name or equal requirements for a total reflection x-ray fluorescence spectrometer procurement, along with details of a related federal contract opportunity posted on the Federal Business Opportunities website.

The attachment provides specifications for an x-ray fluorescence spectrometer capable of elemental detection from sodium to americium with detection limits below 10 parts per billion for most elements. Required capabilities include analysis of samples in various formats up to 90 discs or 50 wafers. The Bruker S4 TStar TXRF spectrometer is given as an example meeting requirements. The federal contract opportunity is posted by the USDA Agricultural Research Service seeking this type of spectrometer to be delivered and installed at their facility in Parlier, California, with associated training.

View the file

Other files for this federal contract opportunity

Other files attached to Total Reflection X-Ray Fluorescence Spectrometer - USDA ARS Parlier, CA, newest first.
File Type Posted
Combined Synopsis Solicitation 12905B23Q0238 RFQ.pdf PDF

On GovTribe

Work with this file on GovTribe

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Brand Name or Equal Bruker S4 TStar TXRF

SJVASC-CPQ needs an x-ray fluorescence spectrometer, for elemental detection and quantitation. The XRF needs to provide accurate and precise measurements of elements from sodium to americium with low detection limits (≤ 10 ppb for most elements). The Bruker S4 TStar total reflection x-ray fluorescence spectrometer is a compact bench top instrument, able to make sub-ppb level measurements for many elements (LOD < 1 pg for Ni) and is simple to operate. We have conducted market research and have concluded the Bruker S4 TStar is an example XRF that will meet our needs.

SPECIFICATIONS: Total Reflection X-Ray Fluorescence Spectrometer

• Element range: Na - Am.

• Detection Limit: < 1 pg for Ni

• Sample capacity: 90, 30 mm discs, or 50 2” wafers, or 30 microscopy slides

• Primary excitation source: 17.5 keV, 50 x 50 µm hot spot, micro focus x-ray tube with

Mo target.

• Secondary excitation source: 35 keV, 50 x 50 µm hot spot, micro focus x-ray tube with

W target.

• X-ray detector: Si drift detector, 60 mm2 active area, energy resolution 145 eV for Mn-K

@ 100 kcps, max. count rate > 100 kcps, dead time < 10 % at 40 kcps,

• Thermo-electrically cooled X-Ray source, no consumption of liquid nitrogen

• Nitrogen purge system

Additional requirements:

• FOB destination shipping

• Installation and training

• Advance user training and sample preparation (2 days on-site)