This document contains specifications for the procurement of five (5) Animal Change Stations designed to provide dual protection for both employees and laboratory animals during cage transfer operations. The units must protect employees from occupational exposure to animal allergens while protecting animals from the macro-environment, with ELISA-proven allergen containment systems serving as the primary safety function.
The specifications require units with dual access front faces on both sides to accommodate comfortable seating for two operators, one-piece solid stainless-steel work surfaces for decontamination, and frameless, shatter-proof sash. Units must be height-adjustable through button or touch screen controls, accommodate entrance heights of 51-78 inches with a work surface area of 43x18x22 inches, and feature a push bar for mobility. Additional required features include V-shaped and center grills to separate clean and dirty zones, large tray handles for easy cleaning, folding addition shelves for supplies, quality wheels with reduced noise and vibration, and a convenience outlet. The units must operate on 110 volts with 12 amps maximum full load (nominal running current of 2-4 amps with ECM blower motors), produce noise output under 53-55 dBA, and include UV lighting with button or touch screen control. Filtration requirements include ULPA filter exhaust and activated charcoal filters with vertical laminar downflow and inflow systems creating an air curtain to protect operators. Optional features include an alarm system for out-of-range conditions and a feed hopper. Biological safety cabinets are excluded due to annual certification requirements. Best value consideration will be given to units compatible with existing equipment to minimize training, maintenance, and downtime.
View the file
Other files
On GovTribe
Work with this file
- Download the original file
- Contacts named in this file
- Similar government files
- Ask GovTribe AI about this file
Text version
Animal Change Station Specifications Quantity: Five (5)
Must protect the employee for occupation exposer to animal allergens.
Must protect the animal from the macro-environment. For example, the dual protection model.
Must have dual access front face to the change station and in both sides, so both operators can sit down comfortably when doing cage transfer Farmless and Shatter proof sash.
Must have a one piece sold stainless-steel work area. This is for ease of decontamination of the surface area.
Must be adjustable for end-user ergonomic requirements. for example, a button or a touch screen to adjust the height.
Must have ELISA-proven allergen containment system. This protects the user from being exposed to the lab animal allergen, which is the primary function of dual access cage transfer unit.
Must have a push bar for ease of moving the change station.
Must have V-shaped Grill to avoid blocking and center Grill to separate Work zones into clean and dirty zones
Must have a Large Tray Handle for easy lifting to ease to clean the bottom of the unit.
Have a folding addition shelf for changing out the animal cages or holding additional supplies.
Must have good quality wheels to move around the facility and reduce sound and vibration.
Height requirements to enter the animal rooms 51 -78 height and work surface area 43x18x22
Must have Power supply 110 volts and 12 amps or lower Full Load Amps / Max Rating: ~10 A to 12 A for startup and receptacle loads, but Nominal running current for the Change Station withECM blower motors andcontrols is much lower, 2 Amps to 4 amps. Due to university lack of power for large appliance consumption.
Noise output under 53-55 dBA
Have a UV light and a button or on a touch screen
Have an optional feed hooper.
Must have a ULPA filter exhaust an activated charcoal filter and the combination of vertical laminar downflow and inflow system.
Ideally to have an alarm on the unit if out of range.
creates an air curtain to protect the operator from contaminants released from within the work zone.
Must have a convenience outlet of the unit.
No Biological safety Cabinet, due to required annual certification.
Preference, best value consideration for current unit compatibility due to training and maintenance for the end-user with little down time.