Attachment 1 - 26-01761 Reqs and Specs.pdf

PDF 177 KB Posted

Attached to
NB688000-26-01761 High Accuracy Laser Wavelength Meter
Solicitation number
NB688000-26-01761_High_Accuracy_Laser_Wavelength_Meter
Issued by

About this file

This is a Requirements/Specifications document for a federal procurement issued by the National Institute of Standards and Technology (NIST), Precision Measurement Laboratory, Time and Frequency Division (PML/688/80). The requesting organization requires one high accuracy laser wavelength meter capable of measuring laser frequency with precision at the 10 MHz level and simultaneously frequency stabilizing multiple lasers. The meter must have a measurement range from 330 nm to 1180 nm, with measurement accuracy of ±10 MHz or better over 375 nm to 1180 nm, and wavelength resolution of 200 kHz or better. The instrument must feature an integrated Photonic Crystal Fiber (endlessly single mode) 1x8 switch to accommodate up to eight simultaneous laser inputs without accuracy degradation, PID frequency stabilization capability for locking up to eight laser inputs simultaneously with programmable analog voltage control, USB control with graphical user interface software, and must maintain performance specifications across an air pressure range of 440-900 mm Hg and temperature range of 0-40°C.

Delivery of the goods, installation, and training must be completed no later than 14 weeks after receipt of order (ARO). Government acceptance is expected to take 21 days. The document indicates that base manufacturer warranty is acceptable, and training, installation, electronic media delivery, facility utility considerations, maintenance, and government-furnished material sections are marked as not applicable. The document does not specify pricing terms, contract vehicles (such as FSS, GWAC, or IDIQ), set-asides, or a response deadline, as these would typically appear in the associated solicitation rather than in this requirements specification template.

View the file

Other files

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

R

Requirements / Specifications: High accuracy laser wavelength meter 1 Version 1.0 (2019) Double-click on for further assistance.

Use this document to detail specifications/requirements for an item and/or ancillary services.

TITLE: High accuracy laser wavelength meter Requesting Lab/Division/Group: PML/Time and Frequency Division (688/80)

General Statement of Need We require a wavelength meter for measuring laser frequency with an accuracy at the 10 MHz level. The wavemeter must also be capable of frequency stabilizing multiple lasers simultaneously.

Items and/or Ancillary Services

Item 1: Laser wavelength meter Quantity: 1 Specifications: Spec 1.1: Measurement range from 330 nm to 1180 nm; Spec 1.2: Measurement accuracy +/- 10 MHz or better over 375 nm to 1180 nm; Spec 1.3: Wavelength resolution: 200 kHz or better; Spec 1.4: Photonic Crystal Fiber (endlessly single mode) 1x8 switch integrated, to accommodate simultaneous reading of up to 8 laser inputs without accuracy degradation; Spec 1.5: PID frequency stabilization capability for locking up to 8 laser inputs simultaneously with programmable analog voltage control; Spec 1.6: PID frequency stabilization of multiple lasers can operate simultaneously and automatically with external laser wavemeter calibration to ensure accuracy and long-term stability of frequency stabilization;

Spec 1.7: USB control and GUI software interface; Spec 1.8: Must maintain performance specifications over an air pressure range of 440-900 mm Hg and temperature range of 0-40 C.

Separately Priced Options and Services:

Item 2: [List minimum specifications.

Quantity: [Enter quantity here.] Specifications: [Enter quantity here.]

[List any other item(s) or service requirements here using the same format above.]

Service Requirements / Item Specifications

Section Requirements/Specifications Planning Considerations ☒ N/A

☐ List any visits required to allow potential vendors to see site specific constraints.

[Explain requirement here] ☐ US citizen required on-site (foreign nationals require preregistration 30 days prior) ☐ Vendor rep on-site more than three days ☐ Other considerations: [List other considerations to ensure an accurate quote]

Shipping & Delivery ☐ N/A

[Enter shipping and delivery information by item here.] Shipping criteria:

☐ Partial delivery acceptable ☐ Direct delivery to building required

Requirements / Specifications: High accuracy laser wavelength meter 2

[Explain need for requirement here] Building has: ☐ Loading dock ☐ Containerization preference: [Choose container] Other requirements: [List any other requirements here] Delivery date criteria:

☒ Delivery of goods, installation, and training shall be completed no later than:

14 week ARO ☐ Deliver no earlier than: [Insert date here] ☐ Other: [List other shipping, delivery, or specil requirements]

Electronic Media ☒ N/A

[Enter electronic media requirements by item here.] ☐ Software ☐ Electronic manuals ☐ License Key ☐ Other: [List any other Electronic Media requirements here] [Enter email address for software delivery here]

Installation ☒ N/A

[Enter installation requirements by item here.] ☐ Vendor technician on site ☐ Rigging ☐ Uncrating / Unpackaging ☐ Removal of packaging material ☐ Equipment set up ☐ Start-up services ☐ Turn key installation ☐ Other [Enter details here.] Note: When installation is required, TPOC or COR shall be the coordination point.

Facility / Utility Considerations ☒ N/A

[List site specific utility considerations by item here.] ☐ OFPM Work Order Number(s) [List any Work Order Number(s) here] ☐ Power available Voltage: [Enter voltage here] (V) Max Current: [Enter max current draw here] (A) Phase: [Choose phase] Frequency: [Typically 60Hz] (Hz) NEMA Plug Type: [Enter plug type here] (i.e.: NEMA L5-20P) Environment: [Choose environment] Other power consideration: [List any other power considerations here] ☐ Utility connections: [List relevant available utilities, pressure(s), and flowrate(s)] ☐ Other considerations: [List other considerations to ensure an accurate quote]

Requirements / Specifications: High accuracy laser wavelength meter 3

Warranty ☐ N/A

Base Manufacturer Warranty is acceptable [If you require an Extended Manufacturer Warranty, describe the warranty requirements per item here.] What is included?

☐ Labor ☐ Parts ☐ Travel ☐ Cost and liability for returns ☐ Other: [Enter other requirements here] ☐ Warranty length: [Enter warranty length here] Period Length:[Ex. month, 6 month, year etc.]

Training ☒ N/A

[Training requirements (who, when, what, where) here. If on-site, is travel necessary?] ☐ On-site ☐ Off-site location: [Enter location(s) here] ☐ Number of trainees: [Enter number of trainees here] Scope of training:

☐ Operations ☐ Maintenance ☐ Troubleshooting ☐ Safety ☐ Other: [Enter other scope requirements here] ☐ Training materials to be provided: [Describe requirement here] ☐ Other: [Enter other training requirements here] Note: When training is required, TPOC or COR shall be the coordination point.

Inspection and Acceptance ☐ N/A

[Enter inspection requirements per item here] ☐ Planned government acceptance testing (7 business days) ☒ Gov’t acceptance expected to take longer than 7 business days: 21 days ☐ Inspection and acceptance plan: [Explain here] ☐ Other: [Enter other requirements here]

Maintenance ☒ N/A

[Describe your maintenance requirements per item here] Period Length: [Ex. month, 6 month, year etc.] What is included?

☐ Labor ☐ Spare parts ☐ Consumables ☐ Travel ☐ Cost and liability for returns ☐ Software updates ☐ Manuals ☐ Service plan length: [Enter service plan length here] ☐ Other: [Enter other requirements here] ☐ Requirements: [Enter requirements here] (i.e.: onsite within 72 hours, etc.)

☐ Option Period 1 ☐ Option Period 2 ☐ Option Period 3 ☐ Option Period 4

Government Furnished Material / Equipment ☒ N/A

[List any government furnished material/equipment per item here] ☐ Samples for testing ☐ Government equipment which will leave gov't site (TPOC must coordinate with division property custodian) Serial Number: [Enter Serial Number here] Description: [Enter description here]

Requirements / Specifications: High accuracy laser wavelength meter 4

NIST property number: [Enter NIST property number here] ☐ Other: [Enter other details here]

Travel ☒ N/A

[Describe any travel requirements per item here] ☐ Explain: [Explain here] ☐ Other: [Enter other requirements here] Vendor: To the maximum extent possible, include as part of appropriate sections above (maintenance, repair, etc.). Otherwise, it will be a separate line item, reimbursable per the Federal Travel Regulation.

Provide suggestions to improve this form by clicking here.

https://oaam.nist.gov/OAAMAcquisitions/FormFeedback/default.aspx

Use this document to detail specifications/requirements for an item and/or ancillary services.
TITLE: High accuracy laser wavelength meter
Requesting Lab/Division/Group: PML/Time and Frequency Division (688/80)
General Statement of Need
Provide suggestions to improve this form by clicking here.

File details come from the government source that posted it. Updated .