This document outlines the salient characteristics of an oral surgery chair required under the referenced federal contract opportunity. The chair must have a motorized vertical lift base with a minimum of 15-inch stroke, four independently controlled motors for the back, toe, tilt, and lift, an integrated foot control, a two-piece seat with independent toe section control, an all-steel frame, a surgical armboard and accessory rail system with patient restraint straps, snap-on/snap-off upholstery and a foot protector, and the ability to move the base, back, tilt, and toe. The chair must also have a minimum 450-pound patient weight capacity, trendelenberg and flat positioning capabilities, a tapered backrest, two body restraint straps, and a dual articulating headrest. The solicitation is for the Department of the Air Force Air Education and Training Command seeking dental chairs that meet these salient characteristics.

View the file

Other files

File Type Posted
Sources Sought Synopsis Dental Chairs.pdf PDF

On GovTribe

Work with this file

  • Download the original file
  • Contacts named in this file
  • Similar government files
  • Ask GovTribe AI about this file

Text version

Salient Characteristics

6 of the Oral Surgery Chair. Must have the following features to be equal.

Must have a motorized vertical lift base with a minimum of 15" vertical stroke

Must have 4 independently controlled motors for back, toe, tilt and lift control

Must have integrated fully function foot control

Must have a two‐piece seat with independent control of toe section

Must be all steel frame

Must include a surgical armboard and accessory rail system that includes patient restraint straps

Must have Snap‐on/Snap‐off upholstery and a foot protector

Must be able to move:

o base ‐ Cantilevered Lift – minimum 15 inches o Back ‐ Electric Actuator o Tilt ‐ Electric Actuator o Toe ‐ Electric Actuator

Must have a minimum of 450 pound patient weight capacity

Must have trendelenberg positioning and flat positioning capabilities

Must have a tapered style backrest

Must have two body restraint straps

Must have a dual articulating headrest

Must keep plastic cover for existing power plugs on floor‐ cannot be moved