Tritium Fume Hoods Statement of Work

1.0 OBJECTIVE: The government has a need for eight (8) tritium fume hoods.

2.0 SCOPE: The current Labconco tritium fume hoods are nearing end of life. Anniston Army Depot (ANAD) needs to double the number of tritium fume hoods due to the increase in production. ANAD is continuing the use of the Labconco hoods due to the sensitivity of tritium and the nature of business conducted at ANAD to ensure the safety of the employees and items produced.

3.0 REQUIREMENTS:

3.1. The contractor shall provide eight (8) of the Labconco #120400002 Radioisotope Lab Hood, Labconco #3746702 4 feet Base, Labconco #9743211 Digital Monitor, and Labconco #7180800 Fiberglass Blower or Labconco equivalent blower.

3.2. The contractor shall ship all items to ANAD within 120 days of contract award. The contractor shall provide a standard one (1) year warranty.

3.3. The government will install all equipment listed in 3.1 at ANAD.

3.4The contractor shall provide a minimum three each hard copy and one digital version(s) of commercial off the shelf manuals. Manuals shall include instruction for installation, operations, and maintenance.
4.0SCOPE CHANGES: The Contractor shall not proceed with any verbal or written change(s) to the scope of work from sources other than the Contracting Officer, who will provide changes in writing.